Skip to content

Design Options® 552153Recalls & reported problems

B. Braun Medical Inc. · CESK BEN TAUB · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046955214426), received between Nov 9, 2020 and May 9, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2025Open, Classified

    Potential for the lid of the catheter connector to be in the incorrect position.

    Z-0126-2026 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2025Break; Break

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: BROKEN/SNAPPED FX CATHETER. DETAILED INQUIRY DESCRIPTION: END USER REPORTING A BROKEN FX CATHETER USED ON ONE OF THEIR OR PATIENTS PLACED FOR POST-OP PAIN MANAGEMENT. PATIENT WAS MOVED TO A LATERAL POSITION AND THAT IS WHEN THE CATHETER SNAPPED WITH THE SPRINGWOUND COIL EXPOSED. PER THE DOCTOR, NOTHING UNUSUAL NOTED ABOUT THE CATHETER AND THERE WAS NO EXCESSIVE FORCE OR PULLING OF CATHETER TO HAVE CAUSED THIS. THEY SUSPECT THERE WAS A MATERIAL DEFECT IN THE CATHETER MATERIAL. THEY WERE ONLY VISUALLY ALERTED BY IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21997909 ↗
  • MalfunctionNov 9, 2020Difficult to Remove; Difficult to Remove

    AS REPORTED BY THE USER FACILITY: "THE CATHETER KNOTTED WHILE PLACING CSE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10810486 ↗

Other Epidural anesthesia kit devices

All Epidural anesthesia kit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.