n.a. 11272VNURecalls & reported problems
Karl Storz GmbH & Co. KG · Flexible Video Urethro-Cystoscope · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04048551226193), received between Apr 10, 2018 and Apr 10, 2018 — classified as 4 injuries.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 1, 2022Open, Classified
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Z-1064-2022 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Contamination /Decontamination Problem4
Patient problems
- Bacterial Infection4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 10, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem
ALLEGEDLY, THE PHYSICIAN PERFORMED A CYSTOSCOPY PROCEDURE ON PATIENT. POST THE PROCEDURE, PATIENT WAS ADMITTED TO THE ICU. PATIENT'S URINE WAS CULTURED WHICH CONFIRMED PSEUDOMONAS AERUGINOSA INFECTION. PATIENT WAS TREATED BY AN OUTSIDE DOCTOR. THE CLINIC DOES NOT HAVE ANY INFORMATION ON TREATMENT, BUT MOST LIKELY THE PATIENT RECEIVED ANTIBIOTICS.
Patient problems: Bacterial Infection
MAUDE report 7418170 ↗InjuryApr 10, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem
ALLEGEDLY, THE PHYSICIAN PERFORMED A CYSTOSCOPY PROCEDURE ON PATIENT. PATIENT'S URINE WAS CULTURED AND PSEUDOMONAS AERUGINOSA INFECTION WAS CONFIRMED FROM THE LAB RESULTS. PATIENT WAS TREATED WITH ANTIBIOTICS FROM A DOCTOR OUTSIDE OF THE CLINIC. THE DOCTOR REPORTED THAT PATIENT IS DOING FINE.
Patient problems: Bacterial Infection
MAUDE report 7418124 ↗InjuryApr 10, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem
ALLEGEDLY, THE PHYSICIAN PERFORMED A CYSTOSCOPY PROCEDURE ON PATIENT. PATIENT'S URINE WAS CULTURED AND PSEUDOMONAS AERUGINOSA INFECTION WAS CONFIRMED FROM THE LAB RESULTS. PATIENT WAS TREATED WITH ANTIBIOTICS FROM A DOCTOR OUTSIDE OF THE CLINIC. THE DOCTOR REPORTED THAT PATIENT IS DOING FINE.
Patient problems: Bacterial Infection
MAUDE report 7418116 ↗InjuryApr 10, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem
ALLEGEDLY, THE PHYSICIAN PERFORMED A CYSTOSCOPY PROCEDURE ON PATIENT. PATIENT'S URINE WAS CULTURED AND PSEUDOMONAS AERUGINOSA INFECTION WAS CONFIRMED FROM THE LAB RESULTS. PATIENT WAS TREATED WITH ANTIBIOTICS FROM A DOCTOR OUTSIDE OF THE CLINIC. THE DOCTOR REPORTED THAT PATIENT IS DOING FINE.
Patient problems: Bacterial Infection
MAUDE report 7418096 ↗
Other CYSTOURETHROSCOPE devices
- n.a. 11272C1Karl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall16 reports
- n.a. 27050CAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE54 reports
- C-VIEW 11272VUKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- n.a. 11272CU1Karl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VUAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- n.a. 11272VNKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.