PINNACLE® 2112431Recalls & reported problems

B. Braun Medical Inc. · PINNACLE® High Flow Y-Adapter · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04046964201141), received between Mar 20, 2025 and Mar 27, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 6 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem6

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: PARTICULATES FOUND IN FINAL PRODUCT. WE HAVE BEEN NOTICING THE OCCURRENCE OF PARTICULATE MATTER BEING FOUND AT AQL INSPECTIONS AT OUR CAPS LHV FACILITY AND HAD INQUIRED ABOUT THE AVAILABILITY OF A FILTER TO USE FOR APEX COMPOUNDING. WE WERE INFORMED THAT A FILTER FOR APEX IS NOT AVAILABLE AT THIS TIME. WOULD YOU BE ABLE TO PROVIDE US WITH AN OFFICIAL ASSESSMENT THAT EVALUATES THE POSSIBILITY OF USING A FILTER OR SOME KIND OF DOCUMENTATION AS TO WHY THE USE OF A FILTER IS NOT FEASIBLE AT THIS TIME? THE PARTICLES ARE FOUND INSIDE THE FINISHED FINAL PRODUCT. RECENTLY, SOME OF THE PARTICLES THAT HAVE BEEN SENT FOR IDENTIFICATION TESTING HAVE BEEN IDENTIFIED AS A SYNTHETIC LATEX RUBBER FROM THE DRUG VIAL RUBBER STOPPERS. THIS WOULD INDICATE THE POTENTIAL FOR CORING OF THE RUBBER STOPPERS WHEN THE TECHNICIAN INSERTS THE MINI-SPIKE INTO THE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21706439 ↗
  • MalfunctionMar 27, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION PARTICULATES FOUND IN FINAL PRODUCT. WE HAVE BEEN NOTICING THE OCCURRENCE OF PARTICULATE MATTER BEING FOUND AT AQL INSPECTIONS AT OUR CAPS LHV FACILITY AND HAD INQUIRED ABOUT THE AVAILABILITY OF A FILTER TO USE FOR APEX COMPOUNDING. WE WERE INFORMED THAT A FILTER FOR APEX IS NOT AVAILABLE AT THIS TIME. WOULD YOU BE ABLE TO PROVIDE US WITH AN OFFICIAL ASSESSMENT THAT EVALUATES THE POSSIBILITY OF USING A FILTER OR SOME KIND OF DOCUMENTATION AS TO WHY THE USE OF A FILTER IS NOT FEASIBLE AT THIS TIME? THE PARTICLES ARE FOUND INSIDE THE FINISHED FINAL PRODUCT. RECENTLY, SOME OF THE PARTICLES THAT HAVE BEEN SENT FOR IDENTIFICATION TESTING HAVE BEEN IDENTIFIED AS A SYNTHETIC LATEX RUBBER FROM THE DRUG VIAL RUBBER STOPPERS. THIS WOULD INDICATE THE POTENTIAL FOR CORING OF THE RUBBER STOPPERS WHEN THE TECHNICIAN INSERTS THE MINI-SPIKE INTO THE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21706293 ↗
  • MalfunctionMar 20, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: PARTICULATES FOUND IN FINAL PRODUCT. WE HAVE BEEN NOTICING THE OCCURRENCE OF PARTICULATE MATTER BEING FOUND AT AQL INSPECTIONS AT OUR CAPS LHV FACILITY AND HAD INQUIRED ABOUT THE AVAILABILITY OF A FILTER TO USE FOR APEX COMPOUNDING. WE WERE INFORMED THAT A FILTER FOR APEX IS NOT AVAILABLE AT THIS TIME. WOULD YOU BE ABLE TO PROVIDE US WITH AN OFFICIAL ASSESSMENT THAT EVALUATES THE POSSIBILITY OF USING A FILTER OR SOME KIND OF DOCUMENTATION AS TO WHY THE USE OF A FILTER IS NOT FEASIBLE AT THIS TIME? THE PARTICLES ARE FOUND INSIDE THE FINISHED FINAL PRODUCT. RECENTLY, SOME OF THE PARTICLES THAT HAVE BEEN SENT FOR IDENTIFICATION TESTING HAVE BEEN IDENTIFIED AS A SYNTHETIC LATEX RUBBER FROM THE DRUG VIAL RUBBER STOPPERS. THIS WOULD INDICATE THE POTENTIAL FOR CORING OF THE RUBBER STOPPERS WHEN THE TECHNICIAN INSERTS THE MINI-SPIKE INTO THE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21651218 ↗
  • MalfunctionMar 20, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: PARTICULATES FOUND IN FINAL PRODUCT. WE HAVE BEEN NOTICING THE OCCURRENCE OF PARTICULATE MATTER BEING FOUND AT AQL INSPECTIONS AT OUR CAPS LHV FACILITY AND HAD INQUIRED ABOUT THE AVAILABILITY OF A FILTER TO USE FOR APEX COMPOUNDING. WE WERE INFORMED THAT A FILTER FOR APEX IS NOT AVAILABLE AT THIS TIME. WOULD YOU BE ABLE TO PROVIDE US WITH AN OFFICIAL ASSESSMENT THAT EVALUATES THE POSSIBILITY OF USING A FILTER OR SOME KIND OF DOCUMENTATION AS TO WHY THE USE OF A FILTER IS NOT FEASIBLE AT THIS TIME? THE PARTICLES ARE FOUND INSIDE THE FINISHED FINAL PRODUCT. RECENTLY, SOME OF THE PARTICLES THAT HAVE BEEN SENT FOR IDENTIFICATION TESTING HAVE BEEN IDENTIFIED AS A SYNTHETIC LATEX RUBBER FROM THE DRUG VIAL RUBBER STOPPERS. THIS WOULD INDICATE THE POTENTIAL FOR CORING OF THE RUBBER STOPPERS WHEN THE TECHNICIAN INSERTS THE MINI-SPIKE INTO THE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21651200 ↗
  • MalfunctionMar 20, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPROTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION PARTICULATES FOUND IN FINAL PRODUCT. WE HAVE BEEN NOTICING THE OCCURRENCE OF PARTICULATE MATTER BEING FOUND AT AQL INSPECTIONS AT OUR CAPS LHV FACILITY AND HAD INQUIRED ABOUT THE AVAILABILITY OF A FILTER TO USE FOR APEX COMPOUNDING. WE WERE INFORMED THAT A FILTER FOR APEX IS NOT AVAILABLE AT THIS TIME. WOULD YOU BE ABLE TO PROVIDE US WITH AN OFFICIAL ASSESSMENT THAT EVALUATES THE POSSIBILITY OF USING A FILTER OR SOME KIND OF DOCUMENTATION AS TO WHY THE USE OF A FILTER IS NOT FEASIBLE AT THIS TIME? THE PARTICLES ARE FOUND INSIDE THE FINISHED FINAL PRODUCT. RECENTLY, SOME OF THE PARTICLES THAT HAVE BEEN SENT FOR IDENTIFICATION TESTING HAVE BEEN IDENTIFIED AS A SYNTHETIC LATEX RUBBER FROM THE DRUG VIAL RUBBER STOPPERS. THIS WOULD INDICATE THE POTENTIAL FOR CORING OF THE RUBBER STOPPERS WHEN THE TECHNICIAN INSERTS THE MINI-SPIKE INTO THE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21651172 ↗
  • MalfunctionMar 20, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: PARTICULATES FOUND IN FINAL PRODUCT. WE HAVE BEEN NOTICING THE OCCURRENCE OF PARTICULATE MATTER BEING FOUND AT AQL INSPECTIONS AT OUR CAPS LHV FACILITY AND HAD INQUIRED ABOUT THE AVAILABILITY OF A FILTER TO USE FOR APEX COMPOUNDING. WE WERE INFORMED THAT A FILTER FOR APEX IS NOT AVAILABLE AT THIS TIME. WOULD YOU BE ABLE TO PROVIDE US WITH AN OFFICIAL ASSESSMENT THAT EVALUATES THE POSSIBILITY OF USING A FILTER OR SOME KIND OF DOCUMENTATION AS TO WHY THE USE OF A FILTER IS NOT FEASIBLE AT THIS TIME? THE PARTICLES ARE FOUND INSIDE THE FINISHED FINAL PRODUCT. RECENTLY, SOME OF THE PARTICLES THAT HAVE BEEN SENT FOR IDENTIFICATION TESTING HAVE BEEN IDENTIFIED AS A SYNTHETIC LATEX RUBBER FROM THE DRUG VIAL RUBBER STOPPERS. THIS WOULD INDICATE THE POTENTIAL FOR CORING OF THE RUBBER STOPPERS WHEN THE TECHNICIAN INSERTS THE MINI-SPIKE INTO THE VIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21651099 ↗

Other CONTAINER, I.V. devices

All CONTAINER, I.V. devices →