Skip to content

Caresite® 470108Recalls & reported problems

B. Braun Medical Inc. · CARESITE® Needleless Connector, 6" Bonded Standard Bore Extension Set · Class 2

All models of Caresite (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04046964093753), received between Dec 21, 2016 and Sep 30, 2024 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak8
  • Detachment of Device or Device Component1
  • Improper Flow or Infusion1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient2
  • Blood Loss1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2024Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component

    AS REPORTED, IN ER DEPARTMENT, AFTER DRAWING BLOOD FROM THE IV, VACUTAINER WAS DISCONNECTED FROM IV EXTENSION SET AND BLOOD CONTINUED TO FLOW OUT OF EXTENSION SET CONNECTED TO IV CATHETER. THIS WAS BECAUSE A PIECE OF THE VACUTAINER BROKE OFF IN THE HUB OF THE EXTENSION SET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20337804 ↗
  • MalfunctionJan 19, 2021Fluid/Blood Leak; Fluid/Blood Leak

    EVENT 1: BLOOD LEAKAGE OCCURRED FROM THE EXTENSION SET. AS REPORTED BY THE USER FACILITY: "IT APPEARED THAT THE RUBBER PORTION OF THE CARESITE VALVE BECAME DISLODGED," CAUSING BLOOD LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11196418 ↗
  • MalfunctionJan 19, 2021Fluid/Blood Leak; Fluid/Blood Leak

    EVENT 3: BLOOD LEAKAGE OCCURRED FROM THE EXTENSION SET. AS REPORTED BY THE USER FACILITY: "IT APPEARED THAT THE RUBBER PORTION OF THE CARESITE VALVE BECAME DISLODGED," CAUSING BLOOD LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11196361 ↗
  • MalfunctionJan 19, 2021Fluid/Blood Leak; Fluid/Blood Leak

    EVENT 2: BLOOD LEAKAGE OCCURRED FROM THE EXTENSION SET. AS REPORTED BY THE USER FACILITY: "IT APPEARED THAT THE RUBBER PORTION OF THE CARESITE VALVE BECAME DISLODGED," CAUSING BLOOD LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11196352 ↗
  • MalfunctionDec 16, 2020Fluid/Blood Leak; Fluid/Blood Leak

    EVENT 3: AS REPORTED BY THE USER FACILITY: "WHEN CONNECTING THE TUBING, THERE IS NO TIGHTNESS. BLOOD FLOWS AT THE LEVEL OF THE SCREW."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11020326 ↗
  • MalfunctionDec 16, 2020Fluid/Blood Leak; Fluid/Blood Leak

    EVENT 2: AS REPORTED BY THE USER FACILITY: "WHEN CONNECTING THE TUBING, THERE IS NO TIGHTNESS. BLOOD FLOWS AT THE LEVEL OF THE SCREW."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11020324 ↗
  • MalfunctionDec 16, 2020Fluid/Blood Leak; Fluid/Blood Leak

    EVENT 1: AS REPORTED BY THE USER FACILITY: "WHEN CONNECTING THE TUBING, THERE IS NO TIGHTNESS. BLOOD FLOWS AT THE LEVEL OF THE SCREW."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11020322 ↗
  • MalfunctionFeb 12, 2020Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED BY THE USER FACILITY: DURING THE FIRST 15 MINUTES OF THE BLOOD TRANSFUSION, NO ABNORMALITIES WERE NOTED. HOWEVER, 30 MINUTES LATER, THE PRODUCT LEAKED. BLOOD WAS OBSERVED AT THE ANTI-REFLUX VALVE. THE TRANSFUSION WAS STOPPED, THE DEVICE WAS CHANGED, AND THE TRANSFUSION WAS RESUMED WITHOUT ISSUES.

    Patient problems: Blood Loss; No Known Impact Or Consequence To Patient

    MAUDE report 9699370 ↗
  • MalfunctionFeb 11, 2020Improper Flow or Infusion; Improper Flow or Infusion

    AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTED THAT ADDITIONAL PRIMING WITH SALINE IS DONE DUE TO AIR BACK FLOWING INTO THE TUBING AFTER PRIMING IT INITIALLY. ALSO THE CLAMP IS HARDER TO CLOSE. DETAILS OF THE REPORT: "EVEN AFTER PRIMING WITH MORE SALINE THAN USUAL THE NURSES ARE HAVING TO CLAMP OFF THE TUBING TO PREVENT AIR BACKFLOW INTO THE PATIENTS' VEINS WHICH WASN'T AN ISSUE WITH THE OLD TUBING. NO INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9694005 ↗
  • MalfunctionDec 21, 2016Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED BY THE USER FACILITY: EXTENSION SET TUBE BROKE AT LUER END, DURING BLOOD INFUSION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6196303 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page