OsteoBridge™ IDSF Diaphysis Femoral Nail 12 130Recalls & reported problems
Merete GmbH · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048266179623), received between Oct 15, 2020 and May 4, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K181026); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Markings/Labelling Problem1
- Inaccurate Information1
Patient problems
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 4, 2022Inaccurate Information; Inaccurate Information
DURING A SURGERY A MISLABELED OSTEOBRIDGE IDSF FEMORAL NAIL HAS BEEN IDENTIFIED. THE SURGEON WANTED TO IMPLANT A FEMORAL IDSF SYSTEM. AFTER CEMENTED IMPLANTATION OF THE NAILS HE RECOGNIZED THAT THE CLAMPING DIAMETER OF ONE NAIL WAS 14 MM INSTEAD OF 16 MM (FEMORAL IDSF NAILS HAVE 16 MM). THIS LED TOT HE CIRCUMSTANCE THAT THE SPACER WHICH SHALL CONNECT THE NAILS DID NOT FIT. THEREFORE THE SURGEON THEREFORE DECIDED TO USE A TEMPORARY IMPLANT AROUND 14MM CLAMPING AREA OF NAIL TO PROTECT NAIL END. AN IDSF SPACER PROVISIONALLY CLAMPED ONTO CONSTRUCT TO MANAGE LENGTH OF EXTREMITY. PATIENT/ SURGEON HAVE BEEN BE PROVIDED WITH A CUSTOM MADE DEVICE FOR FINAL SURGERY.
Patient problems: Failure of Implant
MAUDE report 14274375 ↗MalfunctionOct 15, 2020Device Markings/Labelling Problem
DURING A SURGERY A MISLABELED OSTEOBRIDGE IDSF FEMORAL NAIL HAS BEEN IDENTIFIED. THE SURGEON WANTED TO IMPLANT A FEMORAL IDSF SYSTEM. AFTER CEMENTED IMPLANTATION OF THE NAILS HE RECOGNIZED THAT THE CLAMPING DIAMETER OF ONE NAIL WAS 14 MM INSTEAD OF 16 MM (FEMORAL IDSF NAILS HAVE 16 MM). THIS LED TOT HE CIRCUMSTANCE THAT THE SPACER WHICH SHALL CONNECT THE NAILS DID NOT FIT. THEREFORE, THE SURGEON THEREFORE DECIDED TO USE A TEMPORARY IMPLANT AROUND 14MM CLAMPING AREA OF NAIL TO PROTECT NAIL END. AN IDSF SPACER PROVISIONALLY CLAMPED ONTO CONSTRUCT TO MANAGE LENGTH OF EXTREMITY. PATIENT/ SURGEON HAVE BEEN BE PROVIDED WITH A CUSTOM MADE DEVICE FOR FINAL SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10682314 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K181026. One submission often covers many models, so check the product and lot details in each record.
- Z-0437-2021 · Sep 18, 2020 · TerminatedOsteoBridge Intramedullary Diaphysis Segment Fixation (IDSF) nails (tibia)
- Z-0436-2021 · Sep 18, 2020 · TerminatedOsteoBridge Intramedullary Diaphysis Segment Fixation (IDSF) nails (femur)
- Z-0438-2021 · Sep 18, 2020 · TerminatedOsteoBridge Intramedullary Diaphysis Segment Fixation (IDSF) nails (humerus)
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page