SPROTTE 041251-29ARecalls & reported problems
Pajunk GmbH Medizintechnologie · Distal Tip: SPROTTE® Proximal End: ISO 80369-7 · Class 2
All models of SPROTTE (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04048223104316), received between Jun 21, 2021 and Jul 2, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Fracture2
Patient problems
- Foreign Body In Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 2, 2025Break; Break
IRN# (B)(4). INCIDENT OCCURRED IN ITALY: WHILE REMOVING THE NEEDLE TO CHANGE POSITION, IT BROKE AND THE STUMP REMAINED INSIDE THE PATIENT, IMPOSSIBLE TO REMOVE WITH BARE HANDS (ABOUT 4.5CM). THEREFORE, THE PATIENT REQUIRED A DIAGNOSTIC-THERAPEUTIC PATHWAY, WHICH ENDED WITH THE REMOVAL OF THE STUMP BY THE NEUROSURGERY COLLEAGUES. NO NEUROLOGICAL DAMAGE WAS REPORTED IN THE PATIENT THROUGHOUT THE PERIPROCEDURAL PATHWAY.
Patient problems: Foreign Body In Patient
MAUDE report 22381699 ↗InjuryJul 26, 2022Fracture; Fracture
IRN-# 191-22_761. EVENT TOOK PLACE IN FRANCE, TENTATIVE TRANSLATION OF INITIAL REPORTER´S NARRATIVE: DURING A LUMBAR PUNCTURE, THE NEEDLE BROKE IN THE PATIENT'S BACK. ON REMOVAL, A PORTION REMAINED IN THE SUBCUTANEOUS TISSUE. USUAL LUMBAR PUNCTURE TECHNIQUE (USE OF INTRODUCTER AND STILET IN THE NEEDLE DURING REMOVAL) OVERWEIGHT PAEDIATRIC PATIENT 137 KG. ADMITTED TO THE NEUROSURGERY UNIT ON 22 JUNE FOR PLACEMENT OF A VENTRICULO-ATRIAL SHUNT AND REMOVAL OF THE NEEDLE FRAGMENT. INFECTION OF THE SHUNT LEADING TO RE-HOSPITALIZATION OF THE PATIENT WITH REMOVAL OF THE SHUNT.
Patient problems: Foreign Body In Patient
MAUDE report 15101393 ↗MalfunctionJun 21, 2021Break; Fracture; Break; Fracture
IRN# (B)(4). TENTATIVE SUMMARIZING TRANSLATION FROM INITIAL REPORTER´S NARRATIVE: "DURING SEVERAL ATTEMPTS TO PLACE THE NEEDLE IN THE SPINAL SPACE AT L2-L3 (AFTER SEVERAL UNSUCCESSFUL ATTEMPTS TO PLACE IT IN L3-L4) THE NEEDLE BROKE AT 3 CM DURING THE REMOVAL PROCEDURE OF THE NEEDLE WITH ITS INTRODUCER. DURING THE PROCEDURE I HAD NEVER CSF FLOW THROUGH THE NEEDLE AND WE HAD NO PARESTESIA. AT THE END THE PATIENT MOVES THE LEGS WITHOUT ANY SENSITIVE OR PHISICAL PROBLEMS. WE CALLED THE SURGEON TO TRY TO SURGICALLY REMOVE THE TIP OF THE NEEDLE WITH LOCAL ANESTHESIA AND SEDATION, AFTER TALKING TO THE PATIENT. FINALLY WE SUCCEDED IN REMOVING THE NEEDLE".
Patient problems: Foreign Body In Patient
MAUDE report 12030751 ↗
Other NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices
- ARROW IPN922632Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)2 recalls0 reports
- ARROW IPN922603Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922604Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- 22G 1 1/2IN Regional Block Needle 408348Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922631Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN923597Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922605Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- BD QUINCKE Spinal Needles 405180Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)23 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page