Skip to content

Isolette® 8000 plus MU20602Recalls & reported problems

Draeger Medical Systems, Inc. · I8000 plus · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04049098090612), received between Jun 19, 2023 and Jul 3, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Contamination with Chemical or Other Material1
  • Excessive Heating1
  • Material Opacification1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2024Material Opacification; Adverse Event Without Identified Device or Use Problem; Material Opacification; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT THE RESULT OF THEIR MEASUREMENTS WAS SIGNIFICANTLY GREATER X-RAY ATTENUATION IN THE BASE OF THE DRAEGER ISOLETTE WHEN COMPARED TO THE CUSTOMER OTHER NON-DRAEGER INCUBATORS. THE CUSTOMER STATED THAT, WHEN USING THE SAME EXPOSURE PARAMETERS(70KV &0.5MAS), THE MEASURED EXPOSURE TO THE IMAGE RECEPTOR IN THE NEW INCUBATOR WAS 45-48% LESS THAN COMPARED TO THEIR OTHER INCUBATOR. IN ORDER FOR THE USERS TO ACHIEVE A BETTER EXPOSURE, THEY INCREASED THE PARAMETERS FROM .05MHS TO .08MHS, AND STATED THAT THIS WOULD INCREASE THE PATIENT DOSE BY ABOUT 60%. THERE WAS NO PATIENT INJURY OR ADVERSE PATIENT IMPACT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19666846 ↗
  • InjuryFeb 1, 2024Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT THE DEVICE CONTINUED TO HEAT AT MAXIMUM POWER IN SKIN MODE EVEN THOUGH THE SKIN TEMPERATURE WAS GOOD. WHICH LED TO A BABY HYPERVENTILATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18621689 ↗
  • MalfunctionJun 19, 2023Smoking; Device Contamination with Chemical or Other Material; Smoking; Device Contamination with Chemical or Other Material

    THE CUSTOMER REPORTED A CONTINUOUS BURNING RUBBER SMELL IN THE INCUBATOR WHEN IN USE. THE INCUBATOR IS IN THE BIOMEDICAL DEPARTMENT BECAUSE THE STAFF IS CONCERNED ABOUT THE SMELL BEING TOXIC FOR THE BABY INSIDE THE INCUBATOR. NO ADVERSE PATIENT IMPACT WAS REPORTED. AS THE CAUSE OF THE SMELL HAS NOT BEEN DETERMINED, WE CANNOT EXCLUDE THE ODOR POTENTIALLY PERMEATING INTO THE PATIENT COMPARTMENT OR DETERMINE IF THE SOURCE IS TOXIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17153857 ↗

Other Incubator, Neonatal devices

All Incubator, Neonatal devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.