Metha® NC088KRecalls & reported problems
Aesculap AG · METHA NECK 12/14 135°/0° · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046963183745), received between Mar 16, 2021 and Aug 19, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Biocompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Biocompatibility2
Patient problems
- Failure of Implant2
- Metal Related Pathology2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 19, 2021Biocompatibility; Biocompatibility
IT WAS REPORTED TO AESCULAP AG THAT A METHA® SHORT HIP SYSTEM (PART # NC088K) WAS IMPLANTED DURING A PROCEDURE PERFORMED APPROXIMATELY 13 YEARS AND 4 MONTHS AGO. ACCORDING TO THE COMPLAINANT, THE PATIENT EXPERIENCED A LARGE OSTEOLYSIS OF THE LEFT PELVIS WITH STABLE CUP. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE MODULAR METHASHAFT WAS DISCOLOURED BLACK AT THE CONE CONNECTION. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER (B)(4) REFERENCE XC (B)(4). INVOLVED COMPONENTS : NC085T - METHA SHORT HIP STEM CAP SIZE 5 - LOT 51414643. NK560 - BIOLOX PROSTHESIS HEAD 12/14 32MM S - LOT UNKNOWN.
Patient problems: Failure of Implant; Metal Related Pathology
MAUDE report 12343276 ↗InjuryMar 16, 2021Biocompatibility; Biocompatibility
ASSOCIATED MEDWATCH: 9610612-2021-00170 ((B)(4)+ NC081T). INVOLVED COMPONENTS: NH050T - PLASMACUP SC SIZE 50MM - 51484353; NH202 - SC/MSC PE-INSERT 32MM 48/50 SYM. - 51953832; AE-QAS-COMPET-IMP - COLLECT.NO.QAS IMPLANTS FROM COMPETITORS - UNKNOWN.
Patient problems: Failure of Implant; Metal Related Pathology
MAUDE report 11495657 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices
- LFIT 6260-9-136Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented166 reports
- Trident 623-00-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented112 reports
- Accolade 6021-0435Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented88 reports
- EXETER 0580-1-044Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall22 reports
- Endo-Model SL Knee Prosthesis System 16-2840/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall17 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall12 reports
- Trident 623-10-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented57 reports
- LFIT 6260-9-336Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented56 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.