OLYMPUS Winter & Ibe GmbH · Telescope "ULTRA", 4 mm, 0° · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus WA4KS400 is a 4 mm rigid telescope used in nasopharyngoscopy to view the nose and throat. The 38 adverse event reports mostly describe the scope breaking, cracking or chipping, along with scratched surfaces and a damaged lens or tip cover glass, with no patient injuries or deaths reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 04042761083034), received between Jan 17, 2023 and Sep 25, 2026 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Break” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 6 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break24
  • Scratched Material6
  • Crack5
  • Detachment of Device or Device Component2
  • Defective Component1
  • Degraded1
  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions40

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Crack

    During on site inspection of Olympus Rigid Rhinoscope, it was found chipped distal end. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895659 ↗
  • MalfunctionSep 10, 2026Break

    During inspection of rigid rhinoscope, it was found distal end chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50704256 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50508073 ↗
  • MalfunctionAug 13, 2026Crack

    No additional information received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50240872 ↗
  • MalfunctionMay 10, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID SCOPE EXHIBITED CHIPPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25133701 ↗
  • MalfunctionMar 11, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD SCATTERED OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24571931 ↗
  • MalfunctionFeb 10, 2026Crack

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306171 ↗
  • MalfunctionFeb 10, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE EXHIBITED A DAMAGED INSERTION TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305197 ↗
  • MalfunctionFeb 10, 2026Break

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD DAMAGED COVER GLASS TIP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305090 ↗
  • MalfunctionFeb 3, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD TIP COVER GLASS DAMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251441 ↗
  • MalfunctionFeb 3, 2026Scratched Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD SCRATCHES ON THE EYEPIECE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251350 ↗
  • MalfunctionFeb 3, 2026Scratched Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD TIP COVER GLASS DAMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251186 ↗
  • MalfunctionJan 30, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID RHINOSCOPE HAD DAMAGE TO THE TIP COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24220412 ↗
  • MalfunctionJan 13, 2026Crack

    IT WAS REPORTED THAT THE RIGID RHINOSCOPE HAD A CRACKED LENS. THIS ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24054940 ↗
  • MalfunctionDec 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE RIGID SCOPE HAD THE ADAPTER CONNECT TO THE LIGHT CABLE CAME OFF. THE ISSUE WAS FOUND DURING A SEDATION FOR A THERAPEUTIC OTOLARYNGOLOGY PROCEDURE FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23774913 ↗
  • MalfunctionOct 16, 2025Crack; Crack

    IT WAS REPORTED THAT THE RIGID SCOPE HAD A CRACKED LENS AT THE DISTAL END. THE ISSUE WAS OBSERVED DURING INSTALLATION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23319785 ↗
  • MalfunctionSep 3, 2025Scratched Material; Scratched Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED A SCRATCH ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22954330 ↗
  • MalfunctionSep 3, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED BROKEN LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951473 ↗
  • MalfunctionJul 3, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD BROKEN LENS DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22394767 ↗
  • MalfunctionMay 19, 2025Break; Break

    IT WAS REPORTED THE LENS OF THE RIGID SCOPE WAS ¿SUSPECTED TO BE BROKEN¿, FOUND DURING SET UP. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22047766 ↗
  • MalfunctionMar 27, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED BROKEN INTERNAL LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21707280 ↗
  • MalfunctionMar 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21698917 ↗
  • MalfunctionMar 7, 2025Poor Quality Image; Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE HAD EXHIBITED DAMAGE TO THE INTERNAL LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21545600 ↗
  • MalfunctionFeb 21, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE HAD BROKEN LENS DURING SET UP/ INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21436712 ↗
  • MalfunctionFeb 10, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21350158 ↗