Babyleo TN500 2M60000Recalls & reported problems

Drägerwerk AG & Co. KGaA · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
1
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04048675436607), received between Jul 5, 2019 and Feb 13, 2026 — classified as 1 death, 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event2
  • Use of Device Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Decreased Sensitivity1
  • Device Sensing Problem1
  • Excessive Heating1
  • Insufficient Heating1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Bradycardia1
  • Burn(s)1
  • Death1
  • Fever1
  • Hypernatremia1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 13, 2026Insufficient Heating; Use of Device Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT EXTREMELY PREMATURE (26 WEEKS + 4 DAYS), TRANSFERRED TO THE DEPARTMENT IN A HURRY FOR AN EMERGENCY CAESAREAN SECTION. TEMPERATURE OF 34.5°C AT BIRTH, NEWBORN PLACED IN AN INCUBATOR WITH TEMPERATURE INCREASE IN MANUAL MODE AND OBSERVATION THAT THE TEMPERATURE PROBE WAS NO LONGER IN PLACE, EXPLAINING THE LOW TEMPERATURE. THE TEMPERATURE PROBE WAS REPLACED ON THE NEWBORN, AND IN THE RUSH BEFORE EMERGENCY INTUBATION, AN ERROR WAS MADE BY THE PROFESSIONAL WHO DID NOT PROPERLY VALIDATE THE ¿SKIN MODE¿ ON THE INCUBATOR (REQUIRING TWO PRESSES OF THE BUTTON). THE SKIN MODE ALLOWS THE TEMPERATURE TO BE ADJUSTED TO THE NEWBORN'S TEMPERATURE, WHEREAS THE DEFAULT MANUAL MODE INCREASES THE INCUBATOR TEMPERATURE ONLY DURING THE FIRST 15 MINUTES OF USE. DURING INTUBATION, THE INCUBATOR DISPLAYS 34.7 °C (TOO LOW) AND THE HEART RATE IS MEASURED AT BETWEEN 106 AND 118/MINUTE. FOLLOWING INTUBATION, THE NEWBORN'S TEMPERATURE IS MEASURED IN THE ARMPIT AT 33°C: IT IS THEN NOTED THAT THE INCUBATOR - PREVIOUSLY IN MANUAL MODE - IS NO LONGER HEATING (THE 15 MINUTES OF USE HAVE BEEN REACHED). THE PROFESSIONAL CERTAINLY MISTAKENLY THOUGHT THAT THE TEMPERATURE PROBE WAS NO LONGER IN PLACE,

    Patient problems: Bradycardia

    MAUDE report 24331163 ↗
  • MalfunctionJan 12, 2026Use of Device Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT INCONSISTENT TEMPERATURES WERE DISPLAYED ON THE DEVICE. UNIT PULLED FROM SERVICE. BCHS0005001 SN: (B)(6) BABYLEO TESTED WITH 2 NEW DISPOSABLE PROBES. PN:MX11000 LOT#241220 EXP.DATE:2027-11-19 STATIC TEMPERATURE TEST READ AT 36 DEGREES WITH A SET TEMPERATURE OF 36 DEGREES FOR BOTH CENTRAL AND PERIPHERAL USING TEST KIT 7911314 (DRAEGER) THIS SHOWS THE TN500 ITSELF WAS READING RESPECTIVE RESISTANCES PROPERLY. TESTING WITH THE 2 DISPOSABLE TEMPERATURE PROBES (USING THE FLUKE INCU II SKIN TEMPERATURE TEST KIT) (PROBE 1 SET TO 36.04 AND WAS READ AT 35.2 DEGREES. SHOWING A FAIL WITH PROBE 1) - SHOWING A CONSISTENCY WITH THE REPORTED FAILURE. (PROBE 2 SET TO 36.00 AND WAS READ AT 35.2 DEGREES. SHOWING A FAIL WITH PROBE 2) - SHOWING A CONSISTENCY WITH THE REPORTED FAILURE. BOTH PROBES HAD A RESISTANCE OF 1458 OHMS WHEN INCU II WAS SET TO 36.0 DEGREES. THE UNIT IS PROGRAMMED FOR 1412 OHMS AT 36 DEGREES AS PER SERVICE MANUAL. MAIN ISSUE : ¿BABYLEO IN NICU OVERHEATED BABY GIVING THEM A FEVER, THIS DELAYED DISCHARGE BABY'S ADMISSION TEMP 38.1C AND WAS PLACED UNDER NICU 9A BABYLEO WITH SKIN TEMP PROBE PLACED AND WARMER SET TO "SKIN MODE" AT 1320. BABY'S TEMPERATURE TRENDING

    Patient problems: Fever

    MAUDE report 24035526 ↗
  • InjuryMar 28, 2023Adverse Event Without Identified Device or Use Problem; Excessive Heating; Adverse Event Without Identified Device or Use Problem; Excessive Heating

    IT WAS REPORTED, THAT A NEONATE WITH AN APPROXIMATE WEIGHT OF 700 GRAMS GAINED SKIN BURN AT THE BACK DURING USE WITH THE HEATED MATTRESS IN A BABYLEO TN500. BASED ON THE CURRENT INFORMATION A FAILURE AT THE DEVICE / MATTRESS THAT RESULTED IN A SKIN BURN CANNOT BE EXCLUDED.

    Patient problems: Burn(s)

    MAUDE report 16626093 ↗
  • InjuryDec 22, 2020Unexpected Therapeutic Results; Decreased Sensitivity; Device Sensing Problem; Unexpected Therapeutic Results; Decreased Sensitivity; Device Sensing Problem

    IT WAS REPORTED BY THE USER: ¿PREMATURE BABY IN THE INCUBATOR, CONSPICUOUS WITH INCREASING LIFE-THREATENING HYPERNATREMIA (MAX. 160 MMOL/L) AND HYPERKALEMIA (MAX 7.5 MMOL/L) DESPITE VERY HIGH FLUID INTAKE UP TO 220ML/(KG*H). NO WEIGHT GAIN, LITTLE URINE EXCRETION AND TEMPERATURE LABILITY DESPITE THE ATYPICALLY HIGH INCUBATOR TEMPERATURE OF 37°C. WHEN VIEWED TOGETHER, THE ONLY POSSIBLE CAUSE IS VERY STRONG TRANSDERMAL FLUID LOSS. INCUBATOR REGULATED UP FROM 80% TO 99% RELATIVE HUMIDITY (STATED MEASURED VALUES OF THE INCUBATOR IN THE EXPECTED RANGE, BUT NO PRECIPITATION ON THE INCUBATOR WALLS). THEREFORE THE INCUBATOR WAS REPLACED, WHICH RESULTED IN THE CHILD GRADUALLY STABILIZING.¿

    Patient problems: Hypernatremia

    MAUDE report 11054309 ↗
  • DeathJul 5, 2019Therapeutic or Diagnostic Output Failure

    PLEASE REFER TO THE INITIAL REPORT.

    Patient problems: Death

    MAUDE report 8763674 ↗

Other Incubator, Neonatal devices