Skip to content

Filterflow® 473994Recalls & reported problems

B. Braun Medical Inc. · 1.2 Micron Filtered Extension Set · Class 2

All models of Filterflow (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04046963789695), received between Feb 16, 2018 and Sep 11, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device4
  • Inadequate Filtration Process3
  • Improper Flow or Infusion1
  • Infusion or Flow Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2025Leak/Splash; Air/Gas in Device; Leak/Splash; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: THE DEVICE LEAKED EVEN WHEN CLAMPED IN MULTIPLE PLACES. AFTER BEING ATTACHED TO PATIENT, BLOOD AND AIR WERE PULLED INTO LINE BEFORE FILTER. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23026353 ↗
  • MalfunctionSep 11, 2025Inadequate Filtration Process; Inadequate Filtration Process

    AS REPORTED BY THE USER FACILITY: AIR IS GETTING IN THE SYSTEM DURING PRIMING. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23025890 ↗
  • MalfunctionJul 25, 2025Inadequate Filtration Process; Inadequate Filtration Process

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: SET NOT WORKING PROPERLY. DETAILED INQUIRY DESCRIPTION: (B)(6) LOCATION WAS DOING A TRIAL ON THIS FILTERED EXTENSION. THEY WERE ONLY USING IT TO FILTER OUT BUBBLES IN THE LIPIDS THEY WERE GIVEN THEIR PATIENTS. THE PRODUCT DID NOT WORK PROPERLY. IT DID NOT ELIMINATE ALL THE BUBBLES. THE TARGETED POPULATION IS VERY SMALL, AND A BUBBLE CAN CAUSE HARM. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612215 ↗
  • MalfunctionDec 30, 2024Inadequate Filtration Process; Inadequate Filtration Process

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: THE INLINE 1.2-MICRON IV FILTERS FOR USE WITH A IV FLUID. AFTER USING THE FILTERS, WE NOTICED THAT DESPITE IT BEING A 1.2 U FILTER, THERE WERE PARTICLES OVER 10 U AFTER FILTRATION. THEY DID A PARTICLE ASSESSMENT AND THE FOUND 10-MICRON PARTICLES IN FLUID AFTER FILTRATION DESPITE USING A 1.2 MICRON FILTER. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21036062 ↗
  • MalfunctionOct 11, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED, WE HAVE HAD COMPLAINTS FROM OUR MD'S & RN'S ABOUT THE FILTER EXT SET #473994. THIS IS INTENDED TO KEEP AIR OUT OF THE IV IN FOR INFANTS; IT IS NOT WORKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20428815 ↗
  • MalfunctionFeb 1, 2023Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: CUSTOMER REPORTS AIR BUBBLES ARE FORMING IN THE FILTER WHEN RUNNING AT LOW RATE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16277489 ↗
  • MalfunctionJan 6, 2023Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: CUSTOMER STATES FILTER 473994 FILLED WITH AIR. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16115656 ↗
  • MalfunctionApr 23, 2019Improper Flow or Infusion; Improper Flow or Infusion

    AS REPORTED BY THE USER FACILITY: THE FILTER WAS PRIMED PRIOR TO INITIATING MEDICATION TO ENSURE ALL AIR WAS REMOVED, BUT AN AIR BUBBLE BYPASSED THE FILTER AND WAS TRAVELING TOWARD THE PATIENT. THIS WAS NOTICED DURING USE. NO INJURIES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8539990 ↗
  • MalfunctionFeb 16, 2018Infusion or Flow Problem; Infusion or Flow Problem

    AS REPORTED BY USER FACILITY: THE FILTER FILLED WITH AIR DURING THE MIDDLE OF A FAT EMULSION INFUSION SET TO A RATE OF 0.1 - 2 ML/HR. NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7276310 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page