Skip to content

n.a. 37270DLRecalls & reported problems

Karl Storz GmbH & Co. KG · Suction and Coag.Cannula, l.30 cm · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551155530), received between Nov 12, 2025 and Nov 12, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • Insufficient Information2
  • Unintended Radiation Exposure2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 12, 2025Break; Break

    IT WAS REPORTED THAT THE SURGEON WAS USING THE ELECTRODE PROBE TO PUSH DOWN THE FIBROID. IT WAS THEN THE SURGEON REALIZED THAT THE TIP OF THE INSTRUMENT WAS MISSING. IT WAS AGREED UPON TO SHOOT X-RAY TO DETERMINE WHETHER THE TIP WAS LEFT INSIDE THE PATIENT OR DROPPED OUT. AFTER X-RAY, IT WAS FOUND THAT TIP WAS LEFT INSIDE THE PATIENT. THE PATIENT HAD TO UNDERGO REMOVAL OF THE TIP.

    Patient problems: Insufficient Information; Unintended Radiation Exposure

    MAUDE report 23534502 ↗
  • InjuryNov 12, 2025Break; Break

    IT WAS REPORTED THAT THE SURGEON WAS USING THE ELECTRODE PROBE TO PUSH DOWN THE FIBROID. IT WAS THEN THE SURGEON REALIZED THAT THE TIP OF THE INSTRUMENT WAS MISSING. IT WAS AGREED UPON TO SHOOT X-RAY TO DETERMINE WHETHER THE TIP WAS LEFT INSIDE THE PATIENT OR DROPPED OUT. AFTER X-RAY, IT WAS FOUND THAT TIP WAS LEFT INSIDE THE PATIENT. THE PATIENT HAD TO UNDERGO REMOVAL OF THE TIP.

    Patient problems: Insufficient Information; Unintended Radiation Exposure

    MAUDE report 23534355 ↗

Other COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES) devices

All COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.