Skip to content

CARESAFE™ 354304Recalls & reported problems

B. Braun Medical Inc. · CARESAFE IV ADMIN SET w/AIRSTOP, 114 IN. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046955259960), received between Nov 16, 2023 and Nov 16, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Free or Unrestricted Flow” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0623-2026 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Free or Unrestricted Flow1
  • Inaccurate Flow Rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 16, 2023Inaccurate Flow Rate; Free or Unrestricted Flow; Inaccurate Flow Rate; Free or Unrestricted Flow

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: ROLLER CLAMP IS NOT WORKING PROPERLY. DETAILED INQUIRY DESCRIPTION: WE HAD ISSUES WITH CONTROLLING THE RATE OF FLOW WITH THE ROLLERBALL. IT HAD TO BASICALLY BE CLOSED TO REDUCE THE DRIP RATE MAKING IT DIFFICULT TO TITRATE THE FLOW. MANY PATIENTS RECEIVED ALMOST THE ENTIRE LITER OF FLUID BEFORE REACHING THE OR EVEN WHEN THE NURSES FELT THEY HAD ADEQUATELY CLOSED THE ROLLERBALL. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18151034 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page