Infinity Central Station. Model Number: MS26800 recallZ-1066-2025 · Draeger Medical Systems, Inc. · initiated Jan 14, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.

FDA root cause category: Device Design

What the company is doing

Draeger notified consignees on about 01/14/2025 via Urgent - Medical Device Correction letter. Consignees were instructed to inspect the device to identify if the speaker is configured with the ICS in USB mode or Analog mode. For the speaker set to Analog mode there is an analog cable (aux cable) connected to the backside of the speaker and a light on frontside of the speaker display. If the light is green, there is no issue of speaker audio annunciation. If the light is red, the issue of speaker audio annunciation is present. If an issue with alarm annunciation occurs while operating, follow the steps included in the letter. Ensure all affected users or customers, if further distributed, are notified of the recall notification. They were instructed to complete and return the provided reply card.

Recall details

Recall number
Z-1066-2025
Status
Open, Classified
Initiated by the company
Jan 14, 2025
Posted by the FDA
Jan 30, 2025
Recalling firm
Draeger Medical Systems, Inc.
Quantity
1,013 units
Product code
MHX
FDA submissions
K151860
Distribution
Worldwide - US Nationwide distribution in the states of AL, AS, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, NE, NY, OK, PA, SD, TN, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Botswana, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kenya, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Romania, Samoa, American, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkiye, United Kingdom.

Product description

Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Infinity CentralStation (ICS) Model Number: MS26800. UDI-DI Number: 04049098001878. Edifier Speaker Model Number: MS34036. UDI-DI Number: 04048675548027. Serial Numbers: 12946M00269, 12946M00314, 12946M00506, 12946M00519, 12946M00518, 12946M00553, 1OQ7WR-00103, 1OQ8CA-00249, 1OQ8CA-00254, 1OQ8CA-00229, 1OQ8CA-00221, 1OQ8CA-00282, 1OQ8CA-00314, 1OQ8CA-00193, 1OQ8CA-00200, 1OQ8CA-00194, 1OQ8CA-00181, 1OQ8CA-00178, 1OQ8CA-00248, 1OQ8CA-00125, 1OQ8CA-00071, 1OQ8CA-00032, 1OQ8CA-00171, 1OQ8CA-00164, 1OQ8CA-00149, 1OQ8CA-00069, 1OQ8CA-00057, 1OQ8CA-00153, 1OQ8CA-00150, 1OQ8CA-00148, 1OQ8CA-00165, 1OQ8CA-00155, 1OQ8CA-00097, 1OQ8CA-00113, 1OQ8CA-00106, 1OQ8CA-00075, 1OQ8CA-00052, 1OQ8CA-00067, 1OQ8CA-00128, 1OQ8CA-00132, 1OQ8CA-00268, 1OQ8CA-00198, 1OQ8CA-00255, 1OQ8CA-00151, 1OQ8CA-00342, 1OQ8CA-00391, 1OQ8CA-00321, 1OQ8CA-00402, 1OQ8CA-00395, 1OQ8CA-00297, 1OQ8CA-00240, 1OQ89M-00691, 1OQ89M-00690, 1OQ89M-00708, 1OQ89M-00693, 1OQ89M-00687, 1OQ89M-00707, 1OQ89M-00703, 1OQ89M-00679, 1OQ89M-00683, 1OQ89M-00692, 1OQ89M-00610, 1OQ89M-00613, 1OQ89M-00669, 1OQ89M-00635, 1OQ89M-00664, 1OQ89M-00697, P1OQ84F-00080, 1OQ89M-00684, 1OQ89M-00676, 1OQ89M-00695, 1OQ8CA-00086, 1OQ89M-00677, 1OQ8CA-00031, 1OQ8CA-00027, 1OQ8CA-00065, 1OQ8CA-00092, 1OQ8CA-00082, 1OQ89M-00602, 1OQ89M-00680, 1OQ89M-00675, 1OQ89M-00593, 1OQ89M-00686, 1OQ89M-00696, 1OQ89M-00681, 1OQ89M-00704, 1OQ89M-00674, 1OQ89M-00661, 1OQ89M-00663, P1OQ89M-00559, P1OQ89M-00540, 1OQ8CA-00134, 1OQ8CA-00095, 1OQ8CA-00043, 1OQ8CA-00074, 1OQ8CA-00762, 1OQ8CA-00688, 1OQ8CA-00670, 1OQ8CA-00637, 1OQ8CA-00633, 1OQ8E8-00037, 1OQ8E8-00053, 1OQ8CA-00773, 1OQ8CA-00683, 1OQ8CA-00500, 1OQ8CA-00476, 1OQ8CA-00449, 1OQ8CA-00420, 1OQ8CA-00460, 1OQ8CA-00463, 1OQ8CA-00507, 1OQ8CA-00350, 1OQ8CA-00442, 1OQ8CA-00433, 1OQ8CA-00447, 1OQ8CA-00601, 1OQ8CA-00604, 1OQ8CA-00695, 1OQ8CA-00513, 1OQ8CA-00575, 1OQ8CA-00577, 1OQ8CA-00591, 1OQ8CA-00197, 1OQ8CA-00516, 1OQ8CA-00541, 1OQ8CA-00542, 1OQ8CA-00568, 1OQ8CA-00499, 1OQ8CA-00498, P12051M00227, 1OQ8CA-00593, 12946M00532, 12946M00256, 12946M00278, 1OQ7WR-00203