Perifix® 332082Recalls & reported problems
B. Braun Medical Inc. · PERIFIX® 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX 20 Ga. Closed Tip SoftTip Catheter, 5 mL Glass Luer Slip LOR Tray (Kit) · Class 2
All models of Perifix (57)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964177286), received between Sep 21, 2017 and Sep 21, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2017Break; Break
PER MEDWATCH REPORT (B)(4) PER MEDWATCH REPORT, " CRNA PLACED THE EPIDURAL CATHETER, REMOVED THE NEEDLE, GAVE A TEST DOSE, TAPED THE CATHETER IN PLACE AND STARTED THE EPIDURAL INFUSION. CRNA HAS SOME DIFFICULTY THREADING THE EPIDURAL CATHETER. PATIENT REPORTED HAVING PAIN, SEEN BY ANESTHESIOLOGIST. NOTED CATHETER DIDN'T APPEAR DEEP ENOUGH IN RELATION TO THE CATHETER MARKINGS, THAT IT SHOULD HAVE BEEN GIVEN AN EPIDURAL SPACE DEPTH OF 7 CM, CATHETER PULLED BACK 1/2 CM, RE-TAPED, PATIENT RE-DOSED. RE-DOSE DID NOT HAVE AN EFFECT ON RELIEVING THE LABOR PAIN SO EPIDURAL WAS REMOVED. ON INSPECTION OF THE EPIDURAL CATHETER, THE BLUE TIP END OF THE CATHETER WAS NOT PRESENT, 1/2 OF THE CATHETER WAS SHEARED OFF. X-RAY OF THE LUMBAR SPINE ON (B)(6) 2017. NO SIGNS OF RETAINED CATHETER, PER NEUROSURGERY CLINICAL NOTE, CATHETER PIECE EITHER HAS COME OUT OR IS TOO SMALL TO SEE AGAINST THE BACKGROUND OF THE BONE."
Patient problems: Foreign Body In Patient
MAUDE report 6882797 ↗
Other Epidural anesthesia kit devices
- Design Options® 552116B. Braun Medical Inc. · Epidural anesthesia kit2 recalls7 reports
- Design Options® 552126B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552179B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552177B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Perifix® 332079B. Braun Medical Inc. · Epidural anesthesia kit1 recall45 reports
- Perifix® 332097B. Braun Medical Inc. · Epidural anesthesia kit1 recall39 reports
- Perifix® 332220B. Braun Medical Inc. · Epidural anesthesia kit1 recall21 reports
- Perifix® 332078B. Braun Medical Inc. · Epidural anesthesia kit1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page