Skip to content

MAGNUS 1180.16F6Recalls & reported problems

MAQUET GmbH · MAGNUS carbon-fibre table top · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046768118355), received between Jan 28, 2025 and Jan 28, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Interrogate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Interrogate1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2025Positioning Failure; Failure to Interrogate; Positioning Failure; Failure to Interrogate

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR TABLE TOP - 118016F6 - MAGNUS CARBON FIBRE TABLE TOP, 205CM BOLUS, USA USED WITH PHILIPS C-ARM DEVICE. AS IT WAS STATED, THE TABLE MOVEMENTS STOPPED WORKING AND THE FOLLOWING ALERT WAS DISPLAYED: "PHILIPS TUBE NOT WORKING. USE AT YOUR OWN RISK." ACCORDING TO THE PROVIDED INFORMATION, THE STAFF COULD NO LONGER SLIDE THE TABLE TOP BUT WERE ABLE TO MOVE THE PHILIPS C-ARM TO GET INTO THE SAME POSITION AND USE THE OVERRIDE FUNCTIONS TO MOVE THE COLUMN UP AND DOWN. DUE TO THE ISSUE, THE PATIENT WAS MOVED TO ANOTHER TABLE, WHICH CAUSED A DELAY OF APPROXIMATELY 10 MINUTES IN THE ENDOVASCULAR PROCEDURE. ACCORDING TO INFORMATION PROVIDED FOLLOWING SERVICE VISIT ON SITE, THE LATERAL SHIFT DRIVE AND CONTROL UNIT WERE REPLACED. THE INCIDENT WAS INITIALLY ASSESSED AS REPORTABLE DUE TO A PROCEDURAL DELAY RELATED TO TRANSFER OF THE PATIENT TO ANOTHER OPERATING ROOM DURING PROCEDURE. RECENTLY, THE INCIDENT HAS BEEN REEVALUATED. IT HAS BEEN CONCLUDED THAT THE PROCEDURAL DELAY HAD MINOR CLINICAL RELEVANCE IN THIS SITUATION AND DID NOT LEAD TO ANY HEALTH CONSEQUENCES FOR THE PATIENT. THEREFORE, THE SCENARIO DESCRIBED IN THE RECORD IS CONSIDERED AS NON-R

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21251956 ↗

Other Table, Radiologic devices

All Table, Radiologic devices →

Sources and updates