Skip to content

PENCAN® 333865Recalls & reported problems

B. Braun Medical Inc. · Pencan® 25 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) - Safety Engineered Tray · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964180804), received between Nov 18, 2019 and Nov 26, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 6, 2026Open, Classified

    These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.

    Z-2276-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Break4

Patient problems

  • Foreign Body In Patient3
  • Discomfort1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2024Break; Break

    ACCORDING TO THE EVENT DESCRITPION: A PATIENT WITH CHRONIC AND COMPLEX ANAL FISTULA FOR SEVERAL YEARS. THE PATIENT WAS OPERATED ON FOR PROCTOLOGICAL EXPLORATION. DURING PUNCTURE FOR SPINAL ANESTHESIA, WHEN THE NEEDLE WAS WITHDRAWN, THE LAST TWO CENTIMETERS OF THE NEEDLE WERE FOUND TO BE MISSING: THE NEEDLE HAD BROKEN OFF, WITH A CLEAN CUT APPROXIMATELY 2 CENTIMETERS (CM) FROM THE TIP. THE MISSING END OF THE NEEDLE REMAINED IN THE PATIENT. SPINAL ANESTHESIA WAS DISCONTINUED AND THE PATIENT SWITCHED TO A MUCH HEAVIER GENERAL ANESTHESIA, WHICH ENTAILED CONSIDERABLE RISK FOR THE PATIENT. FOLLOWING NEUROSURGICAL ADVICE AND A CT SCAN, THE NEEDLE TIP IN THE PATIENT WILL NOT BE REMOVED UNLESS COMPLICATIONS ARISE. INCREASED INFECTIOUS SURVEILLANCE. REPORTEDLY THE BROKEN PART WAS RECOVERED. THE PATIENT'S BMI WAS 25.95 KG/M² AT THE TIME OF SURGERY. OPERATIVE TIME WAS PROLONGED, AND THE PATIENT HAD TO BE RE-OPERATED TO REMOVE THE REMAINING PART OF THE NEEDLE. THE INTRODUCER WAS IN PLACE AT THE TIME OF PUNCTURE.

    Patient problems: Foreign Body In Patient

    MAUDE report 20789728 ↗
  • InjuryNov 1, 2024Break; Break

    ACCORDING TO THE EVENT DESCRIPTION THERE WAS AN ACCIDENTAL AND UNPREDICTABLE BREAKAGE OF A SPINAL NEEDLE WITH RETENTION OF ITS DISTAL END IN THE SUBCUTIS. REPORTEDLY THERE WAS AN INTERVENTION TO REMOVE DISTAL PART IN THE SUBCUTIS.

    Patient problems: Discomfort

    MAUDE report 20591657 ↗
  • InjurySep 13, 2024Break; Break

    ACCORDING TO THE EVENT DESCRIPTION: DURING AN ATTEMPT TO PERFORM SPINAL ANAESTHESIA ON A DIFFICULT PATIENT AFTER NEEDLE EXTRACTION, THE NEEDLE IS FOUND TO BE DAMAGED. THE NEEDLE IS BROKEN APPROXIMATELY 5 CM FROM THE BEVEL. THE PATIENT WAS THEN SENT TO RADIOLOGY WHERE SHE UNDERWENT X-RAY AND CT SCAN WHICH REVEALED THE PRESENCE OF THE STUMP AT THE LEVEL OF THE SPINAL PROCESS. ARRANGEMENTS ARE THEN MADE WITH THE RELEVANT NEUROSURGERY DEPARTMENT AND TRANSFER IS ARRANGED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20211300 ↗
  • MalfunctionNov 18, 2019Break; Break

    AS REPORTED BY THE USER FACILITY: IT WAS REPORTED THAT UPON REMOVAL, THE PENCAN NEEDLE BROKE OFF IN PATIENT'S BACK. DURING A TELEPHONE CONVERSATION THE NURSE MANAGER IT WAS STATED THAT," THE BROKEN NEEDLE TIP WAS NOT RECOVERED. THE PATIENT DID NOT RECEIVE SURGICAL INTERVENTION AT THE TIME OF THE EVENT, OR AS OF CURRENT DATE. IT WAS DECIDED TO SEND THE PATIENT FOR NEURO-SURGERY CONSULT, AND OBSERVE. NO COMPLICATIONS AS OF TODAY."

    Patient problems: Foreign Body In Patient

    MAUDE report 9335006 ↗

Other Spinal anesthesia kit devices

All Spinal anesthesia kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page