Skip to content

STIMULONG SONO 531156-35URecalls & reported problems

Pajunk GmbH Medizintechnologie · 531156-35U · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04048223117262), received between Feb 11, 2022 and Feb 11, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2022Material Separation; Material Separation

    IRN# (B)(4). DR. WAS HAVING DIFFICULTY FEEDING THE CATHETER ONCE IT REACHED THE END OF THE TOUCHY NEEDLE AND DECIDED TO ABORT THE PROCEDURE AND PULLED THE CATHETER OUT. WHEN CATHETER WAS REMOVED HE NOTICED THE GOLD TIP OF THE CATHETER DID NOT RETURN AND WAS DISLODGED INSIDE THE PATIENT. HE DESCRIBED THAT HE COULD NOT RETRIEVE IT WITH THE TOUCHY NEEDLE AND WAS CONSULTING WITH THE SURGEON ON HOW TO REMOVE. DR. (B)(6) IS OFF FOR A COUPLE DAYS AND PLANNING ON MEETING WITH HIM NEXT WEEK TO GAIN MORE DETAILS. HE HAD TO SEND THE PRODUCT TO PATH/RISK AND HE IS NOT AWARE AS TO HOW THEY WILL RELEASE IT TO US FOR INVESTIGATION BUT WE WILL COVER WITH THEM IN OUR MEETING NEXT WEEK. ANOTHER PROCEDURE WAS DONE TO REMOVE THE PIECE LEFT BEHIND AND THEY WERE SUCCESSFUL IN REMOVING IT. (DOCTOR) DID NOT DISCLOSE IF THERE WERE ANY COMPLICATIONS.

    Patient problems: Foreign Body In Patient

    MAUDE report 13511667 ↗

Other Anesthesia conduction kit devices

All Anesthesia conduction kit devices →