n.a. 8575KARecalls & reported problems

Karl Storz GmbH & Co. KG · Laryngoscope Holder, GÖTTINGEN model · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Karl Storz Göttingen laryngoscope holder holds a rigid laryngoscope in place during procedures. Reports mostly describe the holder jamming during adjustment and metal shavings or debris falling out, including during surgery. There are 26 reports naming its identifier, with no deaths and 2 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 04048551210901), received between Sep 27, 2024 and Sep 17, 2026 — classified as 2 injuries, 29 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 12 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam22
  • Damaged Thread8
  • Material Fragmentation4
  • Mechanical Problem3
  • Insufficient Device Problem Information2
  • Break1
  • Material Integrity Problem1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions30
  • Insufficient Information3

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Damaged Thread

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812704 ↗
  • MalfunctionSep 17, 2026Damaged Thread

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812703 ↗
  • MalfunctionSep 17, 2026Damaged Thread

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812701 ↗
  • MalfunctionSep 17, 2026Damaged Thread

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812700 ↗
  • MalfunctionSep 17, 2026Damaged Thread

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812699 ↗
  • MalfunctionAug 18, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50304022 ↗
  • MalfunctionApr 24, 2026Mechanical Jam

    IT WAS REPORTED THAT METAL SHAVINGS FROM THE LARYNGOSCOPE HOLDER WERE OBSERVED FALLING OUT DURING SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24987509 ↗
  • MalfunctionApr 24, 2026Mechanical Jam

    IT WAS REPORTED THAT METAL SHAVINGS FROM THE LARYNGOSCOPE HOLDER WERE OBSERVED FALLING OUT DURING SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24986810 ↗
  • MalfunctionApr 3, 2026Mechanical Jam

    IT WAS REPORTED THAT DURING THE OPERATION, METAL DEBRIS WAS OBSERVED WHEN THE ADJUSTMENT WHEEL WAS MOVED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24775908 ↗
  • MalfunctionMar 10, 2026Mechanical Jam

    IT WAS REPORTED THAT THERE WAS OBSERVED A GEAR MALFUNCTION. NO PATIENT INVOLVEMENT REPORTED NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24565649 ↗
  • MalfunctionFeb 2, 2026Break

    IT WAS REPORTED THAT THE LARYNGOSCOPE CANNOT BE SUPPORTED BY THE INSTRUMENT. IT BREAKS AFTER BEING USED ONCE OR TWICE. WHEN RESISTANCE IS APPLIED TO THE CHEST SUPPORT, IT JAMS IMMEDIATELY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24232835 ↗
  • MalfunctionDec 3, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT METAL SHAVINGS FROM THE LARYNGOSCOPE HOLDER WERE OBSERVED FALLING OUT DURING SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23710342 ↗
  • MalfunctionDec 2, 2025Mechanical Jam; Mechanical Jam

    DURING A ROUTINE CHECK, IT WAS REPORTED THAT THE LARYNGOSCOPE HOLDER BECAME STUCK WHILE BEING ADJUSTED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23698772 ↗
  • MalfunctionNov 6, 2025Mechanical Problem; Mechanical Problem

    NO DEFECT DESCRIPTION WAS GIVEN BY THE CUSTOMER. TECHNICAL INSPECTION REVEALED CHIP FORMATION AT THE THREAD OF THE LARYNGOSCOPE HOLDER. ACCORDING TO THE EVENT DESCRIPTION THERE IS NO INFORMATION THAT THE EVENT CAUSED A HARM OR SERIOUS INJURY TO PATIENT, USER OR THIRD. SINCE THE RISK ANALYSIS IS BASED TO PATIENT, USER OR THIRD PARTY HARMS NO RISK-ID IS APPLICABLE TO THE REPORTED EVENT. NON-SERIOUS EVENTS ARE MONITORED THROUGH THE TRENDING PROCESS (Q1.5.0001.WI002).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23490690 ↗
  • MalfunctionAug 29, 2025Material Fragmentation; Mechanical Jam; Material Fragmentation; Mechanical Jam

    ACCORDING TO THE EVENT DESCRIPTION METAL SHARDS INSIDE THERE IS NO INFORMATION THAT THE EVENT CAUSED A HARM OR SERIOUS INJURY TO PATIENT, USER OR THIRD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22922274 ↗
  • MalfunctionAug 29, 2025Material Fragmentation; Mechanical Jam; Material Fragmentation; Mechanical Jam

    ACCORDING TO THE INFORMATION RECEIVED, METAL SHARDS INSIDE. A PROTEIN TEST WAS PERFORMED AND IT SHOWED NEGATIVE, CONFIRMING IT'S COMING FROM THE GEAR WHEEL INSIDE. ACCORDING TO THE AVAILABLE INFORMATION, THERE WAS NO PATIENT INVOLVED. AND THE ISSUE OCCURRED NOT DURING A PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22922264 ↗
  • MalfunctionAug 20, 2025Material Fragmentation; Mechanical Jam; Material Fragmentation; Mechanical Jam

    IT WAS REPORTED THAT THE CHEST SUPPORT ROD WAS MOUNTED THE WRONG WAY AROUND. WHEN ONE TURNS THE WHEEL TO ADJUST THE POSITIONING, METAL SHAVINGS DROP OFF AND ARE LODGED IN THE WHEEL'S MECHANISM. THERE WAS NO HARM TO THE PATIENT OR USER. A DIFFERENT ARTICLE WAS USED TO COMPLETE THE PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. DUE TO AN INTERNAL REVIEW A SENTINEL CASE WAS FOUND.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22853934 ↗
  • MalfunctionAug 18, 2025Material Fragmentation; Mechanical Jam; Material Fragmentation; Mechanical Jam

    IT WAS REPORTED THAT THE DEVICE IS TOO HARD TO RECLINE NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22834960 ↗
  • MalfunctionAug 5, 2025Mechanical Jam; Damaged Thread; Mechanical Jam; Damaged Thread

    IT WAS REPORTED THAT METAL SHAVINGS FELL OUT DURING SURGERY". THE SHAVINGS WERE COMING OUT OF THE METAL WHEEL OF THE ROD AND WERE OUTSIDE OF THE SURGICAL FIELD. ALL OF THE CHIPS (SHAVINGS) COULD BE RECOVERED IN ALL THREE CASES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO SURGICAL DELAY OR PROLONGATION AND NO MEDICAL/ SURGICAL/ DIAGNOSTICAL INTERVENTION REQUIRED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22722171 ↗
  • MalfunctionAug 5, 2025Mechanical Jam; Damaged Thread; Mechanical Jam; Damaged Thread

    IT WAS REPORTED THAT METAL SHAVINGS FELL OUT DURING SURGERY". THE SHAVINGS WERE COMING OUT OF THE METAL WHEEL OF THE ROD AND WERE OUTSIDE OF THE SURGICAL FIELD. ALL OF THE CHIPS (SHAVINGS) COULD BE RECOVERED IN ALL THREE CASES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO SURGICAL DELAY OR PROLONGATION AND NO MEDICAL/ SURGICAL/ DIAGNOSTICAL INTERVENTION REQUIRED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22722108 ↗
  • MalfunctionAug 5, 2025Mechanical Problem; Mechanical Jam; Damaged Thread; Mechanical Problem; Mechanical Jam; Damaged Thread

    IT WAS REPORTED THAT METAL SHAVINGS FELL OUT DURING SURGERY". THE SHAVINGS WERE COMING OUT OF THE METAL WHEEL OF THE ROD AND WERE OUTSIDE OF THE SURGICAL FIELD. ALL OF THE CHIPS (SHAVINGS) COULD BE RECOVERED IN ALL THREE CASES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO SURGICAL DELAY OR PROLONGATION AND NO MEDICAL/ SURGICAL/ DIAGNOSTICAL INTERVENTION REQUIRED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22722080 ↗
  • MalfunctionAug 4, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE DEVICE HAD BEEN WEARING OUT WITH SOME METAL POWDER ON IT, AND THE DEFECT WAS NOTED AFTER REPROCESSING PROCEDURE AND NOT INVOLVING MEDICAL PROCEDURE" THERE IS NO INFORMATION THAT THE EVENT CAUSED A HARM OR SERIOUS INJURY TO PATIENT, USER OR THIRD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22709115 ↗
  • MalfunctionAug 1, 2025Mechanical Jam; Mechanical Jam

    ACCORDING TO THE EVENT DESCRIPTION "NO DEFECT DESCRIPTION" THERE IS NO INFORMATION THAT THE EVENT CAUSED A HARM OR SERIOUS INJURY TO PATIENT, USER OR THIRD. CROSS REFERENCE MDR: 9610617-2025-01329. 9610617-2025-01340.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22691218 ↗
  • MalfunctionAug 1, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT "IT'S ALREADY BEEN WEARING OUT WITH SOME METAL POWDER ON IT, AND THE DEFECT WAS NOTED AFTER REPROCESSING PROCEDURE AND NOT INVOLVING MEDICAL PROCEDURE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22691180 ↗
  • MalfunctionAug 1, 2025Mechanical Jam; Mechanical Jam

    ACCORDING TO THE EVENT DESCRIPTION "NO DEFECT DESCRIPTION" THERE IS NO INFORMATION THAT THE EVENT CAUSED A HARM OR SERIOUS INJURY TO PATIENT, USER OR THIRD. CROSS REFERENCE MDR: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22691146 ↗