Skip to content

Perfusor® 8713040URecalls & reported problems

B. Braun Medical Inc. · PERFUSOR SPACE ACM-KIT · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046963687182), received between May 17, 2018 and Jul 13, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jun 14, 2024Open, Classified

    Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

    Z-2436-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Excess Flow or Over-Infusion3
  • Insufficient Flow or Under Infusion1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2020Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    AS REPORTED BY USER FACILITY: EVENT 2: A FENTANYL INFUSION WAS SET TO RUN AT 2MCG/KG/HR, AND WHEN THE PUMP WAS CHECKED AGAIN, PROBABLY WITHIN AN HOUR, THE RATE WAS 7MCG/KG/HR. NO REPORTED INJURY TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10264702 ↗
  • MalfunctionJul 13, 2020Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    AS REPORTED BY USER FACILITY: EVENT 1: A FENTANYL INFUSION WAS SET TO RUN AT 2MCG/KG/HR, AND WHEN THE PUMP WAS CHECKED AGAIN, PROBABLY WITHIN AN HOUR, THE RATE WAS 7MCG/KG/HR. NO REPORTED INJURY TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10264700 ↗
  • MalfunctionSep 12, 2018Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    AS REPORTED BY USER FACILITY: OVER INFUSION; THE PUMP INFUSED ALL OF THE IRON AT ONCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7868262 ↗
  • MalfunctionMay 17, 2018Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    AS REPORTED BY USER FACILITY: SYRINGE STOPPED INFUSING WITH ADDITIONAL VOLUME TO BE INFUSED. NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7522504 ↗