n.a. 11272C1Recalls & reported problems
Karl Storz GmbH & Co. KG · Cysto-Urethro-Fiberscope · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 04048551226100), received between Oct 3, 2019 and May 7, 2026 — classified as 6 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Chemical Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 1, 2022Class IIOpen, Classified
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Z-1061-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Chemical Problem3
- Fracture3
- Insufficient Device Problem Information3
- Leak/Splash3
- Material Integrity Problem3
- Microbial Contamination of Device3
- Calcified2
- Use of Device Problem2
Patient problems
- Insufficient Information9
- Urinary Tract Infection5
- No Clinical Signs, Symptoms or Conditions3
- Bowel Perforation1
- Foreign Body In Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryMay 7, 2026Use of Device Problem
THE FOLLOWING WAS REPORTED: THE CUSTOMER HAS A FAULTY 11272C1 (LEAKING) CYSTO- URETHRO-FIBERSCOPE. APPARENTLY, TWO PATIENTS DEVELOPED A BLADDER INFECTION, WHICH LED TO THE DISCOVERY THAT THE ENDOSCOPE WAS LEAKING. DUE TO DEVELOPED BLADDER INFECTION THE EVENT IS DEEMED REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25117935 ↗InjuryMay 6, 2026Use of Device Problem
THE FOLLOWING WAS REPORTED: THE CUSTOMER HAS A FAULTY 11272C1 (LEAKING) CYSTO-URETHRO-FIBERSCOPE. APPARENTLY, TWO PATIENTS DEVELOPED A BLADDER INFECTION, WHICH LED TO THE DISCOVERY THAT THE ENDOSCOPE WAS LEAKING. DUE TO DEVELOPED BLADDER INFECTION THE EVENT IS DEEMED REPORTABLE.
Patient problems: Urinary Tract Infection
MAUDE report 25107407 ↗InjuryMar 23, 2026Microbial Contamination of Device
ACCORDING TO THE CUSTOMER, SEVERAL PATIENTS REPORTED URINARY TRACT INFECTIONS AFTER TREATMENT. DESPITE PROPER DISINFECTION AND CORRECT REPROCESSING, THE CUSTOMER NOTED THAT THE DEVICE REPEATEDLY SHOWED MICROBIOLOGICAL ABNORMALITIES WITH EVIDENCE OF BACTERIAL CONTAMINATION. SINCE A DEVICE-RELATED INFECTION CANNOT BE EXCLUDED AND THE ISSUE HAS OCCURRED REPEATEDLY, THE EVENT IS DEEMED REPORTABLE. SINCE SEVERAL EVENTS WERE REPORTED, THE FOLLOWING THREE CASES WERE LINKED TO ONE ANOTHER: 20260004292; 20260004295; 20260004299.
Patient problems: Urinary Tract Infection
MAUDE report 24670877 ↗MalfunctionNov 14, 2025Fracture; Chemical Problem; Fracture; Chemical Problem
IT WAS REPORTED THE SHEATH OF THE DEVICE BROKE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 23553697 ↗MalfunctionNov 5, 2025Fracture; Chemical Problem; Fracture; Chemical Problem
IT WAS REPORTED THE SHEATH OF THE DEVICE BROKE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 23476375 ↗MalfunctionAug 26, 2025Optical Problem; Device Difficult to Maintain; Optical Problem; Device Difficult to Maintain
IT WAS REPORTED THE IMAGE QUALITY OF THE FLEXIBLE SCOPE IS NOT CLEAR DURING SURGERY. IT IS NOT A BRAND NEW SCOPE AND IMAGE QUALITY IS BAD. DIFFICULTY IN DOING THE CYSTOSCOPY PROCEDURE. PROCEDURE COMPLETED SUCCESSFULLY. SPARE DEVICE WAS USED TO COMPLETE PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22900893 ↗InjuryAug 22, 2025Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE FLEX IS LEAKING. BLADDER INJURY, UNNOTICED EMPTYING OF THE BLADDER.
Patient problems: Bowel Perforation
MAUDE report 22873171 ↗MalfunctionAug 5, 2025Fracture; Chemical Problem; Fracture; Chemical Problem
IT WAS REPORTED THAT THE DEVICE SHOWS A BROKEN SHEATH. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 22717911 ↗InjuryDec 21, 2023Calcified; Material Invagination; Microbial Contamination of Device; Calcified; Material Invagination; Microbial Contamination of Device
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE FRONT COVER OF THE PRODUCT (DEFECTIVE). INFECTIONS OCCURRED SEVERAL TIMES AFTER ENDOSCOPY. DUE TO THE REPORTED POSSIBLE INFECTIONS (CAUSALITY NOT CONFIRMED) THE RISK FOR SERIOUS DETERIORATION IN STATE OF HEALTH IS NOT NEGLIGIBLE.
Patient problems: Insufficient Information
MAUDE report 18378987 ↗InjuryDec 21, 2023Calcified; Calcified
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE FRONT COVER OF THE PRODUCT (DEFECTIVE). INFECTIONS OCCURRED SEVERAL TIMES AFTER ENDOSCOPY. DUE TO THE REPORTED POSSIBLE INFECTIONS (CAUSALITY NOT CONFIRMED) THE RISK FOR SERIOUS DETERIORATION IN STATE OF HEALTH IS NOT NEGLIGIBLE.
Patient problems: Insufficient Information
MAUDE report 18375032 ↗MalfunctionMay 10, 2023Contamination; Microbial Contamination of Device; Contamination; Microbial Contamination of Device
THE EVENT OCCURRED IN GERMANY. THE CUSTOMER REPORTS A POSSIBLE INCIDENT DUE TO GERMS IN THE INSTRUMENT. NO DETAILS REGARDING MALFUNCTION, PATIENT INVOLVEMENT, OR NEGATIVE IMPACT ON THE STATE OF HEALTH WERE REPORTED. FURTHER INFORMATION WAS REQUESTED BUT NOT RECEIVED AT THIS TIME. THUS, IN TERMS OF CAUTION, THIS CASE IS DEEMED REPORTABLE.
Patient problems: Insufficient Information
MAUDE report 16904841 ↗MalfunctionMar 23, 2023Fire; Fire
ADDITIONAL INFORMATION WAS PROVIDED THAT SMOKE WAS SEEN AT THE TOP OF THE WORKING SHAFT DURING ENDOSCOPIC SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16599562 ↗MalfunctionApr 29, 2021Leak/Splash; Material Integrity Problem; Insufficient Device Problem Information; Leak/Splash; Material Integrity Problem; Insufficient Device Problem Information
AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6): FROM (B)(6) 2021, HOSPITALIZATION OF 3 PATIENTS FOR KLEBSIALLA PNEUMONIAE EBSLE URINARY TRACT INFECTION. THIS REPORT IS FOR HOSPITALIZATION OF PATIENT 2 OF 3.
Patient problems: Insufficient Information; Urinary Tract Infection
MAUDE report 11744780 ↗MalfunctionApr 29, 2021Leak/Splash; Material Integrity Problem; Insufficient Device Problem Information; Leak/Splash; Material Integrity Problem; Insufficient Device Problem Information
AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(4): FROM (B)(6) 2021, HOSPITALIZATION OF 3 PATIENTS FOR KLEBSIALLA PNEUMONIAE EBSLE URINARY TRACT INFECTION. THIS REPORT IS FOR HOSPITALIZATION OF PATIENT #1 OF 3.
Patient problems: Insufficient Information; Urinary Tract Infection
MAUDE report 11744779 ↗MalfunctionApr 29, 2021Leak/Splash; Material Integrity Problem; Insufficient Device Problem Information; Leak/Splash; Material Integrity Problem; Insufficient Device Problem Information
AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6): FROM (B)(6) 2021, HOSPITALIZATION OF 3 PATIENTS FOR KLEBSIALLA PNEUMONIAE EBSLE URINARY TRACT INFECTION. THIS REPORT IS FOR HOSPITALIZATION OF PATIENT 3 OF 3.
Patient problems: Insufficient Information; Urinary Tract Infection
MAUDE report 11744778 ↗MalfunctionOct 3, 2019Material Fragmentation
PER THE CUSTOMER, DURING A ELECTIVE LEFT PERCUTANEOUS NEPHROLITHOTOMY, LEFT NEPHROSTOMY TUBE INSERTION, AND LEFT NEPHROSTOGRAM PROCEDURE, THE LEFT RENAL CALCULUS WAS EXTRACTED SUCCESSFULLY UTILIZING THE FLEXIBLE NEPHROSCOPE, AND THE LEFT NEPHROSTOGRAM DEMONSTRATES NO EXTRAVASATION. SURGEON OBSERVED BLACK PARTICLE DURING THE PROCEDURE, REALIZED THAT IT CAME FROM THE END OF THE FLEXIBLE CYSTOSCOPE. THEY WERE ABLE TO RETRIEVE IT AND THE PIECE MATCHED THE END OF THE FLEXIBLE CYSTOSCOPE. INSTRUMENT HAD BROKEN PIECES OF COATING ON THE TIP OF THE SCOPE. THE SCRUB TECH, THE SURGEON AND THE RN IN THE CASE EXAMINED THE SCOPE AND DECIDED THAT ALL PIECES HAD BEEN RETRIEVED. THE PATIENT TOLERATED THE PROCEDURE, AND WAS THEN TRANSFERRED TO A POST-ANESTHESIA CARE UNIT. PATIENT WAS STABLE AND WAS THEN DISCHARGED HOME.
Patient problems: Foreign Body In Patient; Patient Problem/Medical Problem
MAUDE report 9152668 ↗
FDA clearance and classification
The FDA regulates product code FBO (cystourethroscope) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the n.a. 11272C1
Has the n.a. 11272C1 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1061-2022 (Class II), began on Apr 1, 2022: Failure to achieve the expected six-log reduction in microorganisms following the disinfection process. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the n.a. 11272C1?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Oct 3, 2019 and May 7, 2026. In those reports, the most common specific device problems are “Chemical Problem” (3), “Fracture” (3) and “Leak/Splash” (3); the most common patient problems are “Urinary Tract Infection” (5), “Bowel Perforation” (1) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the n.a. 11272C1?
None of the reports are classified as deaths, 6 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the n.a. 11272C1, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand n.a., model 11272C1). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CYSTOURETHROSCOPE devices
- n.a. 27050CAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE54 reports
- n.a. 11272VNUKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall4 reports
- C-VIEW 11272VUAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VUKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- n.a. 11272CU1Karl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- n.a. 11272VNKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page