Skip to content

AESCULAP PV010Recalls & reported problems

Aesculap AG · AESCULAP AEOS · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
3

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04046955206742), received between Mar 10, 2021 and Jun 28, 2021 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (2 reports).

3 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. May 13, 2024Class IIOpen, Classified

    Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

    Z-2139-2024 · Root cause: Process control · Recall details · FDA record ↗

  2. Jul 14, 2022Class IITerminated

    Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recovery: 3. Autofocus not available:

    Z-1516-2022 · Root cause: Software design · Recall details · FDA record ↗

  3. Feb 19, 2021Class IICompleted

    Malfunction of the robotic arm preventing further movement of the robotic arm and requiring restart of the robotic digital microscope system.

    Z-1409-2022 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem2
  • Unexpected Shutdown2
  • Break1
  • Display or Visual Feedback Problem1
  • Loose or Intermittent Connection1
  • Material Integrity Problem1
  • Mechanical Problem1
  • Operating System Becomes Nonfunctional1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 28, 2021Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PV010 - AESCULAP AEOS. ACCORDING TO THE COMPLAINT DESCRIPTION, THE AESCULAP AEOS SOFTWARE CRASHED AFTER BEING DRAPED AND WHEN READY TO BE PUSHED IN FOR SURGERY. AFTER MULTIPLE REBOOT ATTEMPTS AND HARD RESTARTS TO THE ENTIRE AEOS SYSTEM, THE SURGEON DECIDED TO USE THE TRADITIONAL MICROSCOPE. THERE WAS A SURGICAL DELAY OF 5-10 MINUTES. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. THIS OCCURRED DURING THE BEGINNING OF AN ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) PROCEDURE; THE PATIENT WAS UNDER ANESTHESIA AND THE INITIAL INCISION HAD BEEN MADE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12079265 ↗
  • InjuryJun 28, 2021Operating System Becomes Nonfunctional; Unexpected Shutdown; Operating System Becomes Nonfunctional; Unexpected Shutdown

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PV010 - AESCULAP AEOS. ACCORDING TO THE COMPLAINT DESCRIPTION, THE AESCULAP AEOS SOFTWARE CRASHED AFTER BEING DRAPED AND WHEN READY TO BE PUSHED IN FOR SURGERY. AFTER MULTIPLE REBOOT ATTEMPTS AND HARD RESTARTS TO THE ENTIRE AEOS SYSTEM, THE SURGEON DECIDED TO USE THE TRADITIONAL MICROSCOPE. THERE WAS A SURGICAL DELAY OF 5-10 MINUTES. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. THIS OCCURRED DURING THE BEGINNING OF AN ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) PROCEDURE; THE PATIENT WAS UNDER ANESTHESIA AND THE INITIAL INCISION HAD BEEN MADE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12076969 ↗
  • MalfunctionApr 20, 2021Reset Problem; Reset Problem

    ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00322 (400508474 PV010), 9610612-2021-00323 (400508475 PV010).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11697540 ↗
  • MalfunctionApr 20, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem

    ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00322 (400508474 PV010), 9610612-2021-00323 (400508475 PV010).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11697539 ↗
  • MalfunctionApr 20, 2021Computer Software Problem; Computer Software Problem

    ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00321 (400508468 PV010), 9610612-2021-00323 (400508475 PV010).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11697538 ↗
  • MalfunctionApr 7, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED TO AESCULAP INC. THAT A AESCULAP AEOS (PART # PV010) WAS BEING SETUP FOR PRODUCT DISPLAY. ACCORDING TO THE COMPLAINANT, THE AEOS UNIT WAS SETUP IN A HOTEL CONFERENCE CENTER TO DISPLAY TO LOCAL SURGEONS. REPORTEDLY, DURING SETUP TWO (2) HARDWARE ISSUES WERE OBSERVED: THE WHEEL OF THE SYSTEM WAS BROKEN; RIGHT HAND CONTROL WAS LOOSE. NO PATIENT INVOLVEMENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE XC (B)(4). ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00275 ((B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628140 ↗
  • MalfunctionApr 7, 2021Break; Loose or Intermittent Connection

    IT WAS REPORTED TO AESCULAP INC. THAT A AESCULAP AEOS (PART # PV010) WAS BEING SETUP FOR PRODUCT DISPLAY. ACCORDING TO THE COMPLAINANT, THE AEOS UNIT WAS SETUP IN A HOTEL CONFERENCE CENTER TO DISPLAY TO LOCAL SURGEONS. REPORTEDLY, DURING SETUP TWO (2) HARDWARE ISSUES WERE OBSERVED: THE WHEEL OF THE SYSTEM WAS BROKEN. RIGHT HAND CONTROL WAS LOOSE. NO PATIENT INVOLVEMENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE XC (B)(4). ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00276 ((B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628042 ↗
  • MalfunctionMar 10, 2021Computer Software Problem; Mechanical Problem; Computer Software Problem; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PV010 - AESCULAP AEOS. ACCORDING TO THE COMPLAINT DESCRIPTION, THE AEOS SYSTEM GENERATED REBALANCE (20 SECONDS) HAPPENS EVERY TIME AEOS IS MOVED INTO "DRAPE" OR "DRAPE STANDBY." IN THE PAST, THE LONGER REBALANCE WOULD ONLY OCCUR WHEN THE SYSTEM WAS INITIALLY MOVED INTO "DRAPE" POSITION. THE LONGER REBALANCE ESPECIALLY WHEN BRINGING THE SYSTEM INTO SURGERY CAUSES A DELAY UNTIL THE SYSTEM COMPLETES THE ACTION. THE DOCTOR COMPLETED THE CASE WITH AEOS AND THE PATIENT WOKE UP FROM ANESTHESIA NORMALLY. THERE WAS NO DESCRIBED PATIENT HARM. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11457003 ↗

FDA clearance and classification

The FDA regulates product code FSO (microscope, surgical, general & plastic surgery) under 21 CFR 878.4700 as a Class I device. 21 CFR 878.4700 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AESCULAP PV010

Has the AESCULAP PV010 been recalled?

Yes. 3 FDA recall records name this device's identifier: 1 still open, 1 completed by the company but not yet closed by the FDA and 1 closed (terminated) by the FDA. The most recent, Z-2139-2024 (Class II), began on May 13, 2024: Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the AESCULAP PV010?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Mar 10, 2021 and Jun 28, 2021. In those reports, the most common specific device problems are “Computer Software Problem” (2), “Unexpected Shutdown” (2) and “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AESCULAP PV010?

None of the reports are classified as deaths, 2 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AESCULAP PV010 FDA approved?

Devices of this type (product code FSO) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Aesculap AG.

Who makes the AESCULAP PV010, and is it still sold?

The FDA's device database lists Aesculap AG as the company responsible for it (brand AESCULAP, model PV010). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other MICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY devices

All MICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources and updates