P-GRIP 8546Recalls & reported problems
Karl Storz GmbH & Co. KG · Handle Sleeve · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551210130), received between Dec 5, 2023 and Dec 5, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of or Failure to Bond2
- Break1
Patient problems
- Insufficient Information2
- Laceration(s)1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 5, 2023Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THAT DURING A PROCEDURE ON NOV-10-2023, THE DEVICE BROKE AND CUT THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 18271380 ↗InjuryDec 5, 2023Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break
PER NEW INFORMATION PROVIDED BY THE CUSTOMER, THE DEVICE WAS USED FOR INTUBATION ON A MALE PATIENT (76.7 KG, 52 YEARS OLD) ON NOV-10-2023. THE DEVICE BROKE DURING THE INTUBATION AND CUT THE PATIENT'S LIP. INSTANT CARE WAS PROVIDED TO THE PATIENT RIGHT AFTER THIS HAPPENED. THERE IS NO LASTING IMPACT ON THE PATIENT ACCORDING TO THE CUSTOMER. THEY COMPLETED THE PROCEDURE WITH A BACK-UP UNIT, WITH A DELAY OF ABOUT FIVE MINUTES.
Patient problems: Insufficient Information; Laceration(s)
MAUDE report 18271377 ↗
FDA clearance and classification
The FDA regulates product code CCW (laryngoscope, rigid) under 21 CFR 868.5540 as a Class I device. 21 CFR 868.5540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the P-GRIP 8546
Has the P-GRIP 8546 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551210130), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the P-GRIP 8546?
The FDA's MAUDE database lists 2 adverse event reports for this device, received on Dec 5, 2023. In those reports, the most common specific device problems are “Loss of or Failure to Bond” (2) and “Break” (1); the most common patient problem is “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the P-GRIP 8546?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the P-GRIP 8546 FDA approved?
Devices of this type (product code CCW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Karl Storz GmbH & Co. KG.
Who makes the P-GRIP 8546, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand P-GRIP, model 8546). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other LARYNGOSCOPE, RIGID devices
- McGRATH MAC 301-000-000Covidien LP · LARYNGOSCOPE, RIGID2 recalls201 reports
- McGRATH MAC 300-000-000Covidien LP · LARYNGOSCOPE, RIGID1 recall1 death203 reports
- n.a. 8404HXKarl Storz GmbH & Co. KG · LARYNGOSCOPE, RIGID235 reports
- McGRATH 340-000-000Covidien LP · LARYNGOSCOPE, RIGID2 deaths194 reports
- McGRATH MAC 301-000-000Covidien LP · LARYNGOSCOPE, RIGID2 recalls90 reports
- n.a. 8404BXCKarl Storz GmbH & Co. KG · LARYNGOSCOPE, RIGID132 reports
- n/a 8404AXCKarl Storz GmbH & Co. KG · LARYNGOSCOPE, RIGID94 reports
- McGRATH 340-000-000Covidien LP · LARYNGOSCOPE, RIGID84 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page