n.a. 10366SRecalls & reported problems
Karl Storz GmbH & Co. KG · Scissors, straight, 2.5 mm, 55 cm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551046982), received between Apr 11, 2026 and Apr 11, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 11, 2026Material Separation
THE FOLLOWING WAS REPORTED: ON SUNDAY EVENING, DURING AN ESOPHAGEAL FOREIGN BODY OPERATION, AN OTOLARYNGOLOGIST USED AN ESOPHAGOSCOPE FINE SCISSORS TO CUT A FISHBONE IN THE ESOPHAGUS. THE SCISSORS MALFUNCTIONED DURING USE. AFTER SEVERAL UNSUCCESSFUL ATTEMPTS, THE FINE SCISSORS WERE REMOVED FROM THE PATIENT'S BODY AND TESTED ON A SURGICAL DRAPE, WHICH THEY ALSO FAILED TO CUT. THE SCISSORS WERE THEN PLACED ON THE SURGICAL INSTRUMENT CART. AFTER THE OPERATION WAS COMPLETED, THE NURSE FOUND THE INSTRUMENT TO BE ABNORMALLY SHAPED DURING INSTRUMENT RETRIEVAL AND SUSPECTED A DEFECT. SHE INSTRUCTED THE ON-DUTY NURSE TO KEEP THE INSTRUMENT FOR FURTHER VERIFICATION THE NEXT WORKDAY. BY THIS TIME, THE PATIENT HAD BEEN SAFELY RETURNED TO THE WARD. ON MONDAY, THE INSTRUMENT NURSE COMPARED THE ODDLY SHAPED INSTRUMENT WITH PHOTOS TAKEN DURING INSTRUMENT PACKAGING AND CONFIRMED THAT THE DISTAL TIP OF THE SCISSORS HAD BROKEN OFF. THE ON-DUTY SURGICAL NURSE AND OPERATING SURGEON WERE IMMEDIATELY NOTIFIED, THE SURGICAL COURSE WAS REVIEWED, AND THE INCIDENT WAS REPORTED ALTHOUGH THE BROKEN TIP HAS NOT BEEN RECOVERED, RETENTION OF THE FRAGMENT INSIDE THE PATIENT HAS BEEN RULED OUT. DUE TO THE UNKNOWN W
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 24850210 ↗InjuryApr 11, 2026Material Separation
THE FOLLOWING WAS REPORTED: ON SUNDAY EVENING, DURING AN ESOPHAGEAL FOREIGN BODY OPERATION, AN OTOLARYNGOLOGIST USED AN ESOPHAGOSCOPE FINE SCISSORS TO CUT A FISHBONE IN THE ESOPHAGUS. THE SCISSORS MALFUNCTIONED DURING USE. AFTER SEVERAL UNSUCCESSFUL ATTEMPTS, THE FINE SCISSORS WERE REMOVED FROM THE PATIENT'S BODY AND TESTED ON A SURGICAL DRAPE, WHICH THEY ALSO FAILED TO CUT. THE SCISSORS WERE THEN PLACED ON THE SURGICAL INSTRUMENT CART. AFTER THE OPERATION WAS COMPLETED, THE NURSE FOUND THE INSTRUMENT TO BE ABNORMALLY SHAPED DURING INSTRUMENT RETRIEVAL AND SUSPECTED A DEFECT. SHE INSTRUCTED THE ON-DUTY NURSE TO KEEP THE INSTRUMENT FOR FURTHER VERIFICATION THE NEXT WORKDAY. BY THIS TIME, THE PATIENT HAD BEEN SAFELY RETURNED TO THE WARD. ON MONDAY, THE INSTRUMENT NURSE COMPARED THE ODDLY SHAPED INSTRUMENT WITH PHOTOS TAKEN DURING INSTRUMENT PACKAGING AND CONFIRMED THAT THE DISTAL TIP OF THE SCISSORS HAD BROKEN OFF. THE ON-DUTY SURGICAL NURSE AND OPERATING SURGEON WERE IMMEDIATELY NOTIFIED, THE SURGICAL COURSE WAS REVIEWED, AND THE INCIDENT WAS REPORTED ALTHOUGH THE BROKEN TIP HAS NOT BEEN RECOVERED, RETENTION OF THE FRAGMENT INSIDE THE PATIENT HAS BEEN RULED OUT. DUE TO THE UNKNOWN W
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 24850208 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page