Caresite® 470116Recalls & reported problems

B. Braun Medical Inc. · Gravity Burette Set with 1 CARESITE® Injection Site · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046964183973), received between Mar 21, 2017 and Jun 4, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device1
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Non specific EKG/ECG Changes1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: THE DEPARTMENT CLAIMS THAT THE ITEM IS CREATING AIR BUBBLES IN THE LINE. SHANTA SENT ME THE SUB TO LOOK AT WITH DR. ROBINSON FROM THE ANESTHESIA TEAM. UNFORTUNATELY THE SUB #407116 ISN'T ACCEPTABLE. WHEN TESTED WE FOUND THAT THERE'S OPPORTUNITY FOR AIR TO TRAVEL THROUGH THE TUBING ONCE THE BURETROL RUNS DRY. "THE DEPARTMENT CLAIMS THAT THE ITEM IS CREATING AIR BUBBLES IN THE LINE." NO INJURY REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19460990 ↗
  • MalfunctionMar 21, 2017Infusion or Flow Problem; Infusion or Flow Problem

    AS REPORTED BY USER FACILITY: AIR IN LINE DURING USE. AIR WAS INTRODUCED INTO THE PATIENT CAUSING ELEVATED T-WAVE REQUIRED STAY OVERNIGHT AND OBSERVATION.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 6420237 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →