Modular universal table top 115030B0Recalls & reported problems
MAQUET GmbH · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046768003347), received between Mar 7, 2024 and Dec 6, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close3
- Unintended Movement3
- Failure to Conduct1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 6, 2024Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20864954 ↗MalfunctionNov 18, 2024Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR TABLE TOPS - 115030B0 - MODULAR UNIVERSAL TABLE TOP. AS IT WAS STATED, THE UPPER BACK PART OF THE OPERATING BOARD RELEASED UPON LOADING WITH AN ANESTHETIZED PATIENT WITH A HIGHER WEIGHT RESULTING IN CHANGE IN THE PATIENT'S POSITION DURING UNSPECIFIED LEG/KNEE SURGERY. THE STAFF WAS ABLE TO CAPTURE AND STABILIZE THE LOOSE BACK PART WITH THE PATIENT. THE INCIDENT RESULTED IN A NON-SIGNIFICANT DELAY. ACCORDING TO PROVIDED INFORMATION, AT THE TIME OF INCIDENT, THE PATIENT WAS SECURED WITH BELTS. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY UNINTENDED MOVEMENT OF THE TABLE DURING SURGERY CREATING A RISK OF NERVE OVERSTRETCHING, WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20704044 ↗MalfunctionApr 30, 2024Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement
ON 16TH APRIL 2024, GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR TABLE TOPS - 115030B0 - MODULAR UNIVERSAL TABLE TOP. AS IT WAS STATED, A HEAD REST LOWERED DURING A SURGERY. IT WAS CONFIRMED THAT THE MOVEMENT WAS FAST AND AN ANESTHETIZED PATIENT WAS ON THE TABLE WHEN THE ISSUE OCCURRED. ACCORDING TO THE PROVIDED INFORMATION, THE HEAD RESTRAINT LATCHES DID NOT LOCK AND GAVE IN EASILY. FOLLOWING SERVICE VISIT ON SITE, IT WAS CONFIRMED THAT THERE WAS AN ISSUE AT THE TABLE TOP. THE MOTORS WERE MISALIGNED AND THE HEAD REST FIXING SYSTEM WAS WORN. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY UNINTENDED MOVEMENT OF THE TABLE CREATING A RISK OF NERVE OVERSTRETCHING, WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19211346 ↗MalfunctionMar 7, 2024Failure to Conduct; Positioning Failure; Failure to Conduct; Positioning Failure
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18855305 ↗
Other TABLE, OPERATING-ROOM, AC-POWERED devices
- ORT300 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls21 reports
- ORT200 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls12 reports
- MEERA mobile operating table 720001F2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- MEERA mobile operating table 720001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall19 reports
- MAGNUS operating table system 118001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.