CORIN mobile operating table 770001A2Recalls & reported problems

MAQUET GmbH · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046768145184), received between Apr 17, 2025 and Jun 6, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 14, 2025Open, Classified

    Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

    Z-2420-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Positioning Failure3
  • Failure to Interrogate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2025Positioning Failure; Positioning Failure

    GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES - 770001A2 - CORIN W AUTODRIVE, IPC, EU. AS IT WAS STATED, THE OPERATING TABLE STOPPED WORKING AND WAS NOT RESPONDING TO THE IR AND CABLE REMOTE CONTROLS AS WELL AS THE OVERRIDE PANEL AT THE BEGINNING OF THE COLECTOMY SURGERY. THE TABLE ALSO DID NOT WORK WHEN CONNECTED TO THE MAINS. THE PATIENT HAD TO BE MOVED TO ANOTHER OPERATING TABLE, WHICH RESULTED IN PROLONGED ANESTHESIA TIME. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY THE OPERATING TABLE NOT RESPONDING TO THE REMOTE CONTROL OR OVERRIDE PANEL, LEADING TO THE INABILITY TO POSITION THE TABLE DURING SURGERY, WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22157422 ↗
  • MalfunctionMay 23, 2025Positioning Failure; Failure to Interrogate; Positioning Failure; Failure to Interrogate

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077628 ↗
  • MalfunctionApr 17, 2025Positioning Failure; Positioning Failure

    MANUFACTURER'S REFERENCE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856853 ↗

Other Table, Operating-Room, Ac-Powered devices

All Table, Operating-Room, Ac-Powered devices →