Skip to content

Perifix® 332097 (UDI 04046955899470)Recalls & reported problems

B. Braun Medical Inc. · PERIFIX® 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Closed Tip Catheter, 5 mL Glass Luer Slip LOR Tray (Kit) · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04046955899470), received between May 5, 2025 and Sep 19, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2025Open, Classified

    Potential for the lid of the catheter connector to be in the incorrect position.

    Z-0119-2026 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Contamination /Decontamination Problem4
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • Needle Stick/Puncture1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Break; Break

    AS REPORTED BY THE USER FACILITY: CUSTOMER COMPLAINT ADVOCATE CALLED THE REPORTER WHILE ENTERING THE RECORD AND HE EXPLAINED THAT THE PLASTIC PORTION OF THE CATHETER TORE OFF WHILE TRYING TO ADMINISTER THE SPINAL TO A LABORING MOM WHO WANTED AN EPIDURAL. HE ADDED THAT THE DEVICE TORE WHILE HE WAS TRYING TO PULL THE CATHETER OUT, THAT THE WIRE WAS STILL IN THE PATIENT. AS THE WIRE WAS STRETCHED OUT AND LENGTHENED AND HAD TWISTED ON ITSELF, IT WAS HARD TO TELL IF A PIECE OF THE WIRE WAS EMBEDDED IN THE PATIENT OR JUST A PIECE OF THE PLASTIC. HE IS NOT SURE IF THE WIRE BROKE OFF OR NOT. HE FEELS LIKE MAYBE HE MADE AN ERROR BY TRYING TO READJUST THE NEEDLE WHILE PULLING THE CATHETER OUT AT THE SAME TIME. THAT MAYBE THE CATHETER NEEDLE BUT THE WIRE OR THE CATHETER - HE IS NOT SURE. AS THE PATIENT WANTED TO EPIDURAL, SHE NEEDED PAIN RELIEF - HE REMOVED EVERYTHING THAT HE COULD AND PLACED ANOTHER CATHETER. A PIECE WAS LEFT IN THE PATIENT. A CT SCAN REVEALED A 1-2CM FRAGMENT EMBEDDED IN THE PATIENT. PATIENT DID DELIVER THE BABY WITHOUT AN ADDITIONAL ISSUE AND IS BEING TREATED AT ANOTHER FACILITY AS THIS IS A SMALL ESTABLISHMENT SO THE PATIENT WAS TRANSFERRED TO ANOTHER LOCATION TO BE TREAT

    Patient problems: Foreign Body In Patient; Needle Stick/Puncture

    MAUDE report 23097381 ↗
  • MalfunctionMay 9, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: RESIDUE IN LIDOCAINE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21997855 ↗
  • MalfunctionMay 9, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: RESIDUE IN LIDOCAINE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21997819 ↗
  • MalfunctionMay 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: RESIDUE IN LIDOCAINE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21959864 ↗
  • MalfunctionMay 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: RESIDUE IN LIDOCAINE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21959811 ↗

Other Epidural anesthesia kit devices

All Epidural anesthesia kit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.