Skip to content

EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180 (UDI 04953170339325)Recalls & reported problems

Olympus Medical Systems Corp. · BF-1T180 VIDEOSCOPE 6.0MM DIA 3.0MM CH · Class 2

All models of EVIS EXERA II BRONCHOVIDEOSCOPE (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA II bronchoscope is a camera scope for the airways. Its 405 adverse event reports often describe the image going blank, burnt tip covers, and microbial contamination, and some describe bacterial infections. The reports include 1 death and 49 injuries. All 3 recalls are open, over laser burning, instruction updates and a scope size that can jam in a connector. Reports fell from 98 in the earlier year to 27 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
408
Deaths reported
1
Injuries reported
49
Recalls naming it
3

Summary

The FDA's MAUDE database contains 408 adverse event reports that list this exact device (DI 04953170339325), received between Aug 8, 2018 and Sep 2, 2026 — classified as 1 death, 49 injuries, 358 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (162 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 17 reportsDec 2023: 10 reportsJan 2024: 10 reportsFeb 2024: 14 reportsMar 2024: 12 reportsApr 2024: 12 reportsMay 2024: 5 reportsJun 2024: 7 reportsJul 2024: 7 reportsAug 2024: 5 reportsSep 2024: 5 reportsOct 2024: 8 reportsNov 2024: 74 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 4 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 6 reportsJan 2026: 5 reportsFeb 2026: 8 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202674
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2025Class IOpen, Classified

    Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.

    Z-0046-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2493-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  3. Jun 8, 2023Class IOpen, Classified

    There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.

    Z-2016-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Display/Image162
  • Thermal Decomposition of Device80
  • Microbial Contamination of Device45
  • Peeled/Delaminated32
  • Failure to Clean Adequately22
  • Device Reprocessing Problem17
  • Melted14
  • Break13

Patient problems

  • No Clinical Signs, Symptoms or Conditions354
  • Bacterial Infection25
  • Unspecified Infection14
  • Fungal Infection5
  • Pneumothorax3
  • Hemorrhage/Blood Loss/Bleeding2
  • Insufficient Information2
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 408 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2026No Display/Image; Loose or Intermittent Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50563525 ↗
  • MalfunctionSep 1, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50542661 ↗
  • MalfunctionAug 13, 2026No Display/Image

    Patient problems: Insufficient Information

    MAUDE report 50230619 ↗
  • MalfunctionJul 30, 2026Device Reprocessing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50085265 ↗
  • MalfunctionJul 27, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50042395 ↗
  • MalfunctionApr 30, 2026Break; Crack

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A CLOG IN THE CHANNEL. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, A CHIPPED DISTAL END PLASTIC COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050492 ↗
  • MalfunctionApr 9, 2026Crack

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CRACKED/CHIPPED PLASTIC DISTAL END COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831583 ↗
  • MalfunctionApr 8, 2026Thermal Decomposition of Device

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, COVER BURNED WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816581 ↗
  • MalfunctionFeb 18, 2026Melted

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOVIDEOSCOPE DISTAL END C-COVER HAD A BURNT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24378975 ↗
  • MalfunctionFeb 18, 2026Melted

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE PLASTIC DISTAL END COVER WAS BURNED ON THE BRONCHOVIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377401 ↗
  • MalfunctionFeb 18, 2026Melted

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE WAS BURNT ON THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376593 ↗
  • MalfunctionFeb 14, 2026Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE DISTAL SHEATH WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24341240 ↗
  • MalfunctionFeb 11, 2026No Display/Image; Key or Button Unresponsive/not Working

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24312314 ↗
  • MalfunctionFeb 11, 2026Melted

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE PLASTIC DISTAL END COVER HAD A BURN MARK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24311070 ↗
  • MalfunctionFeb 9, 2026Thermal Decomposition of Device

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED A BURNT DISTAL END COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24298832 ↗
  • MalfunctionFeb 6, 2026Melted

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE PLASTIC DISTAL END COVER WAS BURNED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24277793 ↗
  • MalfunctionJan 27, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED MISSING SEALING GLUE ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192839 ↗
  • InjuryJan 20, 2026Break; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: Foreign Body In Patient

    MAUDE report 24123454 ↗
  • MalfunctionJan 5, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE IMAGE WAS DISTORTED ON THE BRONCHOVIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23957935 ↗
  • MalfunctionJan 2, 2026Component Missing; Detachment of Device or Device Component

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE BRONCHOVIDEOSCOPE HAD THE OBJECTIVE LENS (IMAGE LENS) MISSING GLUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23945114 ↗
  • MalfunctionJan 2, 2026Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE HAD INTERMITTENT FLICKER OF IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944889 ↗
  • MalfunctionDec 29, 2025No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT FLEX VIDEO SCOPE EXHIBITED NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23915417 ↗
  • MalfunctionDec 29, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT FLEX VIDEO SCOPE EXHIBITED A FAILED FLICKER CHECK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23913385 ↗
  • MalfunctionDec 26, 2025Detachment of Device or Device Component; Scratched Material; Detachment of Device or Device Component; Scratched Material

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE BRONCHOVIDEOSCOPE HAD THE DISTAL END, C COVER HAVE DENTS, SCRATCHES AND DETACHED FROM BENDING SECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23895651 ↗
  • MalfunctionDec 23, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE'S DISTAL TIP WAS BLOWN DUE TO A MISSING A-RUBBER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23885047 ↗