Skip to content

EpiSpin Safety NRFit 1022461-30Recalls & reported problems

Pajunk GmbH Medizintechnologie · EpiSpin Safety NRFit Tuohy NRFit 18G x 90mm with back eye and wings · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04048223074800), received between Sep 28, 2026 and Sep 28, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fitting Problem1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Break; Fitting Problem

    IRN# (b)(4)This incident took place in France (Polynésie).An epidural catheter was to be inserted in case additional anesthetic dosingbecame necessary and to provide postoperative epidural analgesia.After completion of the spinal anesthesia, an attempt was made to withdraw the spinal needle (Sprotte) from the epiduralneedle (Tuohy). The metallic portion of the spinal needle remained lodged inside the epidural needle while both werestill in the patient's back, with no possibility of grasping it. The plastic hub of the spinal needle, to which the injectionsyringe is attached and locked, detached and fell to the floor.As there was no longer any external access point to the metallic portion of the spinal needle, it became necessary toremove the epidural needle in order to safely withdraw the spinal needle from the patient's back.Consequently, it was no longer possible to use the epidural needle to thread an epidural catheter and administeradditional anesthetic doses should the surgical procedure be prolonged, nor to provide postoperative epidural analgesia.)

    Patient problems: Foreign Body In Patient

    MAUDE report 50913372 ↗

Other Anesthesia conduction kit devices

All Anesthesia conduction kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page