SafeDAY™ 362006Recalls & reported problems
B. Braun Medical Inc. · 15 DROP UNIVERSAL IV ADMIN SET · Class 2
All models of SafeDAY (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04046964293979), received between Feb 15, 2018 and Feb 15, 2018 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak8
Patient problems
- No Consequences Or Impact To Patient8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 2: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275737 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 1: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275735 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 6: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275733 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 7: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275732 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 8: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275731 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 5: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275729 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 4: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275728 ↗MalfunctionFeb 15, 2018Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 3: CUSTOMER REPORTS, THAT THERE IS AN ONGOING ISSUE WITH THE IV TUBING. WHEN THE TUBING IS FIRST OPENED, THE CONNECTION AT THE STOPCOCK IS ALWAYS LOOSE TO THE POINT WHERE THE TUBING COMES APART DURING PRIMING, UNLESS YOU AGGRESSIVELY TIGHTEN IT AT THE STOPCOCK. THE STAFF WAS EDUCATED TO MAKE SURE IT WAS TIGHTENED PRIOR TO PRIMING TO ADDRESS THE CONCERN. IN THE LAST FEW WEEKS, MULTIPLE INCIDENTS OCCURRED WHERE THE TUBING SEPARATES AFTER THE IV WAS STARTED, RESULTING IN BLOOD BACK FLOWING AND LEAKING OUT OF THE TUBING AND ALL OVER THE FLOOR. THIS OCCURRED EVEN THOUGH THE CONNECTION WAS TIGHTENED AT THE STOPCOCK. NO CHEMO INVOLVED AS THEY ARE AN ENDOSCOPY CENTER GENERAL IV SOLUTION INFUSING. NO INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7275727 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page