Skip to content

EcoFit® cup cementless incl. central hole cover 02200064Recalls & reported problems

implantcast GmbH · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 10, 2021Class IITerminated

    A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.

    Z-2050-2021 · Root cause: Device Design · Recall details · FDA record ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page