Olympus WA96205ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Telescope, 4 mm, 70°, connector for light guide on bottom, autoclavable · Class 2
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04042761070072), received between Mar 8, 2023 and May 18, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Break2
- Thermal Decomposition of Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2026Thermal Decomposition of Device
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE ADHESIVE WAS BURNED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25204392 ↗MalfunctionMar 18, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD DETACHED DISTAL TERMINAL LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24625328 ↗MalfunctionAug 11, 2025Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A BURNT LIGHT GUIDE CONNECTING SECTION AND LIGHT GUIDE CONNECTOR. THE EVENT OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22772492 ↗MalfunctionJun 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING REPROCESSING OF THE SUBJECT DEVICE, THE RIDE GUIDE CONNECTING SECTION BASE CAME OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22151157 ↗MalfunctionJan 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS, THE TELESCOPE HAD A BLOTCHY IMAGE DUE TO DEPOSITS ON THE COVER GLASS. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: LIGHT GUIDE POST DETACHED FROM BASE. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18535232 ↗MalfunctionSep 1, 2023Break; Break
THE CUSTOMER ORIGINALLY RETURNED HIS OLYMPUS AUTOCLAVABLE TELESCOPE DUE TO THE UNIT PRODUCING A CLOUDY IMAGE. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED BUT WERE UNKNOWN BY THE INITIAL REPORTER. THERE WAS NO PATIENT HARM OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT. DURING THE DEVICE EVALUATION, IT WAS DISCOVERED THAT THE DEVICE¿S LENS HAD BEEN DAMAGED. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE DAMAGED LENS CONFIRMED DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17670178 ↗MalfunctionMar 8, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE TELESCOPE, 4 MM, 70°, CONNECTOR FOR LIGHT GUIDE ON BOTTOM, AUTOCLAVABLE HAD A BROKEN LIGHT GUIDE POST. THE ISSUE WAS FOUND DURING INSPECTION FOR USE. THE INTENDED PROCEDURE WAS ENDOSCOPIC SINUS SURGERY. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16503995 ↗
Other Nasopharyngoscope (Flexible Or Rigid) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.