Design Options® 552006Recalls & reported problems
B. Braun Medical Inc. · Continuous Epidural Tray · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964311215), received between Jan 22, 2019 and Feb 9, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 18, 2025Open, Classified
Potential for the lid of the catheter connector to be in the incorrect position.
Z-0126-2026 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Break4
Patient problems
- Foreign Body In Patient4
- Needle Stick/Puncture1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 9, 2022Break; Break
AS REPORTED BY THE USER FACILITY: WHILE GIVING LOCAL ANESTHESIA VIA 30 GAUGE 1 INCH NEEDLE THE TIP OF THE NEEDLE BROKE OFF SUBCUTANEOUSLY IN THE PATIENT. PER END USER; THE NEEDLE WAS REMOVED, THERE WAS NO FURTHER INTERVENTION THAT WAS OUTLINED IN THE REPORT. NO INJURY REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 13497675 ↗MalfunctionFeb 4, 2020Break; Break
AS REPORTED BY THE USER FACILITY: COMPLAINT REPORT STATES THAT A 30 GAGE HYPODERMIC NEEDLE BROKE OFF FROM THE HUB IN THE PATIENT. DETAILS OF THE INCIDENT, THE CLINICIAN CLEANED THE PATIENT'S BACK AND ADMINISTERED THE LOCAL ANESTHETIC (SKIN WHEAL). UPON REMOVING THE SYRINGE, SHE NOTICED THAT THE NEEDLE HAD BROKEN OFF THE HUB OF THE SYRINGE AND REMAINED IN THE PATIENT'S BACK.
Patient problems: Foreign Body In Patient
MAUDE report 9665126 ↗MalfunctionSep 17, 2019Break; Break
AS REPORTED BY USER FACILITY: IT WAS REPORTED THAT, "THE INCIDENT OCCURRED DURING CATHETER INSERTION. THE ADVANCEMENT WAS DIFFICULT AND THE PATIENT NEEDED TO BE REPOSITIONED AND MAYBE THE CATHETER GOT PINCHED. WE DID A CT SCAN AND DID NOT VISUALIZE THE CATHETER. IT LOOKED LIKE THE SPRING WOUND PIECE CAME OUT IN ITS ENTIRETY. ANOTHER CATHETER WAS INSERTED IN A DIFFERENT LOCATION."
Patient problems: Foreign Body In Patient; Needle Stick/Puncture
MAUDE report 9075104 ↗MalfunctionJan 22, 2019Break; Break
AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTS, POTENTIAL EPIDURAL CATHETER SHEARING. CRNA WAS PERFORMING A STANDARD EPIDURAL PLACEMENT. CATHETER WAS PLACED AS USUAL. CRNA TRIED TO REPOSITION THE CATHETER AND ENCOUNTERED RESISTANCE. UPON BEING UNABLE TO MANIPULATE THE CATHETER THE CATHETER AND NEEDLE WERE REMOVED. ONCE REMOVED THE CRNA NOTED THAT THE TIP OF THE CATHETER WAS MISSING AND INNER WIRE OF THE CATHETER WAS EXPOSED. THE CATHETER'S INNER WIRE COIL WAS INTACT BUT THE TIP OF THE CATHETER WAS MISSING ABOUT 2CM OF THE SOFT POLYAMIDE. CATHETER WAS REMOVED AND A SECOND CATHETER WAS PLACED. PATIENT WAS DISCHARGED.
Patient problems: Foreign Body In Patient
MAUDE report 8267650 ↗
Other Epidural anesthesia kit devices
- Design Options® 552116B. Braun Medical Inc. · Epidural anesthesia kit2 recalls7 reports
- Design Options® 552177B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552179B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552126B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Perifix® 332079B. Braun Medical Inc. · Epidural anesthesia kit1 recall45 reports
- Perifix® 332097B. Braun Medical Inc. · Epidural anesthesia kit1 recall39 reports
- Perifix® 332220B. Braun Medical Inc. · Epidural anesthesia kit1 recall21 reports
- Perifix® 332078B. Braun Medical Inc. · Epidural anesthesia kit1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page