Skip to content

Spinocan® 333211Recalls & reported problems

B. Braun Medical Inc. · Spinocan® 25 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) · Class 2

All models of Spinocan (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964179389), received between Jun 27, 2025 and Jun 27, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 6, 2026Open, Classified

    These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.

    Z-2277-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION. ITEM FOUND IN SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22345397 ↗

Other Spinal anesthesia kit devices

All Spinal anesthesia kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page