Skip to content

Universal Robot UR5 arm, PV010204 subcomponent of Aesculap… recallZ-1409-2022 · Aesculap Implant Systems LLC · initiated Feb 19, 2021

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aesculap Implant Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Malfunction of the robotic arm preventing further movement of the robotic arm and requiring restart of the robotic digital microscope system.

FDA root cause category: Material/Component Contamination

What the company is doing

An Urgent Market Withdrawal Notification letter issued February 19, 2021 was mailed USPS certified. The letter identifies the Universal Robot UR5 arm as part of hte Aesculap Aeos System and requests the customer return the acknowledgement form and schedule repair by service technician.

Recall details

Recall number
Z-1409-2022
Status
Completed
Initiated by the company
Feb 19, 2021
Recalling firm
Aesculap Implant Systems LLC
Quantity
1
Product code
FSO
Distribution
United States: NE

Product description

Universal Robot UR5 arm, PV010204 subcomponent of Aesculap Aeos Digital Microscope System, PV010

Code information (models, lots, serial numbers)

UDI-DI: 04046955206742, Serial Number: SN1004

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page