CADD 21-2111-0300-01Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of CADD (123)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This ICU Medical CADD infusion pump is used to deliver medication. Most of its 1,108 reports describe malfunctions, such as alarm problems, failure to infuse and inaccurate delivery. The reports include no deaths and 2 injuries. Reports fell to 60 in the last 12 months from 268 in the previous 12 months. One open recall, from February 2024, has a software-related root cause.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,116
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,116 adverse event reports that list this exact device (DI 10610586038778), received between Feb 26, 2018 and Sep 29, 2026 — classified as 2 injuries, 1,114 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (166 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K130394); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 27, 2024Class IOpen, Classified
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
Z-2420-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Alarm System166
- Failure to Infuse139
- Device Sensing Problem113
- Inaccurate Delivery112
- Excess Flow or Over-Infusion95
- Peeled/Delaminated73
- Failure to Sense68
- Volume Accuracy Problem67
Patient problems
- No Clinical Signs, Symptoms or Conditions870
- Insufficient Information187
- No Information54
- No Patient Involvement38
- No Consequences Or Impact To Patient34
- No Known Impact Or Consequence To Patient7
- Pain4
- Dysgeusia1
Recent adverse event reports
The 25 most recent of 1,116 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Pumping Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940043 ↗MalfunctionSep 23, 2026Pumping Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50875208 ↗MalfunctionSep 21, 2026Pumping Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50846233 ↗MalfunctionSep 16, 2026Excess Flow or Over-Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50792960 ↗MalfunctionSep 11, 2026Device Sensing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50741085 ↗MalfunctionSep 9, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50699371 ↗MalfunctionSep 8, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50677038 ↗MalfunctionSep 2, 2026Failure to Infuse
It was reported that the pump displayed a System Fault 46,510, with the full fault code reported as 46,510, 303,205,308, 537,652,200, 3,328, 0. This high-priority alarm has the potential to interrupt therapy if it recurs during device use. There was no reported patient involvement, and no patient harm was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50564986 ↗MalfunctionAug 25, 2026Failure to Sense
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50439682 ↗MalfunctionAug 24, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50420093 ↗MalfunctionAug 24, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418577 ↗MalfunctionAug 24, 2026Failure to Infuse
It was reported that during testing the CADD solis HPCA pump had an error code 46508 (UI_ERROR_IRQ_UNDEFINED). There was no patient involvement and harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50412629 ↗MalfunctionAug 13, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50241164 ↗MalfunctionAug 10, 2026Failure to Infuse
It was reported that when putting the cassette, the pump presents the following message: not attached properly or reattach cassette. Pump alarms only when no cassette is attached, and the latch is closed. This is because the fluid ingress causes sensor drift; DSO counts are higher than the software expects with no cassette attached, and therefore the pump alarms ¿Cassette not attached properly¿. DSO might need calibration. There was no patient involvement or harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50192451 ↗MalfunctionAug 4, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50134105 ↗MalfunctionJul 15, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25825958 ↗MalfunctionJul 14, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25818585 ↗MalfunctionJul 9, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25769744 ↗MalfunctionJul 7, 2026Failure to Infuse
THE PUMP GENERATED AN UPSTREAM OCCLUSION ALARM DURING INFUSION. THE CONTENTS WERE REMOVED FROM THE BAG, AND NO TUBING KINKS OR CLOSED CLAMPS WERE FOUND. AN ATTEMPT WAS MADE TO RESTART THE PUMP; HOWEVER, THE DEVICE PROMPTED THE USER TO POWER IT OFF. THE BATTERY DOOR WAS OPENED FOR APPROXIMATELY 10 SECONDS, AND THE PUMP WAS POWERED BACK ON, BUT THE ALARM PERSISTED AND AGAIN INSTRUCTED THE USER TO POWER OFF THE DEVICE. THIS PROCESS WAS REPEATED MULTIPLE TIMES, WITH THE ALARM CONTINUING AND REPEATED REQUESTS TO POWER OFF. SUBSEQUENTLY, THE KEYPAD BECAME LOCKED. THE EVENT OCCURRED IN THE PATIENT'S HOME DURING INFUSION. PATIENT INVOLVEMENT WAS REPORTED; HOWEVER, NO PATIENT HARM RESULTED FROM THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25732984 ↗MalfunctionJul 3, 2026Failure to Infuse
IT WAS REPORTED THAT PUMP HAS FALSE OCCLUSION WARNINGS AND DOWNSTREAM OCCLUSION. THE EVENT OCCURRED DURING SWITCHED OUT THE PUMP, THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25704956 ↗MalfunctionJun 29, 2026Failure to Infuse; Material Integrity Problem
IT WAS REPORTED THAT THE CUSTOMER REPEATEDLY RECEIVED A "CASSETTE NOT LATCHED" MESSAGE DESPITE TRYING NEW CASSETTES. THE DETECTOR PINS WERE OBSERVED TO MOVE FREELY. THE CUSTOMER ALSO NOTED APPARENT CRACKING OF THE BLACK RUBBER COMPONENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT, AND NO HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25655298 ↗MalfunctionJun 9, 2026Failure to Infuse
IT WAS REPORTED THAT THE DEVICE HAD FAILED AIR IN LINE SENSOR ISSUE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25449302 ↗MalfunctionJun 3, 2026Failure to Sense
IT WAS REPORTED THAT THE PUMP RANDOMLY SHUT OFF AND WENT INTO MANUAL MODE DURING TESTING. DURING INVESTIGATION FOUND THE DAMAGED AIR DETECTOR WHICH NEEDED REPLACEMENT. THIS MALFUNCTION MAKES THE EVENT REPORTABLE. THERE ARE NO PATIENT INVOLVEMENT AND ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25382292 ↗MalfunctionMay 27, 2026Material Integrity Problem; Intermittent Communication Failure
IT WAS REPORTED THAT THE CADD PUMP DISPLAYED A COMMUNICATION MODULE ERROR DURING SELF-TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25299717 ↗MalfunctionMay 26, 2026Intermittent Communication Failure
IT WAS REPORTED THAT THE DEVICE DISPLAYED A COMMUNICATION MODULE ERROR DURING SELF-TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25285112 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K130394. One submission often covers many models, so check the product and lot details in each record.
- Z-1667-2025 · Apr 10, 2025 · Open, ClassifiedCADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SO
- Z-1813-2015 · May 26, 2015 · TerminatedCADD- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 & Model 2120, Version 1.0. Provide measured drug therapy to patients in hospita
Other Pump, infusion, pca devices
- Alaris 8120Carefusion 303, Inc. · Pump, infusion, pca2 recalls4 deaths14,640 reports
- CADD 21-2111-0402-51Icu Medical, Inc. · Pump, infusion, pca4 recalls1,535 reports
- Alaris™ PCA Module, Model 8120Carefusion 303, Inc. · Pump, infusion, pca2 recalls1 death917 reports
- Nimbus II PlusInfutronix LLC · Pump, Infusion, Pca1 recall1 death800 reports
- CADD 21-2112-0300-01Icu Medical, Inc. · Pump, infusion, pca1 recall406 reports
- CADD 21-6300-51Icu Medical, Inc. · Pump, infusion, pca414 reports
- Nimbus II PainPROInfutronix LLC · Pump, Infusion, Pca2 recalls186 reports
- CADD 21-2111-0402-51LIcu Medical, Inc. · Pump, infusion, pca3 recalls127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record