Corded control device 118090A0Recalls & reported problems

MAQUET GmbH · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046768032217), received between Jan 5, 2024 and Apr 5, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Electrical Shorting” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical Shorting1
  • Positioning Failure1
  • Unintended Movement1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2024Use of Device Problem; Unintended Movement; Use of Device Problem; Unintended Movement

    GETINGE BECAME AWARE OF AN EVENT WITH 118090A0 - CABLE-CONNECTED HAND CONTROL USED WITH 118001A0 - MAGNUS TABLE COLUMN FOR BUILT-IN PLATE AND 118091A0 - IR HAND CONTROL MAGNUS. AS IT WAS STATED, THE EVENT OCCURRED AFTER THE MEDICAL PROCEDURE WAS FINISHED, THE PATIENT WAS EXTUBATED AND STILL ON THE TABLETOP. THE CABLE-CONNECTED HAND CONTROL WAS PLACED ON THE SIDERAIL BETWEEN TABLETOP AND ARM REST. AFTER MOVING THE ARM REST TOWARDS THE TABLETOP, SLOW UNINTENDED MOTION INTO "HEAD DOWN" POSITION OCCURRED DUE TO ACCIDENTAL PRESSING OF HAND CONTROL BUTTONS WITH THE ARM REST. THE STAFF TRIED TO STOP THE MOVEMENT WITH THE IR CONTROLLER WHICH DID NOT WORK. ACCORDING TO THE INFORMATION PROVIDED, IT TOOK SOME TIME FOR THE STAFF TO DETERMINE THAT THE ISSUE WAS CAUSED BY CABLE-CONNECTED HAND CONTROL AND TO RELEASE IT. ACCORDING TO THE CUSTOMERS ALLEGATION IF THE MOTION HAD CONTINUED, THE PATIENT COULD HAVE BEEN AT RISK OF FALLING OFF THE OPERATING TABLE. THE EVENT WAS INITIALLY ASSESSED AS REPORTABLE DUE TO THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY UNINTENDED MOVEMENT OF THE TABLE AND CREATING A RISK OF FALL, WAS TO REOCCUR. RECENTLY, THE CASE HAS BEEN REEVALUATED. AFTER THE CON

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19048934 ↗
  • MalfunctionJan 5, 2024Positioning Failure; Electrical Shorting; Positioning Failure; Electrical Shorting

    ON 20TH DECEMBER 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR COLUMNS - 118001C0 - MAGNUS TABLE COLUMN, MOBILE. AS IT WAS STATED, NO MOVEMENT OF THE TABLE WAS POSSIBLE AT THE END OF LAST INTERVENTION OF THE DAY. IT IS UNCLEAR WHETHER THE PATIENT WAS STILL ON THE OPERATING TABLE WHEN THE ISSUE OCCURRED. ADDITIONALLY, A SURGERY WAS CANCELLED ON THE FOLLOWING DAY DUE TO THE MALFUNCTION. ACCORDING TO INFORMATION PROVIDED BY THE SERVICE TECHNICIAN FOLLOWING VISIT ON SITE, THE TABLE COULD NOT BE TEMPORARILY OPERATED VIA OVERRIDE PANEL AND THE HANDCONTROL WAS SHORT-CIRCUITED AND NOT WORKING AS WELL. HOWEVER, THE TABLE COULD BE MOVED USING ANOTHER HANDCONTROL. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY THE OPERATING TABLE NOT RESPONDING TO REMOTE CONTROL OR OVERRIDE PANEL LEADING TO INABILITY TO POSITION THE TABLE DURING PROCEDURE, WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18452382 ↗

Other TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED devices

All TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED devices →