n.a. 27071TJRecalls & reported problems
Karl Storz GmbH & Co. KG · Foreign Body Forceps, 28 cm · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04048551079119), received between Jan 13, 2021 and May 5, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Break2
- Fracture1
- Material Integrity Problem1
Patient problems
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMay 5, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT ROLOGY GRASPING FORCEPS BROKE WHILST BEING USED. ATTEMPTED RETRIEVAL OF BROKEN PIECE, INITIALLY GRASPED WITH FORCEPS BUT UNABLE TO FIND THE PIECE WHEN NEW FORCEPS REMOVED FROM SCOPE. PATIENT X-RAYED AND NO METAL SEEN .THOROUGH EXAMINATION WITH REPEAT CYSTOSCOPY DONE AND NO METAL PIECE DETECTED. SURGEONS SATISFIED THAT THE BROKEN PART WAS NOT RETAINED AND CONSULTANT SURGEON INFORMED. THE PROCEDURE WAS COMPLETED WITHOUT INJURY TO PATIENT AND WITH AROUND 15 MINS DELAY. HOWEVER, SINCE THE X RAY WAS REQUIRED, THIS CASE IS REPORTABLE.
Patient problems: Foreign Body In Patient
MAUDE report 25086721 ↗InjuryMay 5, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT ROLOGY GRASPING FORCEPS BROKE WHILST BEING USED. ATTEMPTED RETRIEVAL OF BROKEN PIECE, INITIALLY GRASPED WITH FORCEPS BUT UNABLE TO FIND THE PIECE WHEN NEW FORCEPS REMOVED FROM SCOPE. PATIENT X-RAYED AND NO METAL SEEN. THOROUGH EXAMINATION WITH REPEAT CYSTOSCOPY DONE AND NO METAL PIECE DETECTED. SURGEONS SATISFIED THAT THE BROKEN PART WAS NOT RETAINED AND CONSULTANT SURGEON INFORMED. THE PROCEDURE WAS COMPLETED WITHOUT INJURY TO PATIENT AND WITH AROUND 15 MINS DELAY. HOWEVER, SINCE THE X RAY WAS REQUIRED, THIS CASE IS REPORTABLE.
Patient problems: Foreign Body In Patient
MAUDE report 25086713 ↗MalfunctionApr 2, 2026Break
IT WAS REPORTED THAT THE TECH IN THE ROOM DID TEST PRIOR TO IT GOING IN THE PATIENT AND IT WAS OPENING AND CLOSING FINE, ONCE INSIDE IT BROKE WITH JAWS OPEN SO SHE REMOVED THE HANDLE AND USED ANOTHER GRASPER TO PULL FRAYED ENDS AND THAT CLOSED THE JAWS AND SHE SLOWLY PULLED IT ALL OUT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765022 ↗MalfunctionFeb 16, 2026Fracture
IT WAS REPORTED THAT THE DEVICE IS MECHANICALLY DAMAGED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24354460 ↗MalfunctionApr 18, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE BREAKS EVERY TIME SURGEON TRIES TO USE IT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21869130 ↗MalfunctionJan 13, 2021Material Integrity Problem; Material Integrity Problem
PER MEDWATCH REPORT # (B)(4) RECIEVED, THE GRASPER BROKE DURING FIRST ATTEMPT TO GRASP A URETERAL STENT. THE PIECE OF JAW THAT BROKE OFF WAS RETRIEVED AND REMOVED FROM THE PATIENT.
Patient problems: Foreign Body In Patient; Patient Problem/Medical Problem
MAUDE report 11167610 ↗
FDA clearance and classification
The FDA regulates product code OCZ (endoscopic grasping/cutting instrument, non-powered) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the n.a. 27071TJ
Has the n.a. 27071TJ been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551079119), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the n.a. 27071TJ?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jan 13, 2021 and May 5, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (3), “Break” (2) and “Fracture” (1); the most common patient problems are “Foreign Body In Patient” (3) and “Patient Problem/Medical Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the n.a. 27071TJ?
None of the reports are classified as deaths, 2 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the n.a. 27071TJ, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand n.a., model 27071TJ). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic grasping/cutting instrument, non-powered devices
- diagnostic FTRD Set 13Ovesco Endoscopy AG · Endoscopic Grasping/Cutting Instrument, Non-Powered13 reports
- Single Use Retrieval Basket V FG-V432POlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered11 reports
- Single Use Retrieval Basket V FG-V422PROlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered1 death11 reports
- n.a. 27074BKarl Storz GmbH & Co. KG · Endoscopic grasping/cutting instrument, non-powered9 reports
- n.a. 27072AKarl Storz GmbH & Co. KG · Endoscopic grasping/cutting instrument, non-powered9 reports
- Raptor 00711177US Endoscopy · Endoscopic grasping/cutting instrument, non-powered8 reports
- Raptor 00711178US Endoscopy · Endoscopic grasping/cutting instrument, non-powered5 reports
- Single Use Retrieval Basket V FG-V431POlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page