Caresite Micro™ 354268Recalls & reported problems
B. Braun Medical Inc. · CARESITE Micro 8" Standard Bore Extension Set · Class 2
All models of Caresite Micro (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04046955773022), received between Apr 25, 2025 and Jul 9, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash5
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Chemical Exposure1
- Exposure to Body Fluids1
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 9, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CRACKED TUBING/LEAKING. DETAILED INQUIRY DESCRIPTION: THE TUBING SEEMED TO BE CRACKED/LEAKING NEAR THE HUB, THUS HAVING BLOOD LEAK OUT ONTO THE TUBING, PATIENT AND NURSE.
Patient problems: Exposure to Body Fluids; Hemorrhage/Blood Loss/Bleeding
MAUDE report 22457201 ↗MalfunctionJun 15, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 1: BRIEF INQUIRY DESCRIPTION: CARESITE CAP FALLING OFF EXTENSION SET. DETAILED INQUIRY DESCRIPTION: TWO TIMES THIS WEEK, THE END CAP ON THE 8 INCH CARESITE MICRO SET WITH REMOVABLE CARESITE MICRO LUER ACCESS DEVICE AND SPIN LOCK CONNECTOR FELL OFF WHILE DRAWING LABS ON A PATIENT. IF THE END CAP IS REMOVED, THERE IS NO VALVE KEEPING THE BLOOD FLOW IN. WHEN RN WAS DRAWING LABS, THE END CAP OF THE EXTENSION SET FELL OFF CAUSING PATIENT'S BLOOD FLOWS OUT OF THE PIV. BLOOD WENT ON THE PATIENT, AND IT CAUSED A LARGE MESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22208625 ↗MalfunctionJun 15, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 2: BRIEF INQUIRY DESCRIPTION CARESITE CAP FALLING OFF EXTENSION SET DETAILED INQUIRY DESCRIPTION TWO TIMES THIS WEEK, THE END CAP ON THE 8 INCH CARESITE MICRO SET WITH REMOVABLE CARESITE MICRO LUER ACCESS DEVICE AND SPIN LOCK CONNECTOR FELL OFF WHILE DRAWING LABS ON A PATIENT. IF THE END CAP IS REMOVED, THERE IS NO VALVE KEEPING THE BLOOD FLOW IN. WHEN RN WAS DRAWING LABS, THE END CAP OF THE EXTENSION SET FELL OFF CAUSING PATIENT'S BLOOD FLOWS OUT OF THE PIV. BLOOD WENT ON THE PATIENT, AND IT CAUSED A LARGE MESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22208624 ↗MalfunctionMay 1, 2025Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CHEMO LEAKED FROM EXTENSION SET ONTO PATIENT. DETAILED INQUIRY DESCRIPTION: PATIENTS EXTENSION SET SEPARATED FROM THE PIV AND BLOOD AND CHEMO LEAKED ONTO PATIENT. PATIENT HAD DOCETAXEL CHEMOTHERAPY INFUSING WHEN THE CAP ON THE IV CATHETER BECAME DISCONNECTED FROM THE EXTENSION TUBING ON THE PATIENT'S PIV. PATIENT'S PIV THEN BACK FLOWED BLOOD ALL OVER THE PATIENT, THE INFUSION CHAIR, AND THE FLOOR. THE CHEMOTHERAPY THAT WAS INFUSING WHEN THE CAP BECAME DETACHED, LEAKED ONTO THE PATIENT'S SKIN, THE INFUSION CHAIR, AND THE FLOOR. PATIENT PUT ON HER CALL LIGHT WHEN SHE NOTICED HER IV WAS BLEEDING. WHEN WRITER WALKED IN THE ROOM, IMMEDIATELY CLAMPED PATIENT'S PIV TO STOP IT FROM FURTHER BLEEDING AND STOPPED THE PATIENT'S INFUSION PUMP TO STOP FURTHER CHEMOTHERAPY FROM LEAKING ALL OVER. CHEMO SPILL KIT WAS THEN OBTAINED, AND THE CHEMOTHERAPY WAS CLEANED UP PER PROTOCOL. CLEANED UP THE BLOOD ON THE PATIENT, CHAIR, AND FLOOR, AND WASHED PATIENT'S ARM WITH SOAPY WIPES WHERE THE CHEMOTHERAPY CAME INTO CONTACT WITH HER SKIN. OFFERED PATIENT HOSPITAL GOWN AND PANTS TO WEAR. AFTER CHEMO AND BLOOD CLEAN UP. INFUSION WAS RESTARTED WITH
Patient problems: Chemical Exposure
MAUDE report 21941527 ↗MalfunctionApr 25, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: VALVES ON EXTENSION SET ARE "POPPING" OFF OF DURING INFUSIONS OR BLOOD DRAWS. DETAILED INQUIRY DESCRIPTION: ACCOUNT IS REPORTING THAT SOME OF THE VALVES ON THE EXTENSION SETS ARE COMING OFF DURING INFUSIONS OR BLOOD DRAWS. THERE WAS LEAKING OF PATIENT'S OWN BLOOD ONTO THE PATIENT. NO BLOOD WAS FOUND ON THE STAFF MEMBER'S CLOTHING OR SKIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21907704 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page