Skip to content

CP Stent™ 614375Recalls & reported problems

B Braun Interventional Systems, Inc · CMCP026 Mounted Covered CP Stent 4.5 20 · Class 3 · Implant

All models of CP Stent (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964899553), received between Feb 21, 2017 and Feb 21, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult or Delayed Positioning1
  • Migration or Expulsion of Device1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 21, 2017Difficult or Delayed Positioning; Migration or Expulsion of Device

    AS PER THE REPORT FROM BIS: "THE STENT WENT THROUGH THE SHEATH FINE. AFTER UNSHEATHING, IT WAS NOTICED THAT THE STENT WAS MOUNTED DISTALLY BEFORE INFLATION, IT HAD MIGRATED. THE PHYSICIAN WAS ABOUT TO USE THE STENT, IT WAS JUST DIFFICULT TO USE. THE STENT WAS BEING USED FOR A SMALL CONDUIT TEAR. AN INFLATION DEVICE WITH PRESSURE GAUGE WAS USED. THE CATHETER SHAFT WAS NOT KINKED. THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT ANATOMY. THE PATIENT WAS FINE POST PROCEDURE. SALINE AND CONTRAST WERE BOTH USED DURING PREP. THE PHYSICIAN DID INFLATE THE INNER BALLOON BEFORE THE OUTER BALLOON, AND DID NOT ATTEMPT TO PULL THE STENT BACK THROUGH THE HEMOSTASIS VALVE AT ANY TIME."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6347045 ↗

Other Aortic stent devices

All Aortic stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page