Skip to content

Perifix® 332098 (UDI 04046964177385)Recalls & reported problems

B. Braun Medical Inc. · PERIFIX® 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Open Tip Catheter, 5 mL Glass Luer Slip LOR Tray (Kit) · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964177385), received between Jan 13, 2020 and Jun 15, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4

Patient problems

  • Foreign Body In Patient2
  • Needle Stick/Puncture1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2025Break; Break

    AS REPORTED VIA MEDWATCH NUMBER MW5170263: DESCRIBE EVENT OR PROBLEM: DURING NERVE BLOCK FIRST ATTEMPT, DIFFICULTY FEEDING THE CATHETER. REMOVED AND FOUND THAT THE TIP OF THE CATHETER HAD BROKEN, AND THE SPRING HAD UNWOUND AT THE TIP. ADDITIONALLY, BLUE CM MARKERS RUBBED OFF ON PROVIDER'S FINGERTIPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22208621 ↗
  • MalfunctionJan 29, 2024Break; Break

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CATHETER TIP CAME OFF WHILE REMOVING FROM THE PATIENT. DETAILED INQUIRY DESCRIPTION: TUOHY NEEDLE, RESISTANCE MET INTO 3-4 IN SPACE, BLUE TIP WAS NOT ON CATHETER WHEN PULLED OUT OF THE NEEDLE. RELEVANT TESTS/LABORATORY DATA, INCLUDING DATES ON (B)(6) 2024 - XRAY OF SPINE, ON (B)(6) 2024 - MRI OF SPINE.

    Patient problems: Foreign Body In Patient; Needle Stick/Puncture

    MAUDE report 18595359 ↗
  • MalfunctionOct 12, 2022Break; Break

    AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTS THAT A PATIENT WAS HAVING A BLOOD PATCH PROCEDURE IN THE ER, AND A PIECE OF THE EPIDURAL CATHETER BROKE OFF INTO THE PATIENT. PATIENT DECLINED TO HAVE THE PIECE SURGICALLY REMOVED AND HAS OPTED TO HAVE IT MONITORED. NO INJURY REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 15586271 ↗
  • MalfunctionJan 13, 2020Break; Break

    AS PER MEDWATCH NUMBER (B)(4): WHEN CRNA WAS ATTEMPTING TO PUT IN THE CATHETER, SHE FELT A SLIGHT RESISTANCE AND DID NOT CONTINUE THE ADVANCEMENT OF THE CATHETER, AND EVENTUALLY DECIDED TO WITHDRAW THE CATHETER FROM THE PATIENT. IT WAS THEN NOTICED THAT THE BLUE TIP/MARKER AT THE END OF THE CATHETER WAS NOT SEEN (APPROXIMATELY A CENTIMETER) AND EDGES SEEMED FRAYED OR SHEARED OFF. XRAYS WERE TAKEN OF LUMBAR SPINE 2-3 VIEWS, CERVICAL SPINE 2-3 VIEWS AND THORACIC SPINE 2 VIEWS: NO RADIOPAQUE LINEAR DENSITY OR FOREIGN BODY NOTED (IN ALL VIEWS). NO INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9580661 ↗

Other Epidural anesthesia kit devices

All Epidural anesthesia kit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.