Dialog+® 710200KRecalls & reported problems
B.Braun Avitum AG · Dialog+® Evolution Hemodialysis System · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046963686338), received between Mar 1, 2018 and Mar 19, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (K083460, K963440); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Calibration Problem2
- Defective Alarm1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2025Calibration Problem; Calibration Problem
EVENT 4: ACCORDING TO THE EVENT DESCRIPTION: DURING THE INSTALLATION PROCESS OF 42 DIALOG+ HDF ONLINE (EVO) EQUIPMENT (CODE: 7102072) AND DIALOG+ C/ SUP. CARD. BIC (EVO) (CODE: 710200C) IN THE (B)(6), A SIGNIFICANT DEVIATION WAS IDENTIFIED IN THE CALIBRATION PROCESS OF THE MEASURING INSTRUMENTS. WHEN CARRYING OUT THE TESTS, IT WAS OBSERVED THAT FOR A NOMINAL VALUE OF 14.30 MS/CM CONFIGURED IN DIALOG, THE MEASURING INSTRUMENT (B)(6) (MANUFACTURER: IBP, MODEL: HDM97BQ) PRESENTED READINGS CLOSE TO 16.00 MS/CM, SIGNIFICANTLY ABOVE THE ESTABLISHED TOLERANCE OF ±0.1 MS/CM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21643810 ↗MalfunctionMar 19, 2025Calibration Problem; Calibration Problem
EVENT 7: ACCORDING TO THE EVENT DESCRIPTION: DURING THE INSTALLATION PROCESS OF 42 DIALOG+ HDF ONLINE (EVO) EQUIPMENT (CODE: 7102072) AND DIALOG+ C/ SUP. CARD. BIC (EVO)(CODE:710200C) IN THE (B)(6) HOSPITAL, A SIGNIFICANT DEVIATION WAS IDENTIFIED IN THE CALIBRATION PROCESS OF THE MEASURING INSTRUMENTS. WHEN CARRYING OUT THE TESTS, IT WAS OBSERVED THAT FOR A NOMINAL VALUE OF 14.30 MS/CM CONFIGURED IN DIALOG, THE MEASURING INSTRUMENT EAN0473 (MANUFACTURER: IBP, MODEL: HDM97BQ) PRESENTED READINGS CLOSE TO 16.00 MS/CM, SIGNIFICANTLY ABOVE THE ESTABLISHED TOLERANCE OF ±0.1 MS/CM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21643807 ↗MalfunctionMar 1, 2018Defective Alarm; Defective Alarm
AS REPORTED BY THE USER FACILITY: REPORTED THE MACHINE WAS ALARMING AT THE END OF A THERAPY WITH VENOUS AND ARTERIAL PRESSURE ALARMS AND AIR WAS NOTICED GOING TO THE PATIENT. THE USER FACILITY ALSO STATED THE AIR LOOKED LIKE FOAM. THE CUSTOMER ALSO STATED IT MAY BE BECAUSE OF THE BLOODLINES AND CLAMPS. THE PATIENT WAS SENT TO THE EMERGENCY ROOM AS A PRECAUTION ONLY. SHE WAS NOT EXHIBITING ANY SIGNS OR SYMPTOMS THAT WARRANTED AN EMERGENCY ROOM VISIT. THE PATIENT WAS OKAY AND WAS DISCHARGED THE SAME DAY TO RETURN TO THE FACILITY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7306975 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083460, K963440. One submission often covers many models, so check the product and lot details in each record.
- Z-0872-2022 · Jan 28, 2022 · TerminatedConductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be us
- Z-0873-2022 · Jan 28, 2022 · TerminatedConductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be us
- Z-1532-2016 · Apr 1, 2016 · TerminatedDialog+ Hemodialysis machines: Dialog A + 1 Blood Pump 120V Product Code 710200K Serial # >202169 and < 203249 The Dialog+ can be used for implementing and monitoring hemodialysis
- Z-1536-2016 · Apr 1, 2016 · TerminatedDialog+ Hemodialysis machines: Conductivity Sensor (BIC) Item 3456103A The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or ch
- Z-1535-2016 · Apr 1, 2016 · TerminatedDialog+ Hemodialysis machines: Conductivity Sensor (BIC) Item # 3456102A The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or
- Z-1533-2016 · Apr 1, 2016 · TerminatedDialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL Product Code # 710200S Serial # >200805 and <204001 The Dialog+ can be used for implementing and monitoring hemodial
- Z-1531-2016 · Apr 1, 2016 · TerminatedDialog+ Hemodialysis machines: Dialog A + HE/BIC Product Code 710200L Serial # >200318 and < 2039222 The Dialog+ can be used for implementing and monitoring hemodialysis treatments
- Z-1534-2016 · Apr 1, 2016 · TerminatedDialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL-BIC Product Code # 710200U Serial # 202136 and < 203837 The Dialog+ can be used for implementing and monitoring hemo
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.