Modular universal table top 115030B0Recalls & reported problems

MAQUET GmbH · Class 1

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046768003347), received between Mar 7, 2024 and Dec 6, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close3
  • Unintended Movement3
  • Failure to Conduct1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2024Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20864954 ↗
  • MalfunctionNov 18, 2024Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR TABLE TOPS - 115030B0 - MODULAR UNIVERSAL TABLE TOP. AS IT WAS STATED, THE UPPER BACK PART OF THE OPERATING BOARD RELEASED UPON LOADING WITH AN ANESTHETIZED PATIENT WITH A HIGHER WEIGHT RESULTING IN CHANGE IN THE PATIENT'S POSITION DURING UNSPECIFIED LEG/KNEE SURGERY. THE STAFF WAS ABLE TO CAPTURE AND STABILIZE THE LOOSE BACK PART WITH THE PATIENT. THE INCIDENT RESULTED IN A NON-SIGNIFICANT DELAY. ACCORDING TO PROVIDED INFORMATION, AT THE TIME OF INCIDENT, THE PATIENT WAS SECURED WITH BELTS. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY UNINTENDED MOVEMENT OF THE TABLE DURING SURGERY CREATING A RISK OF NERVE OVERSTRETCHING, WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20704044 ↗
  • MalfunctionApr 30, 2024Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement

    ON 16TH APRIL 2024, GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR TABLE TOPS - 115030B0 - MODULAR UNIVERSAL TABLE TOP. AS IT WAS STATED, A HEAD REST LOWERED DURING A SURGERY. IT WAS CONFIRMED THAT THE MOVEMENT WAS FAST AND AN ANESTHETIZED PATIENT WAS ON THE TABLE WHEN THE ISSUE OCCURRED. ACCORDING TO THE PROVIDED INFORMATION, THE HEAD RESTRAINT LATCHES DID NOT LOCK AND GAVE IN EASILY. FOLLOWING SERVICE VISIT ON SITE, IT WAS CONFIRMED THAT THERE WAS AN ISSUE AT THE TABLE TOP. THE MOTORS WERE MISALIGNED AND THE HEAD REST FIXING SYSTEM WAS WORN. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY UNINTENDED MOVEMENT OF THE TABLE CREATING A RISK OF NERVE OVERSTRETCHING, WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19211346 ↗
  • MalfunctionMar 7, 2024Failure to Conduct; Positioning Failure; Failure to Conduct; Positioning Failure

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18855305 ↗

Other TABLE, OPERATING-ROOM, AC-POWERED devices

All TABLE, OPERATING-ROOM, AC-POWERED devices →