Infinity CentralStation Wide (confi) MS26800Recalls & reported problems
Draeger Medical Systems, Inc. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a Dräger Infinity CentralStation, the hub that collects and displays vital signs from bedside telemetry monitors. Reports mostly describe the system dropping patients from the central display on its own, unexpected restarts that interrupted monitoring, missing alarm sounds, and ECG waveforms displayed with missing peaks. There are 31 reports, 1 death and no injuries. Three open recalls: speakers that may stay silent in analog mode (January 2025), software that drops peaks on narrow ECG waveforms (April 2023), and temporary loss of central monitoring (March 2022).
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 04049098001878), received between Nov 19, 2021 and Nov 10, 2025 — classified as 1 death, 30 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (6 reports).
3 FDA recall records name this device's identifier, 3 still open.
1 other recall cover devices cleared under the same FDA submission (K130711, K231477, K240312); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jan 14, 2025Class IIOpen, Classified
Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
Z-1066-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Apr 5, 2023Class IIOpen, Classified
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Z-1471-2023 · Root cause: Software design · Recall details · FDA record ↗
- Mar 4, 2022Class IIOpen, Classified
Software issue resulting in temporary loss of central monitoring functionality.
Z-1012-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Communication or Transmission Problem6
- No Audible Prompt/Feedback6
- Wireless Communication Problem5
- Telemetry Discrepancy4
- Loss of Data3
- Output Problem3
- Unexpected Shutdown3
- Insufficient Device Problem Information2
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Insufficient Information4
- Tachycardia4
- Bradycardia1
- Unspecified Heart Problem1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2025Telemetry Discrepancy; Communication or Transmission Problem; Telemetry Discrepancy; Communication or Transmission Problem
IT WAS REPORTED THAT THE M300 TELEMETRY MONITOR WAS DISCHARGING ON ITS OWN WITHOUT ANY USER INTERACTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509863 ↗MalfunctionNov 4, 2025Telemetry Discrepancy; Communication or Transmission Problem; Telemetry Discrepancy; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT THE M300 MONITOR IS DISCHARGED A PATIENT AUTOMATICALLY WITHOUT USER INITIATION. THERE WAS NO REPORT OF ADVERSE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23465793 ↗MalfunctionOct 22, 2025Telemetry Discrepancy; Communication or Transmission Problem; Telemetry Discrepancy; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT THE M300 MONITOR IS DISCHARGED A PATIENT AUTOMATICALLY WITHOUT USER INITIATION. THERE WAS NO REPORT OF ADVERSE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23359662 ↗MalfunctionAug 13, 2025Computer Software Problem; Unexpected Shutdown; Computer Software Problem; Unexpected Shutdown
IT WAS REPORTED THAT THE INFINITY CENTRAL STATION (ICS) UNEXPECTANTLY RESTARTED DURING THE EXPORT OF LOGFILES USING THE UNIVERSAL SERIAL BUS (USB) PORT ON THE FRONT OF THE HOUSING. PATIENT MONITORING WAS INTERRUPTED AT THE ICS DURING THE RESTART, BUT MONITORING CONTINUED AT ALL LOCAL MONITORING DEVICES THAT WERE CONNECTED TO THE ICS AT THE TIME OF THIS EVENT. THE USB PORT IN THE REAR OF THE HOUSING WAS THEN USED TO EXPORT THE LOGFILES AND THE DEVICE BEHAVED AS EXPECTED.. ADDITIONALLY, THIS RESULTED IN A CORRUPT LOG FILE THAT COULD NOT BE OPENED. THERE WAS NO REPORT OF ADVERSE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22796512 ↗MalfunctionJul 10, 2025Communication or Transmission Problem; Wireless Communication Problem; Communication or Transmission Problem; Wireless Communication Problem
THE INFINITY M300 PATIENT MONITORS WERE REPORTED TO HAVE DISCHARGED FROM THE INFINITY CENTRAL STATION (ICS) WITHOUT ANY USER INTERACTION. ACCORDING TO THE CUSTOMER, THE ICS DISPLAYED A ¿DISCHARGE¿ MESSAGE AT THE TIME OF THE EVENT, WHILE ALL PATIENT DATA REMAINED ACCESSIBLE IN THE TREND TAB. THE M300 MONITORS WERE SUCCESSFULLY READMITTED, ALLOWING PATIENT MONITORING TO CONTINUE WITHOUT INTERRUPTION. NO ADVERSE PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22475438 ↗MalfunctionJul 10, 2025Communication or Transmission Problem; Wireless Communication Problem; Communication or Transmission Problem; Wireless Communication Problem
THE INFINITY M300 PATIENT MONITORS WERE REPORTED TO HAVE DISCHARGED FROM THE INFINITY CENTRAL STATION (ICS) WITHOUT ANY USER INTERACTION. ACCORDING TO THE CUSTOMER, THE ICS DISPLAYED A ¿DISCHARGE¿ MESSAGE AT THE TIME OF THE EVENT, WHILE ALL PATIENT DATA REMAINED ACCESSIBLE IN THE TREND TAB. THE M300 MONITORS WERE SUCCESSFULLY READMITTED, ALLOWING PATIENT MONITORING TO CONTINUE WITHOUT INTERRUPTION. NO ADVERSE PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22475300 ↗MalfunctionJul 10, 2025Communication or Transmission Problem; Wireless Communication Problem; Communication or Transmission Problem; Wireless Communication Problem
THE INFINITY M300 PATIENT MONITORS WERE REPORTED TO HAVE DISCHARGED FROM THE INFINITY CENTRAL STATION (ICS) WITHOUT ANY USER INTERACTION. ACCORDING TO THE CUSTOMER, THE ICS DISPLAYED A ¿DISCHARGE¿ MESSAGE AT THE TIME OF THE EVENT, WHILE ALL PATIENT DATA REMAINED ACCESSIBLE IN THE TREND TAB. THE M300 MONITORS WERE SUCCESSFULLY READMITTED, ALLOWING PATIENT MONITORING TO CONTINUE WITHOUT INTERRUPTION. NO ADVERSE PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22473721 ↗MalfunctionJan 23, 2025Loss of Data; Unexpected Shutdown; Loss of Data; Unexpected Shutdown
THE CUSTOMER REPORTED THAT THE (ICS) INFINITY CENTRAL STATION REBOOTED ON ITS OWN. APPROXIMATELY TWO HOURS LATER THE STAFF NOTICED THE TRENDS AND FULL DISCLOSURE WERE MISSING. IN ADDITION, IT WAS REPORTED THAT THE ASSOCIATED BEDSIDE MONITOR IS A (M300) TELEMETRY. THERE WAS NO REPORT OF ADVERSE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21217447 ↗MalfunctionDec 19, 2024Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT THE (ICS) INFINITY CENTRAL STATION REBOOTED. WHEN THE (ICS) REBOOTS THERE IS NO PATIENT MONITORING AT THE CENTRAL. IN ADDITION, IT WAS REPORTED THAT THE ASSOCIATED BEDSIDE MONITOR (M300) DID NOT MONITOR A PATIENT DURING THE ICS REBOOT. A LOSS OF MONITORING AT THE CENTRAL AND BEDSIDE MAY DELAY PATIENT TREATMENT AND CARE AND HAS THE POTENTIAL FOR AN ADVERSE EVENT TO OCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20978066 ↗MalfunctionDec 18, 2024Failure to Transmit Record; Loss of Data; Failure to Transmit Record; Loss of Data
IT WAS REPORTED THAT A LOSS OF FULL DISCLOSURE DATA WAS NOTICED ON THE INFINITY CENTRAL STATION WHEN THE CLINICIAN TRIED TO RETRIEVE THE PATIENT DATA. A LOSS OF FULL DISCLOSURE PATIENT DATA COULD IMPACT A CLINICIAN'S ABILITY TO FULLY DIAGNOSE AND OR TREAT A PATIENT. DRAEGER HAS STARTED AN INVESTIGATION, A FOLLOW UP WILL BE PROVIDED WHEN THE INVESTIGATION IS COMPLETE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 20965728 ↗MalfunctionOct 29, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE CUSTOMER REPORTED THAT THE ALARM LIMITS AT THE INFINITY CENTRAL STATION (ICS) COULD NOT BE CHANGED ON THE NIGHT OF SEPT 16 - 17TH. REPORTEDLY, THE ALARM LIMITS COULD NOT BE ADJUSTED TO SUIT THE PATIENT, AND THE DEVICES PROVIDED UNNECESSARY AND EXCESSIVE ALARMING. THIS ISSUE OCCURRED FOR THE TELEMETRY UNITS REGISTERED ON THE ICS. ONE INSTANCE OF THE DEVICE ALARMING WAS WHEN THE HEART RATE LIMITS COULD ONLY BE SET TO 20/MIN OR 70/MIN AS THE LOWER LIMIT FOR HEART RATE AND 120/MIN OR 200/MIN AS THE UPPER LIMIT, WITH NO OTHER OPTIONS. ONE PATIENT HAD 500 BRADYCARDIA ALARMS IN FOUR HOURS , AND THE PROBLEM ALSO CAUSED ALARMS TO RING CONTINUOUSLY THROUGHOUT THE NIGHT FOR SEVERAL PATIENTS. THE CUSTOMER REPORTED THAT THE "REAL ALARMS" WERE NOT AS EFFECTIVELY DETECTED AMONG THE "UNNECESSARY" RINGING, AND THAT A BRADYCARDIA (22/MIN) WAS MISSED, BUT THE NURSES DETECTED THE ISSUE. NO ADVERSE PATIENT IMPACT OR DEATH WAS REPORTED.
Patient problems: Bradycardia
MAUDE report 20557866 ↗MalfunctionSep 30, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT THE INFINITY CENTRAL STATION "STOPPED" ON (B)(6), 8:3PM. NO DETAILS ON THE NATURE OF THE "STOP" WAS PROVIDED. IT WAS THEN RESTARTED AT 9:53 PM. THERE WAS NO REPORT OF ADVERSE PATIENT IMPACT OR PATIENT DEATH.
Patient problems: Insufficient Information
MAUDE report 20339137 ↗MalfunctionSep 12, 2024Inappropriate Audible Prompt/Feedback; No Audible Prompt/Feedback; Inappropriate Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER REPORTED THAT THERE WAS NO SP02 AUDIO ALARMS AT THE CENTRAL STATION, ONLY VISUAL ALARMS. THE DRAEGER TECHNICIAN CONFIRMED THIS REPORT. THE TECHNICIAN FOUND THAT THE ALARM SETTINGS ON THE CENTRAL HAD BEEN CHANGED, AND CHANGED THE SETTINGS. ONCE THE SETTINGS WERE CHANGED, THE AUDIBLE ALARMS IMMEDIATELY FUNCTIONED AS INTENDED. IT WAS REPORTED THAT THE CUSTOMER'S TECHNICAL SUPPORT AND CLINICAL HAD NOT BEEN AWARE THAT THIS SETTING CHANGE WAS POSSIBLE. THE PATIENT WAS REPORTEDLY IN THEIR ROOM, AND THE CUSTOMER STATES THAT THERE WAS NO DELAY TO PATIENT TREATMENT AND NO PATIENT INJURY. THERE WAS NO ADVERSE PATIENT IMPACT OR DEATH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20205099 ↗MalfunctionJul 8, 2024No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER REPORTED THAT THE TELEMETRY MONITOR DID NOT ALARM FOR A SHORT VENTRICULAR TACHYCARDIAC EVENT THAT LASTED 9 BEATS WHILE THE ALARM WAS SET TO MEDIUM. THE EVENT WAS SHOWN TO BE STORED AS AN EVENT AT THE ICS. THE ARRYTHMIA WAS WITNESSED BY THE STAFF PRESENT AT THE BEDSIDE. ADDITIONALLY, THE CUSTOMER REPORTED THAT PATIENT INFORMATION WAS MISSING, BUT DID NOT CONFIRM WHAT PATIENT INFORMATION. NO PATIENT HARM OR ADVERSE PATIENT IMPACT WAS REPORTED. SINCE THE REPORTED MALFUNCTION, THE DEVICE HAS CORRECTLY ALARMED.
Patient problems: Tachycardia
MAUDE report 19690424 ↗MalfunctionJun 28, 2024No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER REPORTED THAT AT 2130 IT IS NOTED THAT THE TELEMETRY ONLY SHOWS VISUAL ALARMS, NO AUDIBLE ALARM SOUND IS HEARD, ALTHOUGH THE SOUND VOLUME IS AT 100 PERCENT. IT IS UNCLEAR HOW LONG THIS HAS BEEN THE CASE. EARLIER IN THE DAY, THE CONTROL CENTER FOR THE TELEMETERS WAS CHANGED. THERE IS NO REPORT OF ANY ADVERSE EVENT OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19632542 ↗MalfunctionJun 28, 2024Output Problem; Output Problem
THE CUSTOMER REPORTED THAT THE PATIENT DATA AFTER A HIGH FREQUENCY OF 150 BPM WAS NOT ABLE TO BE REVIEWED IN THE TREND. THE CUSTOMER CONFIRMED THAT THE BEDSIDE MONITOR CONTINUED TO REAL TIME MONITORING. THE CUSTOMER REPORTED THAT THE INABILITY TO REVIEW THE HISTORICAL TREND OF THE EVENT COULD POTENTIALLY IMPACT DIAGNOSIS AND TREATMENT. THE USER SWITCHED OUT THE TELEMETRY UNITY TO AMEND THE ISSUE. NO ADVERSE PATIENT IMPACT WAS REPORTED.
Patient problems: Insufficient Information; Unspecified Heart Problem
MAUDE report 19632481 ↗MalfunctionJun 28, 2024Telemetry Discrepancy; Inaudible or Unclear Audible Prompt/Feedback; Output Problem; Telemetry Discrepancy; Inaudible or Unclear Audible Prompt/Feedback; Output Problem
THE CUSTOMER REPORTED THAT ECG ALARMS/EVENTS ARE NOT RECORDED, AND THAT PATIENT DATA IS MISSING. ADDITIONALLY, THE CUSTOMER REPORTED THAT, AFTER UPDATING THE CENTRAL TELEMETRY SYSTEM, THE VOLUME OF ALARMS IS MUCH LOWER THAN BEFORE, AND COULD NOT BE CHANGED, STATING THAT USERS DO NOT REACT TO THE NEW ALARMS. ADDITIONALLY, THE UPDATE CHANGED THE DEFAULT SETTINGS FOR THE ALARM LIMITS. THE CUSTOMER CONFIRMED NO ADVERSE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19632475 ↗MalfunctionJun 28, 2024Insufficient Device Problem Information; Wireless Communication Problem; Insufficient Device Problem Information; Wireless Communication Problem
THE CUSTOMER REPORTED THAT THE TELEMETRY PATIENT INFORMATION WAS NOT DISPLAYED AT THE CENTRAL FOR HOURS AFTER THE PATIENT WAS CONNECTED TO THE TELEMETRY UNIT, AND THAT NO ALARMS AND EVENTS WERE REGISTERED, AND NO ANALYSIS OF THE PATIENT HEART WAS ABLE TO BE CONDUCTED DURING THAT TIME. AFTER THE USERS EXCHANGED THE TELEMETRY UNITY, REAL TIME MONITORING WAS RESTORED. NO ADVERSE PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19632470 ↗MalfunctionMay 24, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT THE USB CONNECTION CABLE OF THE DRÄGER KEYBOARD IS BURNT, AND THAT THE CUSTOMER REPLACED THE KEYBOARD WITH A NON-DRAEGER KEYBOARD. THERE WAS NO REPORT OF ANY PATIENT OR USER INJURY OR DEATH.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 19391830 ↗MalfunctionMay 17, 2024Wireless Communication Problem; Wireless Communication Problem
THE CUSTOMER REPORTED THAT THE M300 PLUS MADE A BEEP, REBOOTED A FEW TIMES AND THEN THE PATIENT WAS DISCHARGED ON THE CENTRAL. THE CUSTOMER CHANGED THE BATTERY AND ARE KEEPING THE DEVICE IN USE. NO REPORT OF PATIENT INJURY OR DEATH WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19340974 ↗MalfunctionApr 22, 2024No Audible Prompt/Feedback; No Visual Prompts/Feedback; No Audible Prompt/Feedback; No Visual Prompts/Feedback
THE CUSTOMER REPORTED THAT THE INFINITY CENTRAL STATION FAILED TO ALARM FOR VENTRICULAR TACHYCARDIA ON (B)(6) AND (B)(6). THERE WAS NO REPORT OF PATIENT INJURY OR PATIENT DEATH.
Patient problems: Tachycardia
MAUDE report 19153155 ↗MalfunctionJan 30, 2024No Audible Alarm; Under-Sensing; No Audible Alarm; Under-Sensing
IT WAS REPORTED THAT THE USERS BECAME AWARE THAT THE EXTERNAL LOUDSPEAKERS FOR THE INFINITY CENTRAL MONITORING STATION WILL STAY INACTIVE AFTER RETURN OF ELECTRICAL POWER FOLLOWING AN OUTAGE OR DISCONNECTION. THERE WAS NO PATIENT INVOLVEMENT REPORTED FOR THE PARTICULAR CASE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 18604341 ↗MalfunctionJan 29, 2024No Audible Alarm; Under-Sensing; No Audible Alarm; Under-Sensing
THE CUSTOMER REPORTED THAT THE POWER SUPPLY TO THE EXTERNAL ACTIVE SPEAKERS, SUPPLIED BY DRÄGER FOR THE ICS (EDIFIER R19UMKII) IS BRIEFLY DISCONNECTED. THE SPEAKERS AUTOMATICALLY GO INTO "USB MODE". ONCE THE POWER SUPPLY IS RESTORED, THE RED LED ON THE LOUDSPEAKER LIGHTS UP AND THE LOUDSPEAKERS NO LONGER OUTPUT ANY SOUND. THERE WAS NO ADVERSE EVENT OR PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18596568 ↗MalfunctionJan 22, 2024Sparking; Sparking
IT WAS REPORTED TO DRAEGER THAT CABLE OF THE COMPUTER MOUSE USED IN CONNECTION WITH THE CENTRAL MONITOR STATION ICS EXHIBITED A SHORT CIRCUIT AND WAS SPARKING. THE EVENT DID NOT LEAD TO HEALTH CONSEQUENCES FOR ANY PERSON.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18552587 ↗MalfunctionDec 21, 2023Capturing Problem; Capturing Problem
THERE WAS NO SPECIFIC EVENT REPORTED - THE USER FACILITY STATED THAT THEY WERE UNABLE TO ESTABLISH 12-LEAD ECG MONITORING FOR CRITICAL PATIENTS PROPERLY. THERE WAS NO DETAIL IN THE REPORT WHICH WOULD REASONABLY SUGGEST THAT HEALTH CONSEQUENCES FOR A PATIENT MAY HAVE OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18370716 ↗
FDA clearance and classification
Cleared through 510(k) K240312 on Aug 1, 2024 (decision: Substantially Equivalent). Submitted by Draeger Medical Systems, Inc. as “Infinity CentralStation Wide”. The FDA review took 181 days. FDA record ↗
Cleared through 510(k) K231477 on Oct 17, 2023 (decision: Substantially Equivalent). Submitted by Draeger Medical Systems, Inc. as “Infinity CentralStation Wide, Infinity M300, Infinity M300+”. The FDA review took 148 days. FDA record ↗
Cleared through Special 510(k) K130711 on Apr 11, 2013 (decision: Substantially Equivalent). Submitted by Draeger Medical Systems, Inc. as “Infinity Centralstation With Infinity M300”. The FDA review took 27 days. FDA record ↗
The FDA regulates product code MHX (monitor, physiological, patient(with arrhythmia detection or alarms)) under 21 CFR 870.1025 as a Class II device. 21 CFR 870.1025 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K130711, K231477, K240312. One submission often covers many models, so check the product and lot details in each record.
- Z-0372-2020 · Oct 9, 2019 · Open, ClassifiedInfinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755, MS26076, MS26031, MS18623, MS22862, MS25301
Common questions about the Infinity CentralStation Wide (confi) MS26800
Has the Infinity CentralStation Wide (confi) MS26800 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-1066-2025 (Class II), began on Jan 14, 2025: Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Infinity CentralStation Wide (confi) MS26800?
The FDA's MAUDE database lists 31 adverse event reports for this device, received between Nov 19, 2021 and Nov 10, 2025. In those reports, the most common specific device problems are “Communication or Transmission Problem” (6), “No Audible Prompt/Feedback” (6) and “Wireless Communication Problem” (5); the most common patient problems are “Tachycardia” (4), “Bradycardia” (1) and “Unspecified Heart Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Infinity CentralStation Wide (confi) MS26800?
1 report is classified as a death, 0 as injuries and 30 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Infinity CentralStation Wide (confi) MS26800 FDA approved?
It was cleared through Special 510(k) K130711 on Apr 11, 2013; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Infinity CentralStation Wide (confi) MS26800 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Infinity CentralStation Wide (confi) MS26800, and is it still sold?
The FDA's device database lists Draeger Medical Systems, Inc. as the company responsible for it (brand Infinity CentralStation Wide (confi), model MS26800). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,537 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,214 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,490 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,164 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths940 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls534 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page