n.a. TL300Recalls & reported problems
Karl Storz GmbH & Co. KG · Cold Light Fountain Power LED 300 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04048551378830), received between Jun 29, 2022 and Jun 8, 2026 — classified as 5 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Use of Device Problem2
- Appropriate Device Problem Term/Code Not Available1
- Display or Visual Feedback Problem1
- Electrical /Electronic Property Problem1
- Excessive Heating1
- Fire1
- Intermittent Communication Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Burn(s)3
- Superficial (First Degree) Burn1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SURGERY, A FROZEN IMAGE ERROR WAS TRIGGERED WHILE USING THE 30-DEGREE ENDOSCOPE. LATER IN THE PROCEDURE, THE ISSUE OCCURRED AGAIN, AND A HORIZONTAL LINE APPEARED ACROSS THE MIDDLE OF THE SCREEN AND THE LIGHT SOURCE WAS LOST. THE SURGICAL TEAM SWITCHED TO A 0-DEGREE ENDOSCOPE BECAUSE IT WAS MORE ADVANTAGEOUS TO THE PROCEDURE, WHICH WORKED WITHOUT ANY PROBLEMS. AFTER ATTEMPTING TO SWITCH BACK TO THE 30-DEGREE ENDOSCOPE, THERE WAS NO IMAGE DISPLAYED AT ALL, SO THE SURGICAL TEAM CONTINUED USING THE 0-DEGREE ENDOSCOPE. THE MONITORS WERE STILL SHOWING THAT THE 30-DEGREE ENDOSCOPE WAS CONNECTED EVEN AFTER IT HAD BEEN SWITCHED WITH THE 0-DEGREE ENDOSCOPE. THERE WAS NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25435312 ↗MalfunctionDec 30, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING A ROBOTIC LAPAROSCOPIC RIGHT NEPHRECTOMY DURING THE SUPERIOR ASPECT OF THE RIGHT KIDNEY MOBILIZATION, ALL SCREENS BLACKED OUT ON THE TOWER AND SURGEON CONSOLE. WHEN CHECKING THE ENDOSCOPE SYSTEM, THE LIGHT SOURCE WAS FLASHING AND THERE WAS A BURNING SMELL. THE OPERATING ROOM TEAM REMOVED ALL INSTRUMENTS, UNDOCKED THE SYSTEM, AND MOVED THE ROBOTIC ARMS AWAY FROM THE PATIENT TO SAFELY SHUT DOWN THE SYSTEM. WHEN THE SYSTEM WAS RESTARTED, THE ISSUE RECURRED ALONG WITH THE BURNING SMELL. THE SURGEON CONVERTED THE SURGERY TO A LAPAROSCOPIC SURGERY TO AVOID FURTHER INCIDENT. IT WAS FOUND THE KARL STORZ HARDWARE WAS SMOKING INSIDE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23931022 ↗MalfunctionDec 5, 2025Loose or Intermittent Connection; No Apparent Adverse Event; Loose or Intermittent Connection; No Apparent Adverse Event
IT WAS REPORTED THAT THE "THE TOWER TRIGGERED AN ENDOSCOPE ERROR". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2025-02249. 9610617-2025-02250.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23730162 ↗MalfunctionOct 15, 2025Fire; Excessive Heating; Fire; Excessive Heating
IT WAS REPORTED THAT DURING SURGERY, WHEN THE OBJECTIVE END OF THE ENDOSCOPE WAS PLACED STATIONARY ON THE PATIENT'S BODY, THE HEAT RAPIDLY BURNED THROUGH THE NON-WOVEN FABRIC. THE STANDBY BUTTON OF THE COLD LIGHT SOURCE HAD NOT BEEN PRESSED AT THE TIME. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23280778 ↗InjuryMar 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE INFORMATION RECEIVED, THE USER TEMPORARILY PLACED THE GUIDE BEAM AND LENS ON THE PATIENT DURING THE OPERATION. AFTER THE OPERATION, IT WAS FOUND THAT THE PATIENT WAS SLIGHTLY SCALDED ON THE SKIN WHERE IT HAD BEEN PLACED. AFTER THE SCALDING OBSERVATION FOR A PERIOD OF TIME, NO OTHER ABNORMALITIES WERE FOUND, AND NO ADDITIONAL INTERVENTION WAS DONE.
Patient problems: Burn(s)
MAUDE report 21537905 ↗InjuryMar 5, 2025Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE INFORMATION RECEIVED, THE USER TEMPORARILY PLACED THE GUIDE BEAM AND LENS ON THE PATIENT DURING THE OPERATION. AFTER THE OPERATION, IT WAS FOUND THAT THE PATIENT WAS SLIGHTLY SCALDED ON THE SKIN WHERE IT HAD BEEN PLACED. AFTER THE SCALDING OBSERVATION FOR A PERIOD OF TIME, NO OTHER ABNORMALITIES WERE FOUND, AND NO ADDITIONAL INTERVENTION WAS DONE.
Patient problems: Burn(s)
MAUDE report 21531987 ↗MalfunctionNov 22, 2024Intermittent Communication Failure; Intermittent Communication Failure
IT WAS REPORTED THAT DURING THE PROCEDURE THE PICTURE OF THE CAMERA IS GONE. AFTER A SHORT PERIOD, THE PICTURE IS BACK. NO HARM TO PATIENT, USER OR THIRD REPORTED. NO NEGATIVE CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20760012 ↗InjuryOct 15, 2024Optical Problem; Optical Problem
IT WAS REPORTED THAT PRE-OPERATIVELY, THE STORZ SYSTEM FAILED TO PERFORM A WHITE BALANCE. THE TROUBLESHOOTING INVOLVED RESTARTING THE STORZ SYSTEM THREE TIMES, CHANGING THREE ENDOSCOPES, DISCONNECTING THE EXTENSION CABLE, AND CONNECTING THE ENDOSCOPES WITHOUT THE EXTENSION CABLE. DESPITE THESE EFFORTS, THE ISSUE PERSISTED. A SYSTEM REBOOT WAS ALSO ATTEMPTED, BUT THE ISSUE REMAINED UNRESOLVED. AS A RESULT, THE ROBOTIC SURGERY WAS CONVERTED TO LAPAROSCOPIC SURGERY, AND THE NEXT SCHEDULED OPERATION WAS CANCELLED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20453807 ↗InjuryAug 26, 2022Use of Device Problem; Use of Device Problem
CUSTOMER REPORTED THAT ON OR ABOUT (B)(6) 2022, A PATIENT BURN OCCURRED AS A RESULT OF THE LIGHT SOURCE NOT BEING TURNED OFF OR ON STANDBY WHILE CAMERA (SCOPE, LIGHT CABLE) WAS LAYING ON THE PATIENT OR PATIENT DRAPE DURING A CYSTO CASE. TO DATE, WE HAVE BEEN UNABLE TO OBTAIN ANY FURTHER DETAILS ON THIS CASE.
Patient problems: Burn(s)
MAUDE report 15300713 ↗InjuryJun 29, 2022Use of Device Problem; Use of Device Problem
AS PER MANUFACTURER INCIDENT REPORT #(B)(4) RECEIVED FROM THE FACTORY IN (B)(6): THE REPORTED EVENT HAPPENED IN (B)(6). TOWARDS THE END OF A THORAX SURGERY, THE SURGEON PLACED THE TELESCOPE NEAR THE PATIENT'S FACE WHILE THE LIGHT SOURCE WAS ON. AFTER A SHORT TIME, THE ANESTHETIST NOTICED THAT THERE WAS A SLIGHT BURN IN THE SHEET. THIS CAUSED A SLIGHT BURN NEAR THE PATIENT'S CHIN. THE PATIENT WAS TREATED WITH AN APPROPRIATE OINTMENT.
Patient problems: Superficial (First Degree) Burn
MAUDE report 14852572 ↗
FDA clearance and classification
The FDA regulates product code NTN (led light source) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals
Common questions about the n.a. TL300
Has the n.a. TL300 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551378830), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the n.a. TL300?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jun 29, 2022 and Jun 8, 2026. In those reports, the most common specific device problems are “Use of Device Problem” (2), “Display or Visual Feedback Problem” (1) and “Electrical /Electronic Property Problem” (1); the most common patient problems are “Burn(s)” (3) and “Superficial (First Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the n.a. TL300?
None of the reports are classified as deaths, 5 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the n.a. TL300 FDA approved?
Devices of this type (product code NTN) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Karl Storz GmbH & Co. KG.
Who makes the n.a. TL300, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand n.a., model TL300). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Led light source devices
- VIDEO SYSTEM CENTER CV-170Olympus Medical Systems Corp. · Led Light Source802 reports
- SpyGlass DS M00546650Boston Scientific Corporation · Led light source151 reports
- LUCEA LED® ARDLCA209016AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA506006AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA109008AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA506000AMaquet Sas · Led light source1 recall0 reports
- Olympus WA91502AOLYMPUS Winter & Ibe GmbH · Led Light Source28 reports
- VISERA S VIDEO SYSTEM CENTER OLYMPUS OTV-S500Olympus Medical Systems Corp. · Led Light Source7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page