diagnostic FTRD Set 13Recalls & reported problems
Ovesco Endoscopy AG · The diagnostic FTRD Set is a novel tool for diagnostic and therapeutic full-thickness resection via flexible endoscope in colon and rectum • technology is based on OTSC System • enabling the endoscopist to resect all layers of suitable lesions including the serosa • enables a thorough pathological examination of e.g. motility disorders (e.g. Morbus Hirschsprung) • transection of the tissue happens after the defect has been closed completely • bowel lumen and abdominal cavity won't get into contact • it consists of the following three instruments: diagnostic FTRD (application cap with preloaded clip and integrated HF snare | endoscope sleeve, hand wheel and thread retriever | for colonoscopes) • FTRD Marking Probe (HF coagulation probe | marking of the target lesion in preparation of FTRD use | for detection and complete resection of the right tissue) • FTRD Grasper (specially designed grasping forceps | proper grip on the target tissue | precise retrieval of the sample into the application cap of the FTRD) · Class 2 · Implant
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04260206311072), received between Feb 13, 2019 and Jan 22, 2026 — classified as 13 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem10
- Activation Failure1
- Activation, Positioning or Separation Problem1
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Failure to Advance1
- Leak/Splash1
- Off-Label Use1
Patient problems
- Bowel Perforation8
- Abdominal Distention2
- Abdominal Pain1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Hemorrhage/Blood Loss/Bleeding1
- Internal Organ Perforation1
- Laceration(s) of Esophagus1
- Obstruction/Occlusion1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryJan 22, 2026Use of Device Problem
THE DFTRD WAS USED TO REMOVE A POLYP IN THE CECUM. AFTER TURNING THE HANDWHEEL, THE APPLICATION RING MOVED PAST VISUALIZATION DUE TO A LARGE AMOUNT OF TISSUE IN THE CAP. ASSUMING SUCCESSFUL CLIP APPLICATION, THE RESECTION TOOK PLACE. THE RESULTING PERFORATION WAS CLOSED WITH CLIPS WITHIN THE SAME PROCEDURE. DUE TO SUSPECTED PERITONITIS, THE PATIENT WAS SENT FOR SURGERY.
Patient problems: Bowel Perforation
MAUDE report 24142053 ↗InjuryMay 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE DIAGNOSTIC FTRD WAS USED TO TREAT AN ADENOMA AT THE APPENDICEAL ORIFICE. THE CLIP APPLICATION WAS NOT VERIFIED PRIOR TO TISSUE RESECTION. THE RESULTING PERFORATION WAS CLOSED WITH CLIPS WITHIN THE SAME PROCEDURE.
Patient problems: Bowel Perforation
MAUDE report 22025516 ↗InjuryMay 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE DIAGNOSTIC FTRD WAS USED TO TREAT AN ADENOMA IN THE APPENDICEAL ORIFICE. DUE TO A LARGE AMOUNT OF TISSUE MOBILIZED INTO THE CAP, THE VIEW OF THE WHITE APPLICATION RING WAS LIMITED. AFTER TURNING THE HAND WHEEL THE SUCCESSFUL CLIP APPLICATION WAS NOT VERIFIED PRIOR TO TISSUE RESECTION. THE RESULTING PERFORATION WAS CLOSED WITH A CLIP WITHIN THE SAME PROCEDURE.
Patient problems: Bowel Perforation
MAUDE report 22025339 ↗InjuryMay 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE DIAGNOSTIC FTRD WAS USED TO TREAT A LESION IN THE DUODENUM. CLIP APPLICATION WAS NOT VERIFIED BEFORE TISSUE RESECTION. THE RESULTING PERFORATION WAS CLOSED WITH A CLIP. THIS CLIP HAD TO BE REMOVED DUE TO INCOMPLETE PERFORATION CLOSURE. DURING ENDOSCOPE RETRACTION AND CLIP REMOVAL, THE PATIENT´S ESOPHAGUS WAS ACCIDENTALLY INJURED BY THE CLIP. THE BOWEL PERFORATION WAS CLOSED WITH OVERSTITCHING AND ANOTHER CLIP. THE ESOPHAGEAL TEAR WAS THEN CLOSED WITH ENDOCLIPS. THE PATIENT STAYED IN HOSPITAL OVERNIGHT AND SCANS WERE SCHEDULED FOR THE FOLLOWING DAY.
Patient problems: Bowel Perforation; Laceration(s) of Esophagus
MAUDE report 22024220 ↗InjuryMay 15, 2025Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem
IT WAS REPORTED THAT THE DIAGNOSTIC FTRD WAS USED FOR THE TREATMENT OF A LESION IN THE CARDIA. THE USER MISTAKENLY INTERPRETED THE SLIGHT ONE-SIDED FORWARD MOVEMENT OF THE WHITE APPLICATION RING AS CLIP APPLICATION. THE TISSUE WAS RESECTED SUBSEQUENTLY LEADING TO A PERFORATION. THE PERFORATION WAS CLOSED WITH CLIPS AND THE PATIENT WAS ADMITTED TO THE HOSPITAL.
Patient problems: Internal Organ Perforation
MAUDE report 22024217 ↗InjuryMay 14, 2025Use of Device Problem; Use of Device Problem
DURING AN INTERVENTION IN THE RECTO SIGMOID FOR EFTRD, CLIP DEPLOYMENT WAS MISTAKENLY ASSUMED AND NOT VISUALLY VERIFIED PRIOR TO RESECTION OF THE TARGET TISSUE. THE RESULTING SUPERFICIAL PERFORATION WAS CLOSED WITH 4 MANTIS CLIP. THE PATIENT RECOVERED WITHOUT ANY FURTHER TREATMENT.
Patient problems: Bowel Perforation
MAUDE report 22017651 ↗InjuryMay 14, 2025Use of Device Problem; Use of Device Problem
DURING AN INTERVENTION IN THE CAECUM FOR EFTR, CLIP DEPLOYMENT WAS MISTAKENLY ASSUMED WHILE THE THREAD WAS BROKEN. CORRECT CLIP APPLICATION WAS NOT VISUALLY CONTROLLED, AND THE TARGET TISSUE WAS SUBSEQUENT RESECTED. THE RESULTING PERFORATION WAS CLOSED WITH 7 TTS CLIP. THE PATIENT WAS ADMITTED OVERNIGHT AND DID WELL WITHOUT ANY COMPLICATIONS.
Patient problems: Bowel Perforation
MAUDE report 22017630 ↗InjuryNov 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FEW DAYS AFTER ENDOSCOPIC RESECTION AT THE APPENDICEAL ORIFICE WITH DFTRD, THE PATIENT DEVELOPED APPENDICITIS. THE APPENDICITS REQUIRED SURGICAL APPENDECTOMY.
Patient problems: Abdominal Pain; Unspecified Gastrointestinal Problem
MAUDE report 20693047 ↗InjurySep 13, 2022Use of Device Problem; Use of Device Problem
A DIAGNOSTIC FTRD SET (DFTRD), MANUFACTURER OVESCO ENDOSCOPY AG, WAS USED FOR AN ENDOSCOPIC FULL THICKNESS RESECTION OF A POLYP AT APPENDICAL ORIFICE. THE ENDOSCOPIC INTERVENTIONIST ASSUMED THAT THE CLIP OF THE DIAGNOSTIC FTRD SET IN QUESTION HAD BEEN APPLIED AND, HENCE, PROCEEDED TO PERFORM THE FULL THICKNESS RESECTION. HOWEVER, CLIP APPLICATION WAS VISUALLY NOT VERIFIED BEFOREHAND. THE CAUSED PERFORATION WAS CLOSED BY SURGERY. AS REPORTED THE PATIENT RECOVERED UNEVENTFULLY.
Patient problems: Bowel Perforation
MAUDE report 15410482 ↗InjuryJun 1, 2022Use of Device Problem; Use of Device Problem
A DIAGNOSTIC FTRD SET (DFTRD), MANUFACTURER OVESCO ENDOSCOPY AG, WAS USED FOR AN ENDOSCOPIC FULL THICKNESS RESECTION OF A 2 CM SIZE LESION IN THE CECUM. THE ENDOSCOPIC INTERVENTIONIST ASSUMED THAT THE CLIP OF THE DIAGNOSTIC FTRD SET IN QUESTION HAD BEEN SUCCESSFULLY APPLIED AND, HENCE, PROCEEDED TO PERFORM THE FULL THICKNESS RESECTION. HOWEVER, CLIP APPLICATION WAS VISUALLY NOT VERIFIED BEFOREHAND. THE THEREFORE CAUSED PERFORATION WAS CLOSED BY SURGERY. AS REPORTED THE PATIENT RECOVERED UNEVENTFULLY.
Patient problems: Bowel Perforation
MAUDE report 14559140 ↗InjuryNov 10, 2021Activation Failure
PATIENT WAS IN FOR COLONOSCOPY WITH ENDOSCOPIC ULTRASOUND AND POSSIBLE FULL THICKNESS RESECTION DEVICE (FTRD). LESION FOUND AND ASSESSED BY DOCTORS. THEY FELT IT NEEDED TO BE REMOVED VIA FTRD. AFTER PERFORMING CLIP PLACEMENT AND SNARING OF LESION IT WAS QUICKLY NOTED THAT THERE WAS A SIGMOID COLON PERFORATION. THE SCRUB TECH LOOKED AT THE FTRD SYSTEM THAT WAS REMOVED FROM THE SCOPE AND SAW THAT THE OVESCO CLIP DID NOT DEPLOY. SO, THE CLIP DID NOT GRAB THE TISSUE PRIOR TO SNARING AND CUTTING THE LESION. DOCTOR (WHO HAD PERFORMED FTRD) THEN DECIDED TO STITCH WITH X-TACK SUTURES. PATIENT'S BELLY BECAME SEVERELY DISTENDED AND VERESS NEEDLE HAD TO BE PLACED INTO THE ABDOMEN BY DOCTOR. DOCTOR WAS ABLE TO SUTURE THE PERFORATION, HOWEVER, PATIENT HAD TO BE ADMITTED. FTRD WAS ATTEMPTING TO BE USED TO CLOSE A LARGE DEFECT/HOLE IN THE MUCOSA AFTER A POLYPECTOMY. FOR SOME REASON CLIP THAT CLOSED THE GAP DID NOT ATTACH TO PATIENT AND THE DEFECT/HOLE WAS CLOSED WITH SUTURE INSTEAD. DEVICE RETAINED AND PLACED IN RED BIN. STAFF UNABLE TO TELL IF WAS DEVICE, USER OR ANATOMY .
Patient problems: Abdominal Distention; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Perforation
MAUDE report 12786495 ↗InjuryMar 1, 2019Use of Device Problem; Use of Device Problem
WITHIN THE ENDOSCOPIC INTERVENTION FOR RESECTING A 2 CM FLAT POLYP IN THE SIGMOID COLON, A DIAGNOSTIC FTRD SET (OVESCO ENDOSCOPY USA, INC, (B)(4)) FOR ENDOSCOPIC FULL THICKNESS RESECTION (EFTR) WAS USED. IT WAS DESCRIBED THAT THE FTRD CLIP COULD BE SUCCESSFULLY DEPLOYED; THUS, THE PHYSICIAN PROCEEDED WITH THE FULL THICKNESS RESECTION, AS INTENDED. AFTER FULL THICKNESS RESECTION THE PHYSICIAN NOTICED THAT THE CLIP ADHERED TO THE OPPOSING COLONIC WALL AND ASSUMED IT WOULD CLOSE THE COLON LUMEN. OVESCO'S REMOVE SYSTEM WAS USED TO REMOVE THE FTRD CLIP WITH THE INTENTION TO GAIN MORE LUMINAL SPACE ALTHOUGH THIS WOULD LEAD TO OPENING OF THE BOWEL WALL. TO CLOSE THE RESULTING PERFORATION A OTSC CLIP (14/6T) WAS USED BUT THE CLIP COULD NOT BE APPLIED SINCE THE RELEASE HAND WHEEL DISLODGED. A SECOND OTSC CLIP (14/6T) COULD NOT BE APPLIED EITHER. IT WAS DESCRIBED THAT ONLY THE BOTTOM TEETH DEPLOYED BUT NOT THE TOP SET OF TEETH. A DIFFERENT CLIP TYPE FROM ANOTHER MANUFACTURER WAS THEN USED AND FIRST APPEARED TO CLOSE THE PERFORATION SUCCESSFULLY. BUT CONTRAST AGENT LEAKAGE INTO THE ABDOMINAL CAVITY WAS SEEN UPON FLUOROSCOPY. THUS, THE PATIENT TRANSFERRED TO PACU AND WAS TREATED BY SURGICAL SI
Patient problems: Obstruction/Occlusion
MAUDE report 8381582 ↗InjuryFeb 13, 2019Break; Leak/Splash; Failure to Advance; Activation, Positioning or Separation Problem; Physical Resistance/Sticking
PATIENT PRESENTED TO THE ENDOSCOPY DEPARTMENT TO PLACE A FULL THICKNESS CLIP (FTRD) ON A POLYP IN THE SIGMOID COLON. PER PROCEDURAL NOTE (WHICH PERFECTLY SUMMARIZES THE EVENT: "WHEN THE CLIP WAS DEPLOYED, GI DID PROCEED WITH A FULL THICKNESS RESECTION, AND ATTEMPTED TO PLACE A FULL-THICKNESS CLOSURE DEVICE, UNFORTUNATELY THIS DEVICE ADHERED TO THE OPPOSING COLONIC WALL, THE DEVICE WAS REMOVED, AND A NEW DEVICE WAS PLACED (A PADLOCK), BUT THIS TOO ADHERED TO THE OPPOSING WALL AND WITH INJECTION OF CONTRAST MATERIAL THERE WAS NOTED TO BE A SMALL LEAK FROM THE FULL-THICKNESS RESECTION SITE. BECAUSE OF THIS, THE ENDOSCOPY TEAM DECIDED TO DISCONTINUE THE PROCEDURE, TAKE THE PATIENT OFF ANESTHESIA AND CONSULT GENERAL SURGERY FOR POSSIBLE SIGMOID COLON RESECTION." THE FTRD CLIP (SEE ATTACHED PICTURES) DID NOT CAUSE THE ISSUE-THE AREA WAS DIFFICULT TO ACCESS AND WHEN DEPLOYED, CLOSED THE COLON LUMEN COMPLETELY. THE OVESCO REPS WERE PRESENT AT THE BEDSIDE AND VERBALLY ASSISTED THE PROCESS). THE CLIP WAS SUCCESSFULLY REMOVED USING A OVESCO CLIP CUTTER AND A OVESCO CLIP 14/T WAS DEPLOYED TO CLOSE COLON PERFORATION-IT MALFUNCTIONED. A SECOND 14T WAS DEPLOYED AND AGAIN-MALFUNCTIONED. WRITER SUG
Patient problems: Abdominal Distention; Hemorrhage/Blood Loss/Bleeding
MAUDE report 8334797 ↗
Other Endoscopic Grasping/Cutting Instrument, Non-Powered devices
- Single Use Retrieval Basket V FG-V422PROlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered1 death11 reports
- Single Use Retrieval Basket V FG-V432POlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered11 reports
- n.a. 27074BKarl Storz GmbH & Co. KG · Endoscopic grasping/cutting instrument, non-powered9 reports
- n.a. 27072AKarl Storz GmbH & Co. KG · Endoscopic grasping/cutting instrument, non-powered9 reports
- Raptor 00711177US Endoscopy · Endoscopic grasping/cutting instrument, non-powered8 reports
- n.a. 27071TJKarl Storz GmbH & Co. KG · Endoscopic grasping/cutting instrument, non-powered6 reports
- Raptor 00711178US Endoscopy · Endoscopic grasping/cutting instrument, non-powered5 reports
- NA 0250080265Stryker Corporation · Endoscopic grasping/cutting instrument, non-powered5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.