MICRO-LINE GD450MRecalls & reported problems
Aesculap AG · micro-Line handpiece · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04038653444885), received between Jul 20, 2020 and Jul 30, 2021 — classified as 4 injuries.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Overheating of Device2
- Fluid/Blood Leak1
- Leak/Splash1
- Temperature Problem1
Patient problems
- Burn, Thermal2
- No Clinical Signs, Symptoms or Conditions2
- Superficial (First Degree) Burn1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 30, 2021Leak/Splash; Leak/Splash
InjuryJul 30, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH GD450M -MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM. ACCORDING TO THE COMPLAINT DESCRIPTION, WHEN THE DRILL WAS ACTIVATED, OIL OR WATER CAME OUT OF THE TIP AND DRIPPED INTO THE PATIENT'S MOUTH. THE DOCTOR USED SALINE SOLUTION TO RINSE THE PATIENT'S MOUTH RIGHT AWAY, AND ANOTHER TRAY WAS USED TO CONTINUE SURGERY. THE MALFUNCTION PROLONGED THE SURGERY AND REQUIRED MEDICAL INTERVENTION. THIS OCCURRED DURING A LE FORT PROCEDURE. ADDITIONAL INFORMATION WAS NOT PROVIDED. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12253745 ↗InjuryJul 20, 2020Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM. SPECIFICALLY, IT WAS REPORTED THAT DURING AN ORAL SURGERY, THE DRILL BECAME HOT AND BURNT THE PATIENT'S LIP. IT WAS THE PORTION THAT THE DRILL BIT FITS INTO.THE PATIENT'S BURN WAS TREATED WITH HYDROCORTISONE. THE PURPOSE OF THE PROCEDURE WAS TO CORRECT AN OVERBIT; THE JAW WAS CUT IN THREE PLACES AND MOVED OUT. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. THIS EVENT PROLONGED THE SURGERY FOR 10 MINUTES WHILE A NEW DRILL WAS BROUGHT INTO THE ROOM. THE CORD AND CONSOLE USED IN THE PROCEDURE WERE USED WITH A RECIPROCATING SAW WITHOUT AN ISSUE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4). ASSOCIATED MEDWATCH-REPORTS: 9610612-2020-00268 (400476403 + GA824). INVOLVED COMPONENTS: GA824 - ELAN 4 ELECTRO LOW SPEED MOTOR INTRA - 52278515.
Patient problems: Burn, Thermal
MAUDE report 10297735 ↗InjuryJul 20, 2020Overheating of Device; Temperature Problem; Temperature Problem
ASSOCIATED MEDWATCH-REPORTS: 2916714-2020-00243 (B)(4), 2916714-2020-00242. INVOLVED COMPONENTS: GD450M - MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM - SERIAL NO. (B)(6). ADDITIONAL INFORMATION RECEIVED VIA INFORMATION FROM MW5095342: THE PATIENT WAS A SIX YEAR-OLD FEMALE. THE DEVICE WAS USED A PROCEDURE FOR MAXILLARY HYPOPLASIA, MANDIBULAR HYPOPLASIA, APERTOGNATHIA AND ASYMMETRY. THE PATIENT UNDERWENT A LEFORT 1 SEGMENTAL 3-PIECE OSTEOTOMY, BILATERAL SAGITTAL SPLIT OSTEOTOMY AND GENIOPLASTY. UPON USING THE LEDEMANN BUR, IT WAS NOTED THAT THE AESCULAP DRILL BECAME VERY HOT; DURING THE REMOVAL OF THE DRILL FROM THE PATIENT'S MOUTH, IT WAS NOTED THAT THE EXTENSION OF THE HAND PIECE WAS HOT AND HAD BURNED THE PATIENT'S LOWER RIGHT LIP. COLD SALINE WAS PLACED ON THE LIP IMMEDIATELY, AND SILVER SULFADIAZINE WAS PLACED, AND CONTINUED TO BE PLACED THROUGHOUT THE CASE. A NEW DRILL WAS REQUESTED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THE PROCEDURE, THE STERILIZATION PROCESS WAS REVIEWED AND NO FAILURES IN THE STERILIZATION OF THE DEVICE WERE NOTED.
Patient problems: Burn, Thermal; Superficial (First Degree) Burn
MAUDE report 10297665 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.