EcoFit cup cementless recallZ-2050-2021 · Implantcast GMBH · initiated May 10, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Implantcast GMBH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.
FDA root cause category: Device Design
What the company is doing
On May 10, 2021 Urgent Safety Information notices were emailed to customers who were asked to take the following action: 1. Read this safety information carefully and make sure all relevant departments and officeholders are informed about its content. 2. Keep this safety information. 3. Fill in and return the reply form. Customers with additional questions were encouraged to contact Wendy Ryan-Kettler at [email protected] On 17 May, 2017, a follow-up Medical Device Correction notice was emailed to customers who were asked to take the following additional actions: 4. Destroy the affected (previous) surgical technique and instruction for use (IFU) and replace them with the updated versions that accompanied this safety information.
Recall details
- Recall number
- Z-2050-2021
- Status
- Terminated
- Initiated by the company
- May 10, 2021
- Posted by the FDA
- Jul 9, 2021
- Recalling firm
- Implantcast GMBH
- Quantity
- 41 units (U.S.)
- Product code
- LZO
- FDA submissions
- K163577
- Distribution
- U.S. Nationwide distribution in the states of FL and TX.
Product description
EcoFit cup cementless: Model(Catalog Number)/Description: 02200046 /EcoFit cup cementless 46mm dia incl. central hole cover; 02200048 /EcoFit cup cementless 48mm dia incl. central hole cover; 02200050 /EcoFit cup cementless 50mm dia incl. central hole cover; 02200052 /EcoFit cup cementless 52mm dia incl. central hole cover; 02200054 /EcoFit cup cementless 54mm dia incl. central hole cover; 02200056 /EcoFit cup cementless 56mm dia incl. central hole cover; 02200058 /EcoFit cup cementless 58mm dia incl. central hole cover; 02200060 /EcoFit cup cementless 60mm dia incl. central hole cover; 02200062 /EcoFit cup cementless 62mm dia incl. central hole cover; 02200064 /EcoFit cup cementless 64mm dia incl. central hole cover; 02200066 /EcoFit cup cementless 66mm dia incl. central hole cover; 02200068 /EcoFit cup cementless 68mm dia incl. central hole cover;
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Instructions for Use Reference Number: ECOCUOPU and Surgical Technique Reference Number:09300038US Model(Catalog Number)/Description: 02200046 /EcoFit¿ cup cementless ¿ 46mm incl. central hole cover; UDI: 04048844005368 Lot Numbers: 1736252001; 1736252003; 1736252004; 1736252006; 1736252007; 1736252008; 1736252010; 18481C0027; 18481C0010; 18481C0004; 18481C0013; 18481C0005; 18481C0024; 02200048 /EcoFit¿ cup cementless ¿ 48mm incl. central hole cover; UDI: 04048844005375 Lot Numbers: 17420NM001 17420NM002 17420NM003 17420NM005 17420NM007 17420NM010 17420NM011 17420NM012 17420NM013 17420NM014 17420NM015 17420NM016 17420NM018 17420NM019 17420NM020 17420NM021 17420NM022 17420NM023 17420NM025 17420NM026 02200050 /EcoFit¿ cup cementless ¿ 50mm incl. central hole cover; UDI: 04048844005382 Lot Numbers: 174111N008 174111N012 174111N018 190414M002 190414M003 190414M004 190414M012 190414N016 174111N002 174111N005 174111N006 174111N013 174111N022 174111N023 174111N026 190625J027 190625J026 190625J028 190625J009 190625J012 190625J014 190625J021 190625J020 190625J001 190625J017 190625J006 190625J002 190625J004 190625J018 02200052 /EcoFit¿ cup cementless ¿ 52mm incl. central hole cover; UDI: 04048844005399 Lot Numbers: 1736253002 1736253005 1736253007 1736253008 1736253015 1736253016 1736253017 1736253021 1736253004 1736253006 1736253014 1736253019 1736253020 1736253022 1736253023 190909E021 190909E005 190909E014 190909E004 190909E017 190909E025 190909E002 190909E018 190909E023 190909E006 190909E009 190909E001 02200054 /EcoFit¿ cup cementless ¿ 54mm incl. central hole cover; UDI: 04048844005405 Lot Numbers: 1748194005 1748194007 1748194008 1748194011 1748194013 1748194016 1748194022 1748194023 1748194025 1748194026 1748194001 1748194003 1748194006 1748194012 1748194017 1748194018 1748194020 1748194028 1849208014 1849208004 1849208012 185019I006 1849208001 1849208006 185019I004 1849208009 02200056 /EcoFit¿ cup cementless ¿ 56mm incl. central hole cover; UDI: 04048844005412 Lot Numbers: 174817C002 174817C004 174817C009 174817C010 174817C011 174817C013 17481M2002 17481M2007 174817C001 174817C003 174817C005 174817C007 174817C008 174817C012 17481L7001 17481M2005 17481M2008 17481M2011 19061JB024 19061JB008 19061JB005 19061JB003 19061JB020 19061JB006 19061JB023 19061JB002 19061JB019 02200058 /EcoFit¿ cup cementless ¿ 58mm incl. central hole cover; UDI: 04048844005429 Lot Numbers: 1738104009 173810D001 173810D002 173810D003 173810D004 173810D006 173810D007 173810D008 173810D009 173810D012 173810D014 173810D015 173810D005 173810D010 173810D011 173810D013 02200060 /EcoFit¿ cup cementless ¿ 60mm incl. central hole cover; UDI: 04048844005436 Lot Numbers: 1738116001 1738116002 1738116003 1738116004 1738116005 1738116006 1738116008 1738116018 1738116019 1738116021 1738116023 1738116025 1738116007 1738116010 1738116013 02200062 /EcoFit¿ cup cementless ¿ 62mm incl. central hole cover; UDI: 04048844005443 Lot Numbers: 1741126001 1741126003 1741126004 1741126005 1741126010 1741126011 1741126012 1741126015 1741126017 1741126020 15190KC048 1741126013 1741126016 1741126006 02200064 /EcoFit¿ cup cementless ¿ 64mm incl. central hole cover; UDI:04048844005450 Lot Numbers: 1741104001 1741104002 1741104003 1741104005 1741104007 1741104018 1741104020 1741104021 1741104022 1741104010 1741104004 1741104019 02200066 /EcoFit¿ cup cementless ¿ 66mm incl. central hole cover; UDI: 04048844005467 Lot Numbers: 174110E002 174110E004 174110E007 174110E008 174110E009 174110E011 174110E013 174110E003 174110E012 174110E010 174110E006 174110E005 02200068 /EcoFit¿ cup cementless ¿ 68mm incl. central hole cover; UDI: 04048844005474 Lot Numbers: 17410JK001 17410JK002 17410JK003 17410JK004 17410KM001 17410KM003 17410KM004 17410KM005 17410KM006 17410KM007 17410KM009 17410KM010 17410KM011 1416I6G089 1439O4K008
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- EcoFit® cup cementless incl. central hole cover 02200046implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200048implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200050implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200052implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200054implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200056implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200058implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200060implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200062implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200064implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200066implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
- EcoFit® cup cementless incl. central hole cover 02200068implantcast GmbH · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page