VIDEO SYSTEM CENTER CV-170Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of VIDEO SYSTEM CENTER (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus video system center has 808 adverse event reports, with no deaths and no injuries. Many describe no display or image, a communication error with the scope, and noisy or interrupted images, and most mention no effect on the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 811
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 811 adverse event reports that list this exact device (DI 04953170338496), received between Aug 8, 2021 and Sep 28, 2026 — classified as 811 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (261 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image261
- Communication or Transmission Problem231
- Electrical /Electronic Property Problem50
- Output Problem47
- Poor Quality Image41
- Erratic or Intermittent Display35
- Connection Problem28
- Failure to Power Up26
Patient problems
- No Clinical Signs, Symptoms or Conditions810
- Discomfort1
Recent adverse event reports
The 25 most recent of 811 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026No Display/Image; Moisture Damage; Activation Failure
The customer reported that the Video System Center light did not turn on when the duodenovideoscope was connected. The video cable was disconnected and reconnected to the processor; however, following reconnection, the image was no longer displayed on the monitor. After disconnecting the cable a second time, water was observed within the connector. The event occurred during a therapeutic endoscopic retrograde cholangiopancreatography (ERCP) procedure while the patient was under sedation, and the device was in use. No patient harm or adverse clinical impact was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50920628 ↗MalfunctionSep 28, 2026Communication or Transmission Problem
The customer reported a scope communication error B030 during inspection for use involving an Olympus Video System Center. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916240 ↗MalfunctionSep 21, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50848794 ↗MalfunctionSep 13, 2026No Display/Image; Mechanical Problem; Activation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50747875 ↗MalfunctionSep 11, 2026No Display/Image; Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50732980 ↗MalfunctionSep 8, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50673463 ↗MalfunctionAug 19, 2026Degraded; Output Problem; Activation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50322015 ↗MalfunctionAug 18, 2026Communication or Transmission Problem
The customer reported B30 scope communication error during set up involving an Olympus video system center. There was no patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50300279 ↗MalfunctionAug 17, 2026Erratic or Intermittent Display
The customer returned the video system center to the service center for a separate issue. During the device analysis, the service repair technician indicates that device had interrupted image. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50276778 ↗MalfunctionAug 11, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50214157 ↗MalfunctionAug 11, 2026Incomplete or Inadequate Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50213779 ↗MalfunctionJul 27, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50025480 ↗MalfunctionJul 27, 2026Incomplete or Inadequate Connection
No additional information received from customer.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50025277 ↗MalfunctionMay 29, 2026Poor Quality Image
THE CUSTOMER RETURNED THE VIDEO SYSTEM CENTER, WHICH IS AN OLYMPUS ASSET WITH SEPARATE CASE. DURING THE EVALUATION OF THE DEVICE, NOISY IMAGE WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF NOISY IMAGE WAS DUE TO LOOSEN LOCKING LEVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25331814 ↗MalfunctionMay 15, 2026Poor Quality Image
IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER SCREEN WAS DARK. THE EVENT OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25184287 ↗MalfunctionMay 8, 2026Connection Problem
IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER WAS NOT RECOGNIZED ON THE SCOPE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25125956 ↗MalfunctionApr 23, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER EXHIBIT ERROR B30 DURING INSPECTION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24975232 ↗MalfunctionApr 23, 2026Communication or Transmission Problem
IT WAS REPORTED THE VIDEO SYSTEM CENTER EXHIBITED NOT RECOGNIZED WHEN CONNECTED TO A SCOPES. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966330 ↗MalfunctionApr 20, 2026Connection Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER ENDOSCOPE CABLE CANNOT BE CONNECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24929927 ↗MalfunctionApr 17, 2026Degraded; Infusion or Flow Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER HAD A TUBE CARBONIZATION WHICH PREVENTS AIR FEED FROM MEETING STANDARD VALUES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24916539 ↗MalfunctionApr 16, 2026No Display/Image; Activation Failure
IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER EXPERIENCED A SCREEN BLACKOUT AND SYMPTOM ERRORS. THIS ISSUE WAS IDENTIFIED DURING INSPECTION PRIOR TO USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24904241 ↗MalfunctionApr 14, 2026No Display/Image; Moisture Damage
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER EXHIBITED NO PROJECTED IMAGE AND SHOWED FLUID INVASION ON THE CIRCUIT BOARD (AP BOARD). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24877215 ↗MalfunctionApr 13, 2026Loose or Intermittent Connection; Poor Quality Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER HAD STRIPES IN THE IMAGE AND CONNECTOR NOT LOCKED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24854108 ↗MalfunctionApr 10, 2026No Device Output
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER EXHIBITED FAULTY MAIN LAMP AND LIGHT EMITTING DIODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24838380 ↗MalfunctionApr 6, 2026Poor Quality Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER PRESENTED IMAGE NOISE IN THE ENTIRE DISPLAY AND MAIN BOARD FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790109 ↗
Other Led Light Source devices
- SpyGlass DS M00546650Boston Scientific Corporation · Led light source154 reports
- LUCEA LED® ARDLCA506006AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA506000AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA209016AMaquet Sas · Led light source1 recall0 reports
- LUCEA LED® ARDLCA109008AMaquet Sas · Led light source1 recall0 reports
- Olympus WA91502AOLYMPUS Winter & Ibe GmbH · Led Light Source29 reports
- n.a. TL300Karl Storz GmbH & Co. KG · Led light source11 reports
- VISERA S VIDEO SYSTEM CENTER OLYMPUS OTV-S500Olympus Medical Systems Corp. · Led Light Source9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page