Design Options® 552176Recalls & reported problems
B. Braun Medical Inc. · CESK ALLEGHENY GENERAL HOSPITAL · Class 2
All models of Design Options (116)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046955589487), received between Aug 13, 2024 and Oct 10, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Sep 18, 2025Open, Classified
Potential for the lid of the catheter connector to be in the incorrect position.
Z-0126-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break3
Patient problems
- Needle Stick/Puncture2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 10, 2024Break; Break
AS REPORTED BY USER FACILITY, IDENTICAL ISSUES AT WEXFORD. WE HAVE ALSO HAD MANY ISSUES WITH INJECTING THROUGH THE CATHETER AFTER PLACEMENT. FACILITY CONTACT WHO REPORTED THAT DURING OB CASES THE TEST DOSE CAN NOT BE DELIVERED AND THAT SOME OF THE CATHETERS ARE SHEARING. SHE REPORTED THAT THE SAME OUTCOME WAS REQUIRED AS NO BROKEN PIECES FELL OFF IN THE PATIENT BUT A NEW KIT WAS NEEDED AND A NEW CATHETER NEEDED TO BE PLACED.
Patient problems: Needle Stick/Puncture
MAUDE report 20417803 ↗MalfunctionOct 1, 2024
UNEVENTFUL EPIDURAL PLACED FOR LABOR. AFTER PLACEMENT, UNABLE TO FLUSH CATHETER. CATHETER DEPTH ADJUSTED WITH NO CORRECTION OF PROBLEM. CATHETER HAD TO BE REMOVED, AND NEW CATHETER PLACED. NURSE PERSONALLY FLUSHED THE AFFECTED CATHETER. ALL THREE PORTS FLUSHED WITHOUT DIFFICULTY. CATHETER WILL BE RETURNED TO COMPANY FOR FURTHER EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20343679 ↗MalfunctionAug 13, 2024Break; Break
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION FX CATHETERS SHEARING AND CANNOT INJECT TEST DOSE. DETAILED INQUIRY DESCRIPTION FX CATHETERS SHEARING AND CANNOT INJECT TEST DOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19972424 ↗MalfunctionAug 13, 2024Break; Break
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION FX CATHETERS SHEARING AND KINKING. DETAILED INQUIRY DESCRIPTION TEST DOSE CAN'T BE DELIVERED FOR 552176 AND SOME CATHETERS SHEARING. 1. WERE ANY PIECES OF THE BROKEN CATHETER RETAINED IN THE PATIENT? NO 2. WAS IMAGING REQUIRED? NO 3. WHAT TYPE OF MEDICAL INTERVENTION(S) WERE REQUIRED? NEW CATHETER/KIT WAS USED 4. ARE YOU ABLE TO PROVIDE THE PATIENT GROUP THAT THE DEVICES WERE BEING ADMINISTERED TO (OB, ORTHO, OR OTHER TYPE OF PATIENT)? OB 5. WHAT WAS THE OUTCOME AND PATIENT STATUS? NEW CATHETER PLACED THE ISSUE AT FORBES WAS NOT PLACING THE CATHETER (INSERTION WAS EASY), BUT RATHER BEING UNABLE TO INJECT THROUGH THE CATHETER ONCE PLACED. ADDITIONALLY, THERE HAS BEEN A CATHETER THAT DISCONNECTED DUE TO EITHER BEING BROKEN R DETACHED AT THE ALLIGATOR CLIP.
Patient problems: Needle Stick/Puncture
MAUDE report 19972341 ↗
Other Epidural anesthesia kit devices
- Design Options® 552116B. Braun Medical Inc. · Epidural anesthesia kit2 recalls7 reports
- Design Options® 552126B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552179B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552177B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Perifix® 332079B. Braun Medical Inc. · Epidural anesthesia kit1 recall45 reports
- Perifix® 332097B. Braun Medical Inc. · Epidural anesthesia kit1 recall39 reports
- Perifix® 332220B. Braun Medical Inc. · Epidural anesthesia kit1 recall21 reports
- Perifix® 332078B. Braun Medical Inc. · Epidural anesthesia kit1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page