Perifix® 332098 (UDI 04046964177385)Recalls & reported problems
B. Braun Medical Inc. · PERIFIX® 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Open Tip Catheter, 5 mL Glass Luer Slip LOR Tray (Kit) · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964177385), received between Jan 13, 2020 and Jun 15, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
Patient problems
- Foreign Body In Patient2
- Needle Stick/Puncture1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 15, 2025Break; Break
AS REPORTED VIA MEDWATCH NUMBER MW5170263: DESCRIBE EVENT OR PROBLEM: DURING NERVE BLOCK FIRST ATTEMPT, DIFFICULTY FEEDING THE CATHETER. REMOVED AND FOUND THAT THE TIP OF THE CATHETER HAD BROKEN, AND THE SPRING HAD UNWOUND AT THE TIP. ADDITIONALLY, BLUE CM MARKERS RUBBED OFF ON PROVIDER'S FINGERTIPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22208621 ↗MalfunctionJan 29, 2024Break; Break
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CATHETER TIP CAME OFF WHILE REMOVING FROM THE PATIENT. DETAILED INQUIRY DESCRIPTION: TUOHY NEEDLE, RESISTANCE MET INTO 3-4 IN SPACE, BLUE TIP WAS NOT ON CATHETER WHEN PULLED OUT OF THE NEEDLE. RELEVANT TESTS/LABORATORY DATA, INCLUDING DATES ON (B)(6) 2024 - XRAY OF SPINE, ON (B)(6) 2024 - MRI OF SPINE.
Patient problems: Foreign Body In Patient; Needle Stick/Puncture
MAUDE report 18595359 ↗MalfunctionOct 12, 2022Break; Break
AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTS THAT A PATIENT WAS HAVING A BLOOD PATCH PROCEDURE IN THE ER, AND A PIECE OF THE EPIDURAL CATHETER BROKE OFF INTO THE PATIENT. PATIENT DECLINED TO HAVE THE PIECE SURGICALLY REMOVED AND HAS OPTED TO HAVE IT MONITORED. NO INJURY REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 15586271 ↗MalfunctionJan 13, 2020Break; Break
AS PER MEDWATCH NUMBER (B)(4): WHEN CRNA WAS ATTEMPTING TO PUT IN THE CATHETER, SHE FELT A SLIGHT RESISTANCE AND DID NOT CONTINUE THE ADVANCEMENT OF THE CATHETER, AND EVENTUALLY DECIDED TO WITHDRAW THE CATHETER FROM THE PATIENT. IT WAS THEN NOTICED THAT THE BLUE TIP/MARKER AT THE END OF THE CATHETER WAS NOT SEEN (APPROXIMATELY A CENTIMETER) AND EDGES SEEMED FRAYED OR SHEARED OFF. XRAYS WERE TAKEN OF LUMBAR SPINE 2-3 VIEWS, CERVICAL SPINE 2-3 VIEWS AND THORACIC SPINE 2 VIEWS: NO RADIOPAQUE LINEAR DENSITY OR FOREIGN BODY NOTED (IN ALL VIEWS). NO INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9580661 ↗
Other Epidural anesthesia kit devices
- Design Options® 552116B. Braun Medical Inc. · Epidural anesthesia kit2 recalls7 reports
- Design Options® 552177B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552179B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552126B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Perifix® 332079B. Braun Medical Inc. · Epidural anesthesia kit1 recall45 reports
- Perifix® 332097B. Braun Medical Inc. · Epidural anesthesia kit1 recall39 reports
- Perifix® 332220B. Braun Medical Inc. · Epidural anesthesia kit1 recall21 reports
- Perifix® 332078B. Braun Medical Inc. · Epidural anesthesia kit1 recall12 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.